Which water an autoclave needs depends on the model: three manufacturer manuals disagree on distilled, deionized and softened water.

Written and maintained by CASRAI Editorial Board

Last updated

This page describes products and general use; it is not medical advice.

Short answer: there is no single rule for autoclave water. Some steam sterilizers require steam-process distilled water, some accept distilled or deionized water, and some laboratory autoclaves say distilled water is unsuitable unless the unit has been adapted. The water your model needs is the one its operator’s manual names. The manufacturer manuals for reservoir-fed units exclude tap water.

Who this is for: clinic staff, lab technicians and buyers who fill, or specify the water supply for, a steam sterilizer and need to know what “autoclave water” means for their unit. After reading, you should be able to find the right water statement in your manual and recognise why two manufacturers can give opposite answers.

What autoclave water has to do

An autoclave turns water into steam, and the water also touches heaters, valves, pipework and level sensors. Dissolved minerals can leave scale on heaters and affect level sensors and valves, according to the laboratory-autoclave maker Priorclave. Some units also use the water’s electrical conductivity to detect whether water is present, so water can be wrong in two directions: too many dissolved solids, or too few. For the general design of the machine, see what an autoclave is; for feedwater as part of a site installation, see autoclave installation requirements.

Distilled, deionized, softened and tap water: what three manuals say

The table compares the water statements in three manufacturer documents. They disagree, which is the main point: “use distilled water” is a model-specific instruction, not a universal one.

Document (type of unit) What it says about water Key figures stated
SciCan STATIM 2000/5000 G4 operator’s manual (cassette table-top sterilizer for health care) Use only steam-process distilled water. Deionized, demineralized or specially filtered water should not be used. Never use tap water. Less than 5 ppm total dissolved solids (conductivity below 10 µS/cm). Other impurities or additives will cause an error reading on the display.
SciCan STATCLAVE G4 operator’s manual (table-top steam sterilizer for health care providers) High quality distilled water is recommended. Deionized, demineralized or specially filtered water can also be used. Never use tap water. For the external tank with automatic fill: conductivity of 6.4 ppm / 10 µS/cm or less recommended.
Priorclave “Autoclave Water Quality and Management” (UK laboratory research-grade autoclaves) Distilled or ultra-pure water is unsuitable unless the autoclave has been specially adapted. A sodium ion exchange softener is recommended for hard water. Total hardness as CaCO3 below 50 mg/L; conductivity above 100 µS.

The two SciCan models come from the same manufacturer yet differ on deionized water: the STATCLAVE manual allows it and the STATIM manual forbids it. Always follow the manual for your specific model, not the one for a sibling product.

Why conductivity limits cut both ways

The STATIM manual sets a ceiling: water above its limit, or containing additives, produces an error. Priorclave describes the opposite problem for its laboratory units. Extremely pure water prematurely corrodes copper pipework and brass valves and fittings, and the heater-protection and water-level controls use the water’s conductivity to detect it. That is why Priorclave recommends softened water that still conducts, rather than distilled water, for those machines. Both statements are true for their own machines, because the machines are designed differently: a cassette unit fed from a reservoir versus a laboratory unit whose controls sense the water’s conductivity.

Check the water-quality statement in your own model’s documentation before choosing a water supply.

Is “autoclave water” different from distilled water?

“Autoclave water” is not a separate product category. It is whatever water the manufacturer specifies for that machine: steam-process distilled for the STATIM, high quality distilled (with deionized acceptable) for the STATCLAVE, softened water within stated hardness and conductivity for the Priorclave units. For how laboratory water grades are defined, see Type 1 laboratory water grades.

Which water for which kind of sterilizer

Sterilizer type in the source Source’s water instruction
Cassette table-top (STATIM) Steam-process distilled only; no deionized, demineralized or filtered; no tap.
Table-top with reservoirs (STATCLAVE) Distilled recommended; deionized or demineralized acceptable; no tap.
Laboratory research-grade autoclave (Priorclave) Softened mains water within stated limits; distilled unsuitable unless adapted.

For how steam-generator and house-steam designs differ in their water supply, see steam-generator vs house-steam autoclaves.

Reservoir fill and chamber water basics

  • Fill reservoirs only with the water the manual names. The STATCLAVE quick-start instructs filling both the clean-water and Venturi reservoirs with high quality distilled water.
  • Chamber water becomes more concentrated over repeated cycles. Priorclave advises draining, flushing and cleaning the chamber at least monthly, ideally weekly, and draining spillage immediately.
  • Softened water raises sodium content over time (Priorclave), which is one reason regular flushing is recommended.
  • Where the manual gives a troubleshooting entry for wrong water (the STATIM manual does for water that is not steam-process distilled, or improperly distilled), follow its empty-and-refill instruction.

For routine upkeep see the autoclave preventive maintenance checklist and autoclave troubleshooting.

Common mistakes

  • Using tap water in a reservoir-fed unit. Both SciCan manuals say never to.
  • Assuming deionized equals distilled. One SciCan model allows it and another forbids it.
  • Assuming purer is always better. Priorclave says very pure water can corrode copper and brass and defeat conductivity-based level sensing on its units.
  • Applying another manufacturer’s rule to your machine instead of reading your own manual.

FAQ

Can I use tap water in an autoclave?

Not in the reservoir-fed SciCan units: both manuals say never to use tap water. Priorclave’s laboratory units use mains water treated by a softener to stay below 50 mg/L hardness as CaCO3, so untreated hard tap water is not what that manufacturer recommends either.

Can I use deionized water instead of distilled?

It depends on the model. The STATCLAVE manual says deionized or demineralized water can also be used; the STATIM manual says it should not be.

Do I have to use distilled water in every autoclave?

No. Priorclave states distilled water is unsuitable for its laboratory research-grade autoclaves unless specially adapted. Check your model’s manual.

What is the difference between autoclave water and distilled water?

There is no separate “autoclave water” definition; the term means the water your manufacturer specifies for that machine, which may be distilled, deionized or softened.

Sources

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Autoclave Distilled Water: What the Manuals Require

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →

Ask CASRAI · Regulatory Radar

Research-admin question? Get an answer that links its sources.

An AI assistant specialized in research administration. Every answer links its sources to check before you act. 2 questions free, no account. $29/month after.

  • Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.
  • Every answer numbers its sources and links each one, so you can check the source yourself.