Direct comparison
Graduated vs Volumetric Pipette: Accuracy & Use
Graduated pipette vs volumetric pipette: ISO 835 and ISO 648 tolerances at 1 and 10 ml, accuracy classes A, AS and B, waiting time and calibration.
Written and maintained by CASRAI Editorial Board
Last updated
How do Volumetric pipette and Graduated pipette compare side by side?
The table below compares Volumetric pipette and Graduated pipette across 11 dimensions, from governing standard to when to pick each.
Side-by-side comparison
| Dimension | Volumetric pipette | Graduated pipette |
|---|---|---|
| Governing standard | ISO 648:2008. Its scope is the metrological and constructional requirements for volumetric pipettes with one mark (total delivery); two-mark volumetric pipettes are also covered. It is written for general laboratory use and follows the principles of ISO 384. | ISO 835:2007 (adopted in Europe as EN ISO 835). Its scope is the metrological and constructional requirements for graduated pipettes adequate for general laboratory purposes, also based on ISO 384. |
| What is marked on the stem | A single volume mark (the one-mark type) that defines the one nominal volume the pipette delivers. A two-mark type exists in the standard as well. | A graduated scale with a zero line and a series of graduations, so the pipette can deliver a range of volumes. Sizes of 1, 2, 5, 10 and 25 ml appear in Table 1 of ISO 835. |
| Delivery behaviour | Total delivery: the nominal volume is what runs out when the pipette is emptied from the mark, under the conditions the standard specifies. | Four types are defined. Type 1: partial delivery, zero line to any graduation. Type 2: total delivery from any graduation down to the jet. Type 3: total delivery, zero line to any graduation, with nominal capacity obtained by delivering down to the jet. Type 4: total delivery with the last drop blown out. |
| Accuracy classes | Class A and Class AS (the higher grade) and Class B (the lower grade). | The same three labels: Class A and Class AS (higher grade) and Class B (lower grade). Type 4 pipettes must be Class B. |
| Maximum permissible error at 10 ml | Class A: ±0.02 ml (the example given in the ISO 648 overview). | Class A or AS: ±0.05 ml. Class B: ±0.1 ml. At 10 ml the graduated limit for Class A/AS is 2.5 times the volumetric Class A limit. |
| Maximum permissible error at 1 ml | Class A: ±0.008 ml (the example given in the ISO 648 overview). | Class A or AS: ±0.007 ml (scale division 0.01 ml). Class B: ±0.01 ml. At 1 ml the tabulated graduated Class A/AS limit is not wider than the volumetric Class A example. |
| Other graduated sizes in ISO 835 | The ISO 648 limits quoted on this page are the 1 ml and 10 ml Class A examples; other volumetric sizes are not compared here. | 2 ml (division 0.02): ±0.010 ml Class A/AS, ±0.02 ml Class B. 5 ml (division 0.05): ±0.030 ml Class A/AS, ±0.05 ml Class B. 25 ml (division 0.1): ±0.1 ml Class A/AS, ±0.2 ml Class B. |
| How the error limit is applied | The limit is the permissible deviation of the one delivered volume from the nominal value. | The limits apply at any point on the scale and also to the maximum difference between the errors at any two points, so a graduated pipette is held to a limit on its whole scale, not at a single mark. |
| Waiting time | Class A and Class B are adjusted for delivery without a waiting time. Class AS is adjusted for a specified waiting time of 5 seconds. | Class A and Class B are adjusted without a waiting time. Class AS has a specified waiting time of 5 seconds, observed after the delivery appears complete and before the final reading. |
| Calibration by weighing | NIST SOP 13 treats one-mark pipets individually: fill to the index mark, deliver with the tip touching the inside wall of the vertical receiver, wait two seconds after flow ceases, remove the pipet, weigh the delivered water and repeat in duplicate. | NIST SOP 13 says measuring pipets are calibrated in a manner similar to burets, that is, at the delivered volumes of interest on the scale rather than at one index mark. |
| When to pick each | Pick the volumetric pipette when the work needs one fixed volume delivered to a tight tolerance, for example the 10 ml Class A limit of ±0.02 ml. | Pick the graduated pipette when the work needs several different volumes from one instrument, and read its class and delivery type (1 to 4) before use; a Type 4 pipette is a blow-out pipette and is always Class B. |
Sources
- ISO 648:2008 overview (standards.iteh.ai) https://standards.iteh.ai/catalog/standards/iso/034d83b6-819a-42cf-9e51-314841ad6998/iso-648-2008
- NIST SOP 13 https://www.nist.gov/document/sop1315nov06pdf
- EN ISO 835:2007 overview (standards.iteh.ai) https://standards.iteh.ai/catalog/standards/cen/bea4bf98-fab1-437f-9df3-7d52c532bafb/en-iso-835-2007
Ask CASRAI · free to try
Ask about Graduated vs Volumetric Pipette: Accuracy & Use
2 free questions a day, no account.
