Written and maintained by CASRAI Editorial Board
Published
On this page
The Eppendorf Research plus is a family of mechanical, air-cushion pipettes from Eppendorf for dispensing liquid in the microlitre to millilitre range. According to Eppendorf’s product page it comes as single-channel (fixed or variable volume), 8-, 12-, 16- and 24-channel versions and an adjustable tip-spacing version called Move It, and the whole pipette can be autoclaved. Its operating manuals cover volumes from 0.1 µL to 10 mL, each range paired with matching tips and a colour-coded control button.
Definition and scope
“Research plus” is a product line name, not a generic pipette type. The pipettes work on the air-cushion principle: a piston moves an air column that draws liquid into a disposable tip, and the operator sets the volume mechanically on a dial. Eppendorf lists spring-loaded tip cones (except on the three millilitre ranges), a temporary adjustment feature for liquids that differ from water, and an accessory range that includes a Pipette Carousel 2 holding up to six pipettes and a locking ring that deactivates the spring-loaded cone (Eppendorf product page, checked 2026-11-17).
This page describes the product line and the practice described in its manuals. It is a research-practice reference, not a purchasing guide: it contains no prices, availability or rankings. For other pipette topics on this site see the guides on the P10 pipette, reverse pipetting and filtered pipette tips.
Volume ranges and colour codes
Each single-channel variable-volume range has its own colour of control button, and the manual states that the button colour matches the colour of the corresponding epT.I.P.S. tip tray, which is how a user pairs a pipette with its tips. The data table below lists the ranges in the Research plus operating manual (2010 edition), with the tip range each is intended to use and the pipette’s tip length in the manual’s technical data. The current Eppendorf product page lists the same range structure, for example 0.5–10 µL (medium gray), 2–20 µL (light gray), 100–1,000 µL (blue) and 1–10 mL (turquoise), so check the label on your own pipette against the current documentation because colours have been revised in places: Eppendorf’s product-page FAQ states it changed the colour code of all 20 µL pipettes and its epT.I.P.S. 384 tips to light pink from pearl white.
Which version for which vessel
Eppendorf’s product page gives a vessel-based starting point, summarised here:
| Vessel | Pipette version Eppendorf lists |
|---|---|
| Tubes | Single-channel pipettes, fixed or variable volume |
| 96-well and 24-well plates | 8- and 12-channel pipettes |
| 384-well plates | 16- and 24-channel pipettes (used with epT.I.P.S. 384 tips) |
| Changing between vessel formats | Move It adjustable tip-spacing variant |
The same page says fixed-volume single-channel models (a 12-variant range) can be faster to use and less error-prone than variable models, and that variable pipettes become “fixed-volume” in effect when adjusted for a specific liquid, which is why Eppendorf’s FAQ describes adjustment as valid only for that liquid and volume.
How it is used in practice
The following points come from the operating manual (2010 edition; the 2016 edition agrees on immersion depth, waiting time, tip use and the autoclave conditions). They describe handling of the instrument and its tips only.
- Immersion and timing. Immerse the tip vertically about 4 mm into the liquid, let the control button return slowly, and for large volumes wait about 3 seconds before withdrawing the tip.
- Pre-wetting. The manual recommends wetting each new tip by aspirating and dispensing the liquid one to three times for maximum precision and accuracy.
- Tips. Use each tip once. The ep Dualfilter T.I.P.S. are not to be autoclaved.
- Different liquids and conditions. Liquids that differ strongly from water, and temperature differences between pipette, tip and liquid, change the delivered volume. At 1,000 m above sea level the manual puts the volume error of a 100 µL pipette at about −0.3 %, and says the pipette must be calibrated to ambient air pressure at extreme altitudes.
- Reverse pipetting. Eppendorf’s product page lists forward and reverse pipetting as supported techniques; see reverse pipetting for the method.
Tip cone, locking ring and temporary adjustment
Three features recur in Eppendorf’s documents and explain most of what is distinctive about the line.
Spring-loaded tip cone and locking ring
The spring-loaded tip cone gives a little when a tip is pressed on. Eppendorf’s FAQ says the delivery includes a locking ring that deactivates it, which is useful for tips not designed for Eppendorf cones; fitting it involves separating the upper and lower parts, and the manual carries the illustrated steps. The FAQ states the ring is made of polyphenylene sulfide and can be autoclaved under the same conditions as the pipette (121 °C, 1 bar, 20 min).
Temporary adjustment
Pipettes leave the factory calibrated to distilled water. Eppendorf says its mechanical pipettes allow a temporary user adjustment for liquids that differ from water in density, viscosity, surface tension or vapour pressure, for capillary effects such as DMSO, for altitude, and for non-standard tips such as epT.I.P.S. long. The adjusted pipette delivers the correct volume only for that liquid, and a variable pipette becomes in effect a fixed-volume pipette at one volume while adjusted. The Research plus has an adjustment window showing the change, and the setting can be returned to the factory “0”. Detailed procedures are in the operating manual and Application Note 198.
