An LCR meter is test equipment that measures inductance (L), capacitance (C) and resistance (R) by applying an AC test signal to a component and measuring the resulting impedance. Researchers, core-facility staff and electronics technicians use it to characterise passive components, coils, transformers and sensor elements. The practical point of this page: a reading depends on the test frequency, the equivalent-circuit model and the fixture, so those three choices matter as much as the instrument itself.
What is an LCR meter?
An LCR meter is an instrument that measures inductance, capacitance and resistance. Basic meters convert a measured impedance into an inductance or capacitance value; more advanced meters also report the true L or C, the equivalent series resistance (ESR) of a capacitor and the quality factor (Q) of an inductor, and bridge-type meters can report dissipation factor (Wikipedia, LCR meter, read 2026-11-07). “LCR” is simply the letters for inductance (L), capacitance (C) and resistance (R).
Who uses it: electronics and instrumentation technicians, lab staff maintaining or building instruments, engineers qualifying components, and students learning impedance measurement.
How does an LCR meter work?
The meter applies an AC voltage to the device under test (DUT), measures the voltage across it and the current through it, and divides one by the other to get the impedance magnitude. Advanced meters also measure the phase angle and then compute an equivalent inductance or capacitance and resistance (Wikipedia, read 2026-11-07).
Real components are not ideal. Keysight notes that real capacitors show series resistance and parasitic inductance, and real inductors show winding resistance and parasitic capacitance. Capacitive reactance generally decreases and inductive reactance generally increases as frequency rises (Keysight, Measure passive component impedance, read 2026-11-07). That is why the same part can read differently at 100 Hz and at 100 kHz.
Series or parallel model
The meter must assume a series or parallel equivalent circuit to turn impedance into L, C and R. Inductors with resistance (LR) are usually measured in a series model and capacitors with resistance (CR) usually in a parallel model (Wikipedia, read 2026-11-07). Keysight lists the measurement function and equivalent-circuit model, together with the test frequency, as choices to make before measuring.
Types of LCR meter
| Type | Typical test frequencies | Other features | Source |
|---|---|---|---|
| Handheld | 100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz (typical set) | Example: Keysight U1732C offers 100 Hz, 120 Hz, 1 kHz and 10 kHz; capacitance 20 pF to 20 mF; inductance 20 μH to 2000 H | Wikipedia; Keysight use-case page (both read 2026-11-07) |
| Benchtop | Sometimes above 100 kHz; example Keysight E4980A covers 20 Hz to 2 MHz | Can superimpose DC bias; supports 4-wire Kelvin fixtures for SMD parts, air-core coils and transformers. E4980A: 0.05% basic impedance accuracy, 16 impedance parameters, built-in DC bias of 1.5 V or 2 V, 201-point list sweep | Wikipedia; Keysight E4980A fact sheet (read 2026-11-07) |
Specifications that matter
For the E4980A, Keysight specifies a test signal of 100 μV to 2 Vrms and 1 μA to 20 mA, plus open, short and load compensation and cable extensions of 1, 2 and 4 m. When reading any LCR meter specification, look for what the manufacturer states about frequency range, basic accuracy, test-signal level, DC bias capability and the measurement parameters the meter reports.
Choosing frequency, model and fixture
- Frequency: choose the test frequency first, because reactance changes with frequency (Keysight).
- Model: choose the measurement function and series or parallel equivalent circuit before measuring (Keysight; Wikipedia).
- Fixture: the fixture “should match the component package and expected impedance range” (Keysight). Leads, cables and adapters add to the measured impedance unless they are minimised or compensated.
- Connections: keep them short, mechanically stable and consistent between measurements (Keysight).
Compensation and calibration basics
Keysight describes two fixture corrections. Open compensation reduces stray admittance, such as fixture capacitance. Short compensation reduces residual series impedance from leads, contacts and fixture conductors. Perform compensation with the fixture in the same arrangement used for the component. The E4980A also offers load compensation. For general background on keeping instruments in tolerance, see what is calibration; for a related lab instrument example, see radiation survey meter calibration and response checks.
Common mistakes
- Using a different frequency or model from the one a specification or datasheet assumes, then comparing the readings.
- Measuring with a fixture that does not suit the component package or impedance range.
- Skipping open/short compensation, or compensating with the fixture in a different arrangement from the measurement.
- Letting leads move between measurements.
- Measuring a motor while turning its rotor: Wikipedia cautions that component-oriented LCR meters can be damaged by EMF from turning a permanent-magnet motor rotor.
For another impedance-based technique, see impedance spectroscopy. For a different kind of lab measurement instrument, see what a scanning probe microscope is.
FAQ
What is an LCR meter?
Test equipment that measures inductance (L), capacitance (C) and resistance (R) by applying an AC signal and measuring voltage and current to find impedance.
What is an LCR meter used for?
Measuring the inductance, capacitance and resistance of components, and on advanced meters also capacitor ESR, inductor Q and dissipation factor. Benchtop meters with Kelvin fixtures handle SMD parts, air-core coils and transformers (Wikipedia).
What is the difference between an LCR meter and a multimeter?
An LCR meter applies an AC test signal at a selectable frequency and reports inductance, capacitance and resistance from the measured impedance, using a series or parallel equivalent-circuit model. Because the result depends on the test frequency, fixture and open/short compensation, an LCR meter is the instrument for characterising reactive components, and a general-purpose multimeter is not described that way.
What is meant by LCR meter?
The name comes from the three quantities it measures: inductance (L), capacitance (C) and resistance (R).