An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.
Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.
Works on this site and inside Claude, Cursor and the AI tools you already use.
Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.
Common questions
Common questions about Volumetric pipette vs Graduated pipette
What is the difference between a graduated pipette and a volumetric pipette?
+
A volumetric pipette is built to deliver one nominal volume from a single mark and is covered by ISO 648:2008. A graduated pipette carries a scale and can deliver a range of volumes; it is covered by ISO 835:2007. Both standards use accuracy classes A, AS and B.
Is a graduated pipette less accurate than a volumetric pipette?
+
Usually, but not at every size. At 10 ml the ISO 835 limit for Class A or AS is ±0.05 ml against ±0.02 ml for a Class A volumetric pipette. At 1 ml the graduated Class A/AS limit is ±0.007 ml against the ±0.008 ml ISO 648 example, so the gap closes at small volumes.
Which graduated pipette type blows out the last drop?
+
Type 4 in ISO 835: total delivery with the last drop blown out. The standard requires Type 4 pipettes to be Class B, the lower accuracy grade.
What does Class AS mean on a pipette?
+
Class AS is a higher-grade class, like Class A, but with a specified waiting time of 5 seconds. The wait is observed after delivery appears complete and before the final reading. Classes A and B are adjusted without a waiting time.
How are volumetric and graduated pipettes calibrated?
+
NIST SOP 13 describes gravimetric calibration of volumetric ware. One-mark pipets are filled to the index mark, delivered and weighed in duplicate; measuring pipets are calibrated in a manner similar to burets.
What do graduated pipette types 1, 2 and 3 mean in ISO 835?
+
Type 1 is partial delivery, from the zero line to any graduation. Type 2 is total delivery from any graduation down to the jet. Type 3 is total delivery from the zero line to any graduation, with the nominal capacity obtained by delivering down to the jet. Type 4 is the blow-out type and must be Class B.
Do ISO 648 and ISO 835 use the same basis?
+
Both are written for general laboratory use and follow the principles of ISO 384, which is why the two standards share the Class A, AS and B labels and can be compared side by side. They differ in what they cover: ISO 648 covers one-mark volumetric pipettes (and two-mark types), ISO 835 covers graduated pipettes.
Data: Maximum permissible error of volumetric and graduated pipettes by class and nominal capacity
Maximum permissible errors in ml from the ISO 648:2008 overview (Class A volumetric examples at 1 and 10 ml) and Table 1 of EN ISO 835:2007 (graduated pipettes, Class A/AS and Class B, 1 to 25 ml), compiled into one long-format table.
Period covered: 2007/2008.
| Pipette standard and class | Nominal capacity (ml) | Maximum permissible error (ml) |
|---|---|---|
| Volumetric pipette, Class A (ISO 648) | 1 | 0.008 |
| Volumetric pipette, Class A (ISO 648) | 10 | 0.02 |
| Graduated pipette, Class A or AS (ISO 835) | 1 | 0.007 |
| Graduated pipette, Class A or AS (ISO 835) | 2 | 0.01 |
| Graduated pipette, Class A or AS (ISO 835) | 5 | 0.03 |
| Graduated pipette, Class A or AS (ISO 835) | 10 | 0.05 |
| Graduated pipette, Class A or AS (ISO 835) | 25 | 0.1 |
| Graduated pipette, Class B (ISO 835) | 1 | 0.01 |
| Graduated pipette, Class B (ISO 835) | 2 | 0.02 |
| Graduated pipette, Class B (ISO 835) | 5 | 0.05 |
| Graduated pipette, Class B (ISO 835) | 10 | 0.1 |
| Graduated pipette, Class B (ISO 835) | 25 | 0.2 |
Sources
- ISO 648:2008 overview, iTeh Standards (accessed 2026-11-18)
- EN ISO 835:2007 overview, iTeh Standards (accessed 2026-11-18)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.