Written and maintained by CASRAI Editorial Board
Published
On this page
Incubator repair for a laboratory CO2 incubator splits into two jobs: the operator does cleaning, water, filter and log-keeping tasks on a schedule, and anything that involves internal adjustment or a sensor failure goes to qualified service personnel. Thermo Fisher’s operating manual for its 3010 series states that “All internal adjustments and maintenance must be performed by qualified service personnel.” This guide sets out the routine maintenance that keeps a lab incubator out of the repair queue, how to read an alarm, and where spare parts come from. It is about laboratory incubators, not egg incubators.
Incubator repair: what the operator does and what needs a technician
The three manufacturer and institutional sources used here draw the line in the same place. Operators clean, refill, inspect, replace consumable filters and record what they did. Technicians make internal adjustments, handle sensor faults and supply replacement parts. The overview of laboratory incubators explains the incubator types these rules apply to; the water-jacketed and direct-heat CO2 designs in the manuals below are the ones most labs run for mammalian cell culture.
| Task | Who does it, per the source | Source |
|---|---|---|
| Internal adjustments and maintenance | Qualified service personnel only | Thermo Scientific manual 7013010 Rev. 6 |
| Cleaning and decontaminating the interior, shelves, ducts and humidity pan | The operator, after powering off and unplugging the unit | Thermo manual; Esco maintenance guide |
| Checking the humidity pan water level and the gas tank level | The operator | Esco maintenance guide; Thermo manual preventive maintenance table |
| Replacing air sample, access port and gas filters | Listed as maintenance items in the manual’s preventive maintenance table, with a recommended 6-month interval for the air sample and access port filters | Thermo manual |
| Sensor failure shown on the display | Contact the local distributor or the manufacturer’s technical services department | Thermo manual |
| Calibration of temperature, CO2 and humidity sensors | Listed as a yearly maintenance item in Esco’s schedule; Thermo says to verify and document calibration, with intervals that depend on use, environment and accuracy required | Esco maintenance guide; Thermo manual |
For the calibration task itself, including temperature uniformity mapping, see the separate guide to CO2 incubator calibration and temperature mapping. Esco’s guide adds that a general inspection should happen at least annually and cover the tightness of the glass door seal and hinge screws, whether the door latch works, a functional check of the operating panel and controls, and an electrical safety check per local regulations.
Incubator maintenance checklist and preventive maintenance schedule
Three sources give three different schedules, and the differences matter more than any single number. Thermo’s preventive maintenance table for the 3010 model lists checking CO2 tank levels, inspecting the door latch, hinges and door gasket seal, checking the humidity pan water level, verifying and documenting CO2 and temperature calibration, complete decontamination “between experiments”, changing HEPA and gas filters “or as needed”, and inspecting all tubing and connections. The manual also says that “Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.” Esco states fixed intervals, and the University of Utah Institutional Biosafety Committee states ranges tied to how many people use the unit. The data table below sets the three side by side so you can see where they agree and where they diverge.
Daily and weekly tasks
Esco’s schedule has one daily task: check the CO2 or N2 gas tank level, with the regulator gauge not below 15 psig. Weekly, it says to check the water level in the humidity pan, refilling with sterile distilled water once a week, and to clean the interior and exterior with a general-purpose laboratory cleaning agent. Utah’s page is aimed at shared tissue-culture rooms: check the incubator once a week and discard unused cultures, clean spills immediately, clean the top of the unit every two weeks, and clean the incubator one to two times per month depending on the number of users.
Six-monthly, yearly and as-needed tasks
Thermo recommends replacing the air sample filter and the access port filter every 6 months and replenishing the water jacket rust inhibitor every 2 years on the 3010 model. Esco puts the yearly items together: general inspection, calibration of the temperature, CO2 and humidity sensors, a ULPA filter change on the 170 L and 240 L models (once a year or when visibly dirty), inlet filters changed at least once a year or when dirty, and oxygen sensor calibration yearly and as needed. Its as-needed items are replacing the outer door magnetic gasket (check for brittleness, corrosion, wear or damage), replacing the UV lamp (about 10,000 running hours, which Esco puts at around four years) and decontaminating the inner chamber.
Keep a maintenance and service log
Esco advises keeping a maintenance and service log, and Thermo’s table asks the operator to verify and document calibration. The wider practice of tracking assets, service dates and vendor contracts is covered in biomedical equipment maintenance.
How do I clean my incubator?
Switch the incubator off and unplug it, remove the shelves, ducts, filter tubing and humidity pan, clean the cabinet from the top down with a disinfectant that is compatible with stainless steel, rinse until no disinfectant remains, wipe with 70% ethanol, reassemble and let the unit run before returning cultures. This follows the Thermo 3010 manual and Esco’s procedure, which agree on the order of work and differ only in detail.