Warm, cold and UV-treated pipettes
For warm or cold liquids Eppendorf advises that pipette, tips, liquid and surrounding air should be at the same temperature, because a temperature difference affects the air cushion and can push errors beyond permissible limits; for challenging liquids in general it recommends direct-displacement devices. On decontamination, the FAQ says the pipette may be temporarily exposed to UV light under a sterile bench, that Eppendorf tested 165 hours of intense UV with no visible discoloration, and that it recommends 30 W at 254 nm at a distance of about 60 cm.
Autoclaving and cleaning
The manuals specify one autoclave procedure: 121 °C, 20 minutes, 1 bar overpressure, following the autoclave maker’s own manual and making sure the pipette itself does not exceed 121 °C. The Research plus may be autoclaved as a whole unit or with the lower part removed, and the lower part is not to be disassembled. Eppendorf’s FAQ states the pipettes were tested over 175 autoclaving cycles under those conditions, which it equates to an average of about five years of use. The manual adds that the piston does not need relubricating afterwards and that, if autoclaving follows the described procedure, no gravimetric test or adjustment is needed.
Cleaning limits stated in the manual: do not use acetone, do not use sharp objects, and do not use additional disinfectants, decontamination agents or sodium hypochlorite during autoclaving or UV irradiation. Check the chemical-resistance information Eppendorf supplies before using aggressive liquids.
Where it matters for labs
Calibration and routine checks
Eppendorf’s FAQ cites EN ISO 8655 Part 1 (2022 edition) for calibration “regularly and at a minimum, once per year”, with routine tests at shorter intervals, and says both frequencies should be set by a risk assessment that considers how often the pipette is used, the number of users, how aggressive the liquids are and the acceptable error limits, together with any industry- or organisation-specific requirement. Calibration checks use distilled or deionised water on a balance; for other liquids Eppendorf points to its temporary adjustment (Application Note 198) rather than calibrating with the liquid itself.
Storage and sustainability labelling
Eppendorf recommends keeping pipettes on a carousel rather than in a drawer. A trade-press report dated 17 December 2021 states that the single-channel variable Research plus line-up was certified with an ACT label by the non-profit My Green Lab, an environmental-impact label covering manufacturing, energy and water use, packaging and end-of-life; the product page repeats this for the single-channel variable line-up. The label applies to that line-up, not to every Research plus version.
Common misunderstandings
- “Autoclaving means I must recalibrate.” The manual says no gravimetric test or adjustment is needed after autoclaving as described. A gravimetric test is part of reassembly after servicing, to confirm no parts were swapped.
- “Fully autoclavable covers the tips.” The pipette is autoclavable; the ep Dualfilter T.I.P.S. are not.
- “A pipette’s nominal volume is its mid-range.” Eppendorf states that the nominal volume of a variable pipette is its maximum volume, for example 1,000 µL for a 100–1,000 µL pipette.
- “Colour codes never change.” They are tied to a given edition and tip range, as the 20 µL change above shows.
- “One adjustment fits all liquids.” A temporary adjustment is valid for the liquid and volume it was made for.
Sources
- Eppendorf: Eppendorf Research plus product page and FAQ, checked 2026-11-17
- Eppendorf: Research plus Operating manual (2010 edition, PDF copy), checked 2026-11-17
- Eppendorf: Research plus Operating manual (2016 edition, PDF copy), checked 2026-11-17
- BioSpectrum Asia: Eppendorf Research plus is the first pipette to receive My Green Lab ACT label, 17 December 2021, checked 2026-11-17
Data: Eppendorf Research plus single-channel variable pipettes: range, button colour, tip range and tip length
Compiled from the technical-data table of the Eppendorf Research plus operating manual (2010 edition), cross-read with the current Eppendorf product page. Lists each single-channel variable-volume range with its control-button colour, the tip range specified for it and the pipette tip length in millimetres.
Period covered: 2010 manual edition.
| Volume range | Control button colour | Specified tip range | Tip length (mm) |
|---|---|---|---|
| 0.1–2.5 µL | dark gray | 0.1–10 µL | 34 |
| 0.5–10 µL | medium gray | 0.1–20 µL | 40 |
| 2–20 µL | light gray | 0.5–20 µL (L) | 46 |
| 10–100 µL | yellow | 2–200 µL | 53 |
| 20–200 µL | yellow | 2–200 µL | 53 |
| 30–300 µL | orange | 20–300 µL | 55 |
| 100–1,000 µL | blue | 50–1,000 µL | 71 |
| 0.5–5 mL | purple | 0.1–5 mL | 120 |
| 1–10 mL | turquoise | 1–10 mL | 165 |
Sources
- Eppendorf Research plus Operating manual (2010 edition) (accessed 2026-11-17)
- Eppendorf Research plus product page (accessed 2026-11-17)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.