Skip to main content
v2026.11,986 entries · CC-BY 4.0

Zero Air Generator: Purity Limits, Types and Maintenance

How zero air generators work, how federal methods define zero air purity, and what EPA says about maintaining and verifying one.

Written and maintained by CASRAI Editorial Board

Published

On this page

A zero air generator is a laboratory instrument that takes an existing compressed air supply and cleans it until the remaining contaminants, usually hydrocarbons, fall below what a gas analyzer can detect. It is used wherever an instrument needs air that reads as “zero” on that instrument, for example fuel air for a gas chromatograph flame ionization detector or baseline gas for an ambient air monitor. How clean “zero” has to be depends on the analyzer and the method.

Short answer

Zero air is air purified so that the species an analyzer measures are not detectable in it. A zero air generator makes it on site from compressed air, in contrast to buying it in a cylinder. There is no single purity figure: the US federal methods quoted below range from less than 10 ppmv volatile organic compounds (EPA Method 21) to less than 0.1 ppmv organic material (Method 25A), and a manufacturer of chromatographic units states total hydrocarbons below 0.1 ppm measured as methane.

Definition and scope: what counts as zero air

The phrase is defined by use, not by a fixed composition. 40 CFR Part 53, which governs testing of reference and equivalent methods for air monitoring, requires that all diluent air be “zero air free of contaminants likely to cause a detectable response on the test analyzer” (40 CFR 53.22, 2011 edition). The same regulation lists two ways to obtain it: ambient air purified by scrubbers or other devices, or a cylinder of compressed zero air certified by the supplier or an independent laboratory. A generator is the first option made into a bench or rack-mounted instrument.

Zero air is not nitrogen. Nitrogen generators separate nitrogen from air and are used as carrier, make-up or blanketing gas, while zero air keeps the oxygen and nitrogen of air and removes the contaminants. If your instrument needs an inert gas, see the nitrogen generator guide; for the safety data of the cylinder alternative, see nitrogen gas SDS.

How a zero air generator works

The manufacturer literature and the EPA procedures describe two practical approaches. Both start from compressed air you already have, so the generator does not compress air itself.

Catalytic converter units for gas chromatography

Airgas describes its chromatographic zero air generators (brass line models) as producing a continuous flow of air from an existing compressed air supply, intended for use with hydrogen generators to automate sampling on a gas chromatograph with a flame ionization detector (GC-FID). Its description of the stages is: an inlet 0.5-micron coalescing filter removes particles, oil and water; hydrocarbons are removed when the air passes over a converter holding a heated catalyst; after cooling, a 0.01-micron cellulose filter removes residual particles. The stated result is total hydrocarbon content below 0.1 ppm measured as methane. These are manufacturer claims for one product line, not an industry-wide specification. For what the detector does with the air, see the guide to gas chromatography columns, carrier gases and detectors.

Scrubber-stack units for ambient air analyzer calibration

The US EPA National Performance Audit Program (NPAP) standard operating procedure (version 2.0, March 2023) uses a zero air generator (ZAG) as the source of zero air when auditing ambient monitors, and lists a Teledyne API 701H “or equivalent” in its equipment. It describes consumables that include activated carbon, Purafil and a particulate filter, plus a molecular sieve, a carbon monoxide scrubber and a dryer. This is a different design from the catalytic unit above: the contaminants are removed by scrubbing media that wear out, not by a heated catalyst.

Design (as described in the sources) Typical use named in the source What removes the contaminants Source
Catalytic converter generator Air for GC-FID, with a hydrogen generator Heated catalyst after a 0.5-micron coalescing filter; 0.01-micron filter after cooling Airgas product literature
Scrubber-based ZAG Zero air for audits of ambient monitors Activated carbon, Purafil, particulate filter; molecular sieve, CO scrubber and dryer as needed EPA NPAP SOP v2.0
Certified cylinder of zero air Alternative source for the same purpose Supplier or independent laboratory certification 40 CFR 53.22, Table B-2; EPA NPAP SOP

How a generator is used and maintained in practice

Two operating details come directly from the EPA procedure and apply to any scrubber-based unit. First, consumables have a replacement cycle: the SOP says the scrubber consumables (activated carbon, Purafil and the particulate filter) are replaced annually, while the molecular sieve, CO scrubber and dryer are replaced as needed. After maintenance the unit is powered on, leak checked, and run in maintenance mode for several days to condition it. Second, the output is verified, not assumed: the SOP checks that the untreated ZAG output reads no more than 0.020 ppm carbon monoxide higher than the output of a palladium-scrubbed ZAG or an ultrapure cylinder of zero air. It also notes that cylinder ultrapure zero air typically measures higher in CO than the ZAG output.

The same document says a ZAG gives off heat that can warm a mobile laboratory, which is a practical siting point. Flow also has to match the demand: Airgas lists models from 1,000 to 30,000 cc/min, a delivery pressure range of 2 to 125 psig and a 117 V AC supply for the brass line family, so the compressed air supply, electrical supply and bench space are all part of the specification. Replacement inlet filters are sold as an option for that line, which reinforces that the inlet filter is a maintenance item.

Where it matters in labs and monitoring

Gas chromatography with flame ionization detection

In GC-FID the detector burns the sample in a hydrogen flame supplied with air. Hydrocarbon impurity in that air raises the baseline, so the air stream is cleaned to a very low total hydrocarbon level. The Airgas literature ties its generators specifically to this job.

