Written and maintained by CASRAI Editorial Board
Published
On this page
A repeat pipette, also called a repeater pipette, repetitive pipette, stepper or multi-dispenser, is a hand-held dispenser that draws up a liquid once and then delivers it in many equal portions without refilling the tip. Eppendorf describes its Repeater M4 as aspirating once and dispensing up to 100 times, and names tube and plate filling and aliquoting reagents as the typical jobs.
Who uses it: laboratory staff who fill long series of tubes, plate wells or vials with the same volume. This page compares what two manufacturers state about their systems, from the instruments to the tips, so you can see what separates a repeat pipette from the single-draw, single-dispense pipette described in What Is a Pipette?
What a repeat pipette is
A repeat pipette is a pipette built for one task: dispensing a series of identical volumes from one filling. BRAND’s brochure for its HandyStep touch puts it this way: repetitive pipettes are “the first choice when it comes to efficient dispensing of long series”, and positive displacement pipettes are “often the right solution for pipetting viscous media”. In the systems described here, the tip works like a small syringe rather than an air-cushion tip.
Eppendorf says it pioneered the Repeater and Combitips system in 1978, the first system that let users fill a tip once and dispense several times. The Repeater M4 is described as the successor of the earlier Repeater plus.
The page that looks closest is Micropipette Types and Uses: Single-Channel vs. Multichannel, which covers repeaters in a short subsection. This page goes into the tips, the volume ranges and the accuracy figures that decide how a repeater is used.
How a repeat pipette is built and how it works
Positive displacement, no air cushion
A standard micropipette moves liquid with a column of air between the piston and the sample. A repeat pipette instead drives a piston that sits inside the tip. Eppendorf states that the Repeater M4 works by positive displacement, with no air cushion, so the delivered volume is correct regardless of the density, viscosity or volatility of the liquid. The Combitips advanced tip carries its own integrated piston, and that piston is hermetically sealed, which Eppendorf says prevents aerosol contamination. The same product page calls the system ideal for handling infectious, toxic or radioactive liquids. For the broader family of tips that suit air-displacement instruments, see What Are Filtered Pipette Tips?
The instrument
The Repeater M4 is a mechanical instrument. Its listed features include a step counter, a volume selection dial with 20 volume settings per tip size, a motion sensor, a tip attachment sensor, and a single lever that fills, dispenses and ejects the tip. It recognises the attached tip size automatically. The Starter Kit specification lists a volume range of 1 μL to 10 mL, a single channel, a button-cell battery and a weight of 105 g.
The tip is the consumable
Each tip is a calibrated syringe, so the tip size sets the volumes the instrument can deliver. This is why a repeat pipette is bought and used as a system of instrument plus matching tips.
Types and variants
The two manufacturers whose documents were read for this page sell two different designs, a mechanical repeater and an electronic stepper. The table lists only features each manufacturer states.
| Type | Example from the source | Stated features | Key difference |
|---|---|---|---|
| Mechanical repeater | Eppendorf Repeater M4 | Positive displacement; 1 μL to 10 mL; 20 volume settings per tip size; step counter; automatic tip recognition; up to 100 dispenses | Closed system: uses Combitips advanced or ViscoTip only, not open to third-party tips; not autoclavable |
| Electronic stepper | BRAND HandyStep touch | Variable volume 1.0 μL to 50 mL; Li-ion battery with inductive charging; modes include MULTI-DISP, AUTO-DISP and PIP; aspiration and dispensing speed adjustable to 8 levels | Recognises BRAND PD-Tips II by coding on the piston; the brochure says third-party tips can also be used |
| Electronic stepper with extra modes | BRAND HandyStep touch S | Adds sequential dispensing, multi-aspirating and titrating to the modes above | More operating modes than the base touch model |
The BRAND brochure explains the speed setting in practical terms: reduced aspiration speed suits highly viscous or foaming media, and reduced dispensing speed gives swirl-free, spray-free dispensing into small vessels.
Tips: sizes and purity grades
Tip sizes
Eppendorf lists nine Combitips advanced volume sizes: 0.1, 0.2, 0.5, 1.0, 2.5, 5.0, 10, 25 and 50 mL, with increments as low as 100 nL depending on the dispenser. For each tip, 20 different dispensing volumes can be set, and between 5 and 100 dispensing steps can be carried out with a tip filled to full capacity, depending on the volume required. Elongated 2.5, 5 and 10 mL tips allow complete emptying of common tubes. The ViscoTip is a 10 mL tip for highly viscous liquids, up to 14,000 mPa·s.
