Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General product information only, not medical advice and not a surgical technique guide.
Short answer
A Meyerding retractor is a self-retaining laminectomy retractor: a ratcheting, scissor-style instrument whose blades hold the edges of an incision apart. In FDA data, listings for it can carry product code GZT, a Class II neurosurgery code under 21 CFR 882.4800, which is not the same class as a general hand retractor.
Definition and scope
The FDA regulation for self-retaining retractors for neurosurgery, 21 CFR 882.4800, describes the device as a self-locking device used to hold the edges of a wound open during neurosurgery, and classifies it as Class II (performance standards). The FDA product code GZT is named “Retractor, Self-Retaining, For Neurosurgery” under that regulation, reviewed by the Neurology panel, and is not an implant (openFDA, data last updated 2026-10-19).
In AccessGUDID, a Markham-Meyerding self-retaining hemilaminectomy retractor from Ambler Surgical (DI 00190660139117) is described as a 6 3/4 inch instrument with a 2 1/2 inch by 1 inch wide blade on the left side, a single hook on the right side, and a ring handle with ratchet. Its GMDN term is “Laminectomy retractor, self-retaining”, described as a self-locking instrument that draws aside the margins of an incision so the lamina of the vertebral posterior arch can be excised. This page does not describe how the instrument is used in surgery.
How it is described in practice
Catalogue and regulatory records describe the instrument in a consistent set of attributes. Reading a listing means checking each one:
- Product code and class. The Ambler GUDID record lists product code GZT, which openFDA maps to Class 2.
- Reusable, supplied non-sterile. The same record states the device is not sterile, requires sterilization before use (moist heat or steam), and is not single-use.
- Overall length and blade size. A Teleflex listing for Meyerding laminectomy retractors (item KM55430) shows lengths of 6 1/2 in (16.5 cm), 9 1/2 in (24.0 cm) and 10 in (25.5 cm), blade depth by width of 57.0 x 25.0, 51.0 x 25.0 and 41.0 x 25.0 mm, and a sharp distal style. The listing does not state which blade size goes with which length, so confirm the pairing on the specific item.
- Symmetric or one-sided. The Markham-Meyerding record above has a wide blade on one side and a single hook on the other.
- MRI labeling. The Ambler record carries no MRI safety information.
Types and related names
Names for the same family of instrument vary between records. The table lists the names that appear in the cited records.
| Name in the record | Where it appears | What it shows |
|---|---|---|
| Meyerding laminectomy retractors | Teleflex listing KM55430 | Three lengths, three blade sizes, sharp distal style |
| Markham-Meyerding self-retaining hemilaminectomy retractor | AccessGUDID, Ambler Surgical | Wide blade left, single hook right, ring handle with ratchet |
| Laminectomy retractor, self-retaining | GMDN term on the GUDID record | Generic device-nomenclature name; the term’s status is listed as obsolete |
Where it matters for labs, clinics and supply decisions
The name “retractor” does not determine regulatory class. The same word appears under two different FDA product codes:
| Product code | FDA device name | Regulation | Class |
|---|---|---|---|
| GZT | Retractor, Self-Retaining, For Neurosurgery | 21 CFR 882.4800 | Class 2 |
| GAD | Retractor | 21 CFR 878.4800 (general manual instruments) | Class 1 |
Both rows are from openFDA classification data checked on 2026-10-30. So when comparing records for a retractor, compare the product code and the labeled intended use, not the instrument name. Related hand-held retractors on this site, such as the Richardson retractor and the Weitlaner retractor, are covered in their own guides; see also the overview of surgical retractors, retractor blades and neurosurgical instruments.
Common misunderstandings
- “It arrives sterile and is single-use.” The Ambler GUDID record says the opposite: reusable and non-sterile, with sterilization required before use.
- “All retractors are Class I.” The general Retractor code GAD is Class 1, but the self-retaining neurosurgery code GZT is Class 2.
- “Meyerding” means one fixed size. Records show several lengths and blade sizes under the name.
FAQ
Is a Meyerding retractor Class 1 or Class 2?
A record that carries product code GZT is Class 2 under 21 CFR 882.4800. The Ambler Markham-Meyerding record lists GZT. A retractor listed under code GAD is Class 1 instead, so check the code on the specific product.
Is a Meyerding retractor sterile?
The AccessGUDID record for the Ambler Markham-Meyerding retractor says it is not sterile, is reusable, and needs steam sterilization before use. Follow the manufacturer’s instructions for use for the product in hand.
What sizes are listed?
The Teleflex listing KM55430 shows 6 1/2, 9 1/2 and 10 inch lengths with blade depth by width of 57.0 x 25.0, 51.0 x 25.0 and 41.0 x 25.0 mm; the Ambler record is 6 3/4 inches.
Sources
- eCFR, 21 CFR 882.4800, Self-retaining retractor for neurosurgery, https://www.ecfr.gov/current/title-21/section-882.4800 (checked 2026-10-30).
- openFDA device classification, product codes GZT and GAD, https://api.fda.gov/device/classification.json (data last updated 2026-10-19; checked 2026-10-30).
- NLM AccessGUDID, DI 00190660139117, https://accessgudid.nlm.nih.gov/devices/00190660139117 (checked 2026-10-30).
- Teleflex, Meyerding Laminectomy Retractors KM55430, https://us.myteleflex.com/en/USD/All-Categories/Surgical-Instruments/Orthopedic/Meyerding-Laminectomy-Retractors/p/KM55430 (manufacturer listing, checked 2026-10-30).