Regulatory emissions and air monitoring methods

The 40 CFR Part 60 Appendix A methods name their own zero gas. In Method 21 (leak detection of volatile organic compounds) the zero gas is air with less than 10 ppmv VOC. In Method 25A (total gaseous organic concentration by flame ionization) it is high-purity air with less than 0.1 ppmv organic material (propane or carbon equivalent), or less than 0.1 percent of the span value, whichever is greater. A generator that meets one of these limits does not automatically meet the other. For the measurement side, see volatile organic compound detection.

The data table below compares the figures from each document in one place. The values are quoted from the named sources and are not a recommended specification for any other analyzer.

Common misunderstandings

Zero air is one fixed grade

It is not. The table below shows limits that differ by a factor of one hundred between two federal methods, and the 40 CFR 53.22 wording defines it purely by the analyzer’s response. Always read the method or the instrument manual for the figure that applies.

A generator and a cylinder give identical air

The EPA SOP records a measurable difference: cylinder ultrapure zero air typically measures higher in CO than the ZAG output. That is the reason the procedure includes a verification check between the two.

Zero air removes everything

The Airgas literature quantifies hydrocarbons, expressed as methane. It does not claim removal of every species, and neither does 40 CFR 53.22, which asks only for freedom from contaminants likely to cause a detectable response on the analyzer in use.

A question checklist drawn from the sources

Each question below traces to a statement in the sources above. It is a way to read a datasheet, not buying advice.

  • What contaminant limit does your method or analyzer name (for example 10 ppmv VOC in Method 21, or 0.1 ppmv organic material in Method 25A), and in which units and basis (as methane, as propane)?
  • What flow does the instrument draw, and what inlet pressure and compressed air quality does the generator specify (the Airgas family quotes 2 to 125 psig delivery)?
  • Which consumables does the design use, how often are they replaced, and how long is the conditioning period afterwards (EPA NPAP: annual for carbon, Purafil and particulate filter; several days of conditioning)?
  • How will the output be verified (EPA NPAP: CO difference of no more than 0.020 ppm against a palladium-scrubbed ZAG or ultrapure cylinder zero air)?
  • Where will it sit, and can the room take the heat the unit gives off?

Frequently asked questions

What is zero air used for?

Zero air is used as fuel and carrier air for flame ionization detectors in gas chromatography, as diluent and baseline gas for air quality analyzers, and as the zero gas in regulatory methods such as EPA Methods 21 and 25A. In each case its job is to give the analyzer a reading with no target contaminant in the gas.

What are the downsides of using natural gas generators?

That question is about fuel-burning power generators, which is a different machine from a zero air generator. A zero air generator is a gas purification instrument that runs from an electrical outlet and a compressed air line, for example 117 V AC for the Airgas brass line models. This page does not cover power generators, so it gives no fuel, noise or emissions figures for them.

Which is the best small generator?

This page does not rank products. If you mean a small zero air generator, the lowest flow model in the Airgas brass line family is rated at 1,000 cc/min, and the flow you need is set by the instrument connected to it. If you mean a portable power generator, that is outside this topic.

How long will a 5000 W generator run on 5 gallons of gas?

That is a power generator fuel question and has no connection to zero air generators, which have no fuel tank and are supplied with compressed air and mains electricity. Runtime depends on the specific engine and load, so use the manufacturer’s runtime figures for that unit.

Sources

  • Airgas, “Chromatographic Zero Air, Brass Line Models” product literature, airgas.com, read 2026-11-19. Manufacturer claims.
  • US Code of Federal Regulations, 40 CFR Part 60 Appendix A-7 (Methods 21 and 25A), Legal Information Institute copy, read 2026-11-19.
  • 40 CFR 53.22 and Table B-2, 2011 edition, US Government Publishing Office, read 2026-11-19. The 2011 edition wording is quoted; check the current CFR for later changes.
  • US EPA, National Performance Audit Program Standard Operating Procedure, Version 2.0, March 2023, epa.gov, read 2026-11-19.

Data: Zero air contaminant limits named in US federal methods, an EPA audit SOP and one manufacturer line

Contaminant limits that define zero air in four documents: two EPA reference methods, the EPA NPAP audit SOP and Airgas product literature. Values are quoted from each document with its own analyte and unit; the limits are not interchangeable.

Period covered: 2023/2026. Coverage: United States.

Zero air contaminant limits named in US federal methods, an EPA audit SOP and one manufacturer line
Document and contextContaminant basisLimit (less than, unless stated)UnitQualifier
EPA Method 21 (40 CFR 60 App. A-7), zero gasVolatile organic compounds10ppmvAir; limit as stated in s7.1.1
EPA Method 25A (40 CFR 60 App. A-7), zero gasOrganic material (propane or carbon equivalent)0.1ppmvOr 0.1 percent of span value, whichever is greater
Airgas chromatographic zero air generators, brass lineTotal hydrocarbons as methane0.1ppmManufacturer specification
EPA NPAP SOP v2.0, ultrapure cylinder zero air exampleTotal hydrocarbons0.01ppmAt most; example from certificate of analysis context
EPA NPAP SOP v2.0, ultrapure cylinder zero air exampleCarbon monoxide0.01ppmAt most
EPA NPAP SOP v2.0, ultrapure cylinder zero air exampleNitrogen oxides0.001ppmAt most
EPA NPAP SOP v2.0, ultrapure cylinder zero air exampleSulfur dioxide0.001ppmAt most

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Zero Air Generator: Purity Limits, Types and Maintenance

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →