Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product and regulatory information only; this is not medical advice and not instructions for using any surgical instrument.
Short answer
“Greenberg retractor” is a brand name for a self-retaining retractor system for neurosurgery. Under US rules the device type sits in FDA product code GZT, “retractor, self-retaining, for neurosurgery”, regulated at 21 CFR 882.4800 as Class 2. A catalogue entry for a Greenberg product tells you the maker’s model and blade, while the regulatory category comes from the generic device type.
Definition and scope
The federal regulation defines the category in one sentence: a self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery. The same section classifies it as Class II (performance standards) (21 CFR 882.4800, checked 2026-10-31). FDA’s product classification database lists code GZT under that regulation with device class 2, submission type 510(k), and eligibility for the 510(k) Third Party Review Program (FDA Product Classification, GZT, checked 2026-10-31).
“Greenberg” is therefore a product name inside that category, not a separate regulatory class. In the NIH National Library of Medicine’s AccessGUDID database, Aspen Surgical Products, Inc. lists a Symmetry GREENBERG record (model 50-1531) under product code GZT, described as “Greenberg Retractor Tapered Blade Width 5/32 in x 5/8 in (4 mm x 16 mm); 9 in; 229 mm” (AccessGUDID, checked 2026-10-31).
What a listing records, and what it does not
The public device records describe the product as an item, not a technique. Two Greenberg-branded records in AccessGUDID show the same pattern for the fields below (checked 2026-10-31). Procedure steps are outside the scope of this page; the surgical team’s training and the manufacturer’s instructions for use govern how the system is assembled and used.
| Record field | What the record shows | Source |
|---|---|---|
| Brand and model | Symmetry GREENBERG, model 50-1531; Symmetry Greenberg Simplicity, catalogue 50-1519 | AccessGUDID |
| Product code | GZT for both records | AccessGUDID |
| Description string | Model 50-1531: tapered blade, width 5/32 in x 5/8 in (4 mm x 16 mm), 9 in (229 mm) | AccessGUDID |
| Sterile as packaged | No; sterilization is required before use, by moist heat (steam) | AccessGUDID |
| Single use | No (reusable) | AccessGUDID |
| MRI safety | Labeling does not contain MRI safety information (model 50-1531 record) | AccessGUDID |
Greenberg and related retractors
Greenberg is one of several named retractor families. For the generic device group see what a surgical retractor is; for the instrument set it is used alongside, see neurosurgical instruments. Other named designs have their own guides: the cerebellar retractor and the Bookwalter retractor. Parts that hold tissue are covered in retractor blades.
Where it matters for lab, clinic and supply decisions
A checklist of questions drawn from the fields above. Each question is answered by the maker’s own documentation, not by this page.
- Which model is it? Greenberg-branded records carry different model numbers (50-1531 and 50-1519 above), so a brand name alone does not identify a part.
- What does the description string say? Blade type, width, and length are encoded in text such as “Tapered Blade Width 5/32 in x 5/8 in (4 mm x 16 mm); 9 in; 229 mm”. Compare the string, not the brand.
- Is it sterile when it arrives? The records say no, so the facility’s reprocessing workflow applies. “Non-sterile as packaged” is a labeling fact, not a defect.
- Is it reusable? The records mark single use as No, which means reprocessing instructions in the manufacturer’s labeling matter.
- Does the labeling address MRI? The 50-1531 record says the labeling contains no MRI safety information. Ask the maker rather than assuming.
A discrepancy worth knowing about. The record for the Simplicity retractor (catalogue 50-1519) is described as a “Retractor Kit”, yet the same record does not describe its components, and its own kit indicator does not agree with the description. A database record is not a parts list; confirm contents against the manufacturer’s documentation.
Common misunderstandings
- “Greenberg is a class of device.” It is a brand. The class comes from the generic type, GZT under 21 CFR 882.4800.
- “Class 2 means exempt from review.” The FDA classification page shows submission type 510(k) for GZT; it does not state an exemption.
- “Non-sterile means unsafe.” The records show the item is supplied non-sterile and intended to be sterilized before use.
- “A database record equals instructions for use.” GUDID records identify a device; they are not a substitute for the maker’s labeling.
FAQ
What is a Greenberg retractor?
A Greenberg retractor is a branded self-retaining retractor system for neurosurgery. The US device type is product code GZT.
What FDA class is a self-retaining neurosurgical retractor?
Class II under 21 CFR 882.4800; FDA’s classification database lists class 2 for code GZT.
Is a Greenberg retractor sterile when supplied?
The AccessGUDID records for the Greenberg 50-1531 and Greenberg Simplicity 50-1519 say it is not sterile as packaged and needs sterilization before use.
Is it single use?
No. Both records mark single use as No.
Sources
- eCFR, 21 CFR 882.4800, Self-retaining retractor for neurosurgery, checked 2026-10-31.
- FDA, Product Classification, code GZT, checked 2026-10-31.
- NLM, AccessGUDID record, Symmetry GREENBERG 50-1531, checked 2026-10-31.
- NLM, AccessGUDID record, Symmetry Greenberg Simplicity 50-1519, checked 2026-10-31.