BRAND lists ten PD-Tips II sizes: 0.1, 0.5, 1, 1.25, 2.5, 5, 10, 12.5, 25 and 50 mL. Sterile BIO-CERT liquid-handling variants exist, and the 25 and 50 mL tips need an autoclavable polypropylene adapter.
Purity grades
Eppendorf offers Combitips advanced in five purity grades. They are a useful checklist of the contamination types a lab may need to control.
| Grade | What Eppendorf states |
|---|---|
| Eppendorf Quality | Bulk packaging for routine work |
| PCR clean | Free of human DNA, DNase, RNase and PCR inhibitors |
| Sterile | Sterility assurance level 10-6; pyrogen-free according to Ph. Eur. 2.6.14 |
| Biopur | Individually blister-wrapped and sterile; free of pyrogens, RNases, DNases, DNA, ATP and PCR inhibitors |
| Forensic DNA Grade | Produced to ISO 18385 |
Specifications that matter: accuracy falls at small fractions of the tip volume
The data table below this guide compiles the accuracy and coefficient of variation figures that BRAND publishes for its ten PD-Tips II sizes, measured with distilled water, and adds one calculated column: how much larger the permitted systematic error is at 1% of the tip volume than at 100%. BRAND states that the tolerances specified in ISO 8655 are not exceeded. In every size the permitted error grows as the dispensed volume shrinks relative to the tip. For the 0.1 mL tip it is 1.0% at nominal volume and 8.0% at 1%; for the 50 mL tip it is 0.5% and 3.0%. The coefficient of variation shows the same pattern, for example 0.15% at nominal volume and 9.0% at 1% for the 50 mL tip.
The practical reading, which follows from the numbers rather than from any single instruction: the same instrument is least precise when a large tip is used for a very small portion, so the tip size should be close to the volume actually being dispensed. For the general idea of volume ranges on single-draw pipettes, see Pipette Tips: Selection Guide by Application.
How to choose
These criteria come directly from the manufacturers’ documents. Neither document ranks products against each other, so none of this is a brand recommendation.
- Series of identical volumes. BRAND calls repetitive pipettes the first choice for dispensing long series efficiently. Eppendorf names tube and plate filling and reagent aliquoting.
- Viscous, volatile or foaming liquids. Positive displacement removes the air cushion that makes these liquids hard to dispense. Eppendorf’s ViscoTip is rated up to 14,000 mPa·s, and BRAND lets users slow aspiration for viscous or foaming media.
- Hazardous liquids. Eppendorf states that the sealed tip system suits infectious, toxic or radioactive liquids.
- Tip system. The Repeater M4 is not an open system and takes only Eppendorf tips. The BRAND brochure says third-party tips can be used on the HandyStep touch. Check this before building a workflow around one tip supplier.
- Purity. Pick the tip grade from the contamination the work needs to avoid, using the table above.
- Autoclaving. The Repeater M4 instrument is not autoclavable, according to its specification table.
Related reading: digital pipettes cover the electronic design in more detail, and bottle-top dispensers are the bench-top alternative when a reagent is dispensed straight from its container.
Pre-dispense and post-dispense in electronic repeat mode
Electronic pipettes with a repeat-dispense mode raise a point that mechanical repeater instructions do not stress in the same words. INTEGRA Biosciences, in a white paper hosted on News-Medical (sponsored content, dated 15 November 2022), explains that after aspiration the mechanism must change direction from aspirating to dispensing, which can make the first dispense too low in volume. It advises discarding the first dispense, typically released back into the source vessel; it calls this the pre-dispense and says it is usually 3-5% of the pipette’s total volume.
The same document says all later dispenses can be used but that errors may accumulate, because no pipette is 100% accurate. The liquid left in the tip for the final dispense may not be the amount the user expects, so the final dispense is also discarded; this waste is the post-dispense, also typically 3-5% of the pipette’s total volume. INTEGRA states that overall results are “significantly improved” with both discarded. It is a manufacturer statement about its own electronic pipettes, not a rule that applies equally to every repeater, so check the instructions for the instrument in use. It does explain why Eppendorf’s mechanical sequence includes a reverse stroke before dispensing starts.
