Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General information only; this is not medical advice and not a procedure guide.
Short answer
A Ragnell retractor is a double-ended, hand-held surgical retractor. In the United States it is listed under FDA product code GAD (Retractor), a Class 1 device under 21 CFR 878.4800 that is exempt from 510(k) premarket notification, and the two AccessGUDID records checked for this guide describe it as reusable, not packaged sterile, and requiring sterilization before use.
Definition and scope
“Ragnell” is a pattern name, not a regulatory category. Regulators and device databases classify the underlying type instead. Two AccessGUDID records, one from Sontec Instruments (device 00192896025417) and one from STERIS under the V. Mueller brand (device 10885403086885), both carry the Global Medical Device Nomenclature (GMDN) term “Hand-held surgical retractor, reusable” (code 45918).
For the general idea of retractors and how designs differ, see what a surgical retractor is.
How listings describe the Ragnell retractor
The two records name the same pattern in different words, which is the main reason a Ragnell listing needs reading closely rather than matching by name alone.
| Field | Sontec Instruments record | STERIS / V. Mueller record |
|---|---|---|
| Brand name | Ragnell Retractor | Affinity (V. Mueller), model OA930 |
| Description as listed | “Ragnell Retractor Double Ended Blue Electro Coated” | “Affinity Ragnell Retractor Delicate, Double-Ended. Overall length 6 inches 15.2 cm.” |
| Overall length | 6 inches | 6 inches (15.2 cm) |
| FDA product code | GAD, Retractor | GAD, Retractor |
| Single-use | No | Reusable |
| Packaged sterile | No | No |
| Requires sterilization before use | Yes | Yes; moist heat or steam sterilization listed |
| Blade sizes in the record | Not given beyond overall length | “No Device Sizes” under clinically relevant size |
Both columns are registered listings, checked in AccessGUDID on 2026-10-30. Other manufacturers’ Ragnell listings can differ, so the record for the exact catalogue number is the one to read.
FDA classification in one place
The FDA product classification database lists product code GAD with the device name “Retractor,” Class 1, regulation number 878.4800, 510(k) exempt, and the definition “Manual surgical instrument for general use.” Exempt does not mean unregulated: manufacturers still register with FDA, and the exemption has limits that the manufacturer must confirm under 21 CFR Parts 862 to 892 (FDA classification page checked 2026-10-30).
Where it matters for labs, clinics and supply decisions
- Reprocessing is part of ownership. Both records state the instrument is reusable and not supplied sterile, so it enters a facility as an item that has to go through the facility’s own sterilization workflow before use. The STERIS record lists moist heat or steam as the method.
- Finish and variant are specification details. The Sontec record names a blue electro-coated finish; the STERIS record names a “delicate” version. Neither is a regulatory category, so they are confirmed on the manufacturer’s own listing.
- Blade width is not in the database. The STERIS record shows no device sizes, so blade dimensions come from the manufacturer’s catalogue, not from AccessGUDID.
Checklist for reading a Ragnell listing
- Confirm the exact catalogue or model number, not just the name “Ragnell.”
- Check the overall length against the two listings above (6 inches / 15.2 cm).
- Check whether the listing says standard, “delicate,” or double-ended, and what finish it names.
- Check that it is labelled reusable and note the sterilization instruction the manufacturer gives.
- Look up the device in AccessGUDID by its identifier to compare the product code (GAD) and sterilization fields.
Common misunderstandings
- “Class 1 means no regulation.” Class 1 devices are still subject to FDA general controls and registration.
- “It arrives sterile.” Both records say the opposite: not packaged sterile, sterilization required before use.
- “All Ragnell retractors are the same.” The two records differ in name, finish and the “delicate” designation.
- “The database confirms its surgical use.” The records give a general device description, not indications for a particular procedure.
Related retractors
For other named patterns see the Senn retractor, Richardson retractor, Army-Navy retractor, Deaver retractor and Weitlaner retractor guides.
FAQ
What is a Ragnell retractor?
A double-ended surgical retractor, listed in AccessGUDID as a reusable hand-held surgical retractor (GMDN 45918).
Is a Ragnell retractor FDA regulated?
Yes. It falls under product code GAD, 21 CFR 878.4800, Class 1, exempt from 510(k).
Is it sold sterile?
The two listings state it is not packaged sterile and requires sterilization before use.
How long is a Ragnell retractor?
Both listings give an overall length of 6 inches (15.2 cm).
Sources
- NLM AccessGUDID, device 00192896025417, Sontec Instruments, accessgudid.nlm.nih.gov/devices/00192896025417, checked 2026-10-30.
- NLM AccessGUDID, device 10885403086885, STERIS Corporation (V. Mueller), accessgudid.nlm.nih.gov/devices/10885403086885, checked 2026-10-30.
- FDA Product Classification, code GAD, accessdata.fda.gov classification GAD, checked 2026-10-30.