Safe use and handling basics
- Fit a compatible tip while pushing the operating lever. On the Repeater M4 the tip size is then recognised, and the user selects the volume and the number of steps. The volume can be changed between steps.
- Eppendorf’s described sequence is to immerse the tip, push the lever to aspirate the total volume, push the lever once more to trigger a reverse stroke “to maximise accuracy”, and then dispense by pushing the lever slowly as far as it goes.
- Used tips and instruments may be biohazardous, chemical or radiological waste. Eppendorf advises separate critical and non-critical waste containers following local regulations.
- The instrument may be contaminated and contains electronics and a battery. For disposal, Eppendorf says to contact the biosafety officer or waste management.
Common mistakes
Each point below follows from a statement in the manufacturer documents above.
- Using a tip much larger than the portion. Error limits rise steeply at 1% of the tip volume (see the data table).
- Skipping the reverse stroke step. Eppendorf includes it in the sequence “to maximise accuracy”.
- Assuming any tip fits. The Repeater M4 accepts only its own tips; the HandyStep touch recognises PD-Tips II by piston coding.
- Autoclaving the instrument. The Repeater M4 specification says it is not autoclavable.
- Forgetting the adapter on the largest BRAND tips. The 25 and 50 mL tips need an autoclavable polypropylene adapter.
- Using the first and last dispense in electronic repeat mode. INTEGRA advises discarding both.
- Mixing tip waste. Contaminated tips go in the waste stream that fits what they held.
Sources
- Eppendorf, Repeater M4 product page, read 2026-11-19.
- Eppendorf, Combitips advanced product page, read 2026-11-19.
- BRAND GMBH + CO KG, HandyStep touch brochure (English), read 2026-11-19.
- INTEGRA Biosciences, How to use the repeat dispense mode (sponsored white paper on News-Medical, 15 November 2022), read 2026-11-19.
Data: BRAND PD-Tips II (repetitive pipette tips): volume range, accuracy and coefficient of variation at nominal volume and at 1% of nominal
Compiled from the accuracy table in the BRAND HandyStep touch brochure (distilled water, ISO 8655 tolerances not exceeded). Adds a calculated column: the permitted systematic error at 1% of nominal volume divided by the error at 100%.
| Tip | Nominal volume (mL) | Minimum set volume (uL) | Maximum set volume (uL) | Systematic error limit at 100% of nominal (%) | Systematic error limit at 1% of nominal (%) | Ratio of error limit at 1% to 100% (x) | CV limit at 100% of nominal (%) | CV limit at 1% of nominal (%) |
|---|---|---|---|---|---|---|---|---|
| PD-Tip II 0.1 mL | 0.1 | 1 | 100 | 1 | 8 | 8 | 0.5 | 12 |
| PD-Tip II 0.5 mL | 0.5 | 5 | 500 | 0.9 | 5 | 5.6 | 0.3 | 5 |
| PD-Tip II 1 mL | 1 | 10 | 1000 | 0.6 | 5 | 8.3 | 0.3 | 4 |
| PD-Tip II 1.25 mL | 1.25 | 12 | 1250 | 0.6 | 5 | 8.3 | 0.15 | 4 |
| PD-Tip II 2.5 mL | 2.5 | 25 | 2500 | 0.5 | 3.5 | 7 | 0.15 | 3 |
| PD-Tip II 5 mL | 5 | 50 | 5000 | 0.5 | 3.5 | 7 | 0.15 | 3 |
| PD-Tip II 10 mL | 10 | 100 | 10000 | 0.4 | 3.5 | 8.8 | 0.15 | 4 |
| PD-Tip II 12.5 mL | 12.5 | 125 | 12500 | 0.5 | 3.5 | 7 | 0.15 | 6.5 |
| PD-Tip II 25 mL | 25 | 250 | 25000 | 0.5 | 3 | 6 | 0.15 | 6 |
| PD-Tip II 50 mL | 50 | 500 | 50000 | 0.5 | 3 | 6 | 0.15 | 9 |
Sources
- BRAND HandyStep touch brochure (English), accuracy table (accessed 2026-11-19)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.