The cleaning sequence in the manuals
Thermo’s interior procedure: power off and unplug; remove the shelves, access port filter, side duct sheets, temperature sensors and air sample filter tubing; wash the shelves and ducts with disinfectant and rinse with sterile water (the shelves, ducts and wing nuts may be autoclaved); wash the cabinet from the top down and the inner door on both sides; rinse until the disinfectant is gone; spray 70% alcohol; reassemble. The manual then says: “Allow the unit to run empty for 24 hours before returning to service.” The exterior takes a damp sponge and mild detergent, and the humidity pan is washed with soap and water plus laboratory disinfectant, rinsed with sterile water, treated with 70% alcohol, and may be autoclaved.
Esco’s version starts earlier: wear the protective equipment your laboratory requires, move samples to another incubator, switch off, unplug and tag the unit as being serviced. After removing the shelves, plenum and humidity pan, clean with a mild soap or disinfectant at the contact time the disinfectant manufacturer specifies, rinse twice with distilled water, dry with a lint-free cloth, wipe with 70% ethanol and make sure everything is completely dry before use. It also notes using a stainless steel cleaner on the metal and a glass cleaner on the inner glass door, and running the unit’s decontamination or sterilisation cycle if it has one.
Cleaning agents: what to avoid
| Agent or action | What the sources say | Source |
|---|---|---|
| Bleach and chlorine-based disinfectants | Do not use sodium hypochlorite: it causes pitting and rusting. “Stainless steel is corrosion resistant, not corrosion proof.” | Thermo manual |
| Chlorine-based disinfectants, general | “Never use chlorine-based disinfectants.” | Esco maintenance guide |
| Chlorine cleaners, cell-culture view | They corrode stainless steel and copper and are toxic to cells. | University of Utah IBC |
| Chlorine on stainless steel when it cannot be avoided | Chlorine and chlorine-based compounds pit stainless steel in incubators, water baths and autoclaves. If one is used, follow it with a water wipe to remove residue, then 70% ethanol or 70% isopropanol. | University of Nebraska-Lincoln EHS safe operating procedure |
| Strong alkaline or caustic agents | Do not use them. | Thermo manual |
| Alcohol, including 70% | Volatile and flammable: use in a ventilated area free from open flame, and let it dry before powering the unit on. | Thermo manual |
| Spraying onto sensors or the control panel | Do not spray the temperature or CO2 sensor, the control panel or the area near the electrical panel; use a soaked cloth. The CO2 sensor runs at about 150 °C in operation, so let it cool first. | Thermo manual; Esco maintenance guide |
| Any cleaning method not recommended by the manufacturer | Check with the manufacturer first. | Thermo manual |
For interior surfaces Utah’s page names quaternary ammonium disinfectants and gives three product examples (Lysol No Rinse, Conflikt and Fermacidal-D), and it describes optional humidity-pan additives (Aquaguard-1, Aqua EZ Clean and SigmaClean).
Humidity pan water
Thermo specifies sterile, distilled or demineralised water in the pan, filled to within half an inch of the top. Esco adds never to use chlorinated tap water, because chlorine deteriorates stainless steel and tap water leaves scale. Utah specifies sterile distilled water with a pH between 7 and 9 and conductivity of 1 to 20 microsiemens per cm, changed every two weeks on its page, and keeps the pan full because low humidity raises evaporation from culture medium. Contamination that keeps recurring despite this routine is treated in cell culture contamination.
Incubator troubleshooting: reading the alarm
An alarm tells you which condition the control system detected, not why. The only symptom-to-meaning table in the sources is Thermo’s Table 3-1 for the 3010/3020 models, so the table below is specific to those models. Other makes and models use different thresholds and delays; read your own manual’s alarm table before relying on it.
| Alarm or display (Thermo 3010/3020) | Condition as stated in the manual |
|---|---|
| Over Temp | Temperature above the over-temperature setpoint; no delay; ringback 15 minutes |
| Low Temp | At least 1 °C below setpoint; 15-minute delay |
| High CO2 | At least 1% above setpoint; 15-minute delay |
| Low CO2 | At least 1% below setpoint; 15-minute delay |
| Add Water | Water jacket level low; no delay |
| Door Open | Inner glass door open; 15-minute delay |
| Temperature display flashing | Temperature sensor failure: contact the local distributor or Technical Services |
| CO2 display flashing | CO2 sensor failure: contact the local distributor or Technical Services |
The same manual says the alarms are momentary and clear automatically, that the low-temperature alarm delay is 9 hours after power-up, and that temperature alarms are disabled at a setpoint of 10.0 °C and CO2 alarms at 0.0%.
The sources do not list causes for each symptom, and this guide does not invent them. What they do give you is a short pre-call checklist drawn from the maintenance tasks above: gas tank level and regulator gauge (Esco), humidity pan water or jacket water level (Thermo), door latch, hinges and gasket seal (Thermo), and whether the door has been left open (Esco warns that an open door causes overheating, fire hazard and contamination). If none of those explain the alarm, or a sensor display is flashing, the manufacturer routes the problem to technical services.
What are the parts of an incubator?
Using the terms the manuals use for a CO2 incubator: the interior shelves; the side duct sheets and top duct; the blower wheel and scroll; the humidity pan; the inner glass door and the door gasket (Esco calls its outer gasket a magnetic gasket); the temperature and CO2 sensors; the air sample filter, the access port filter and the HEPA or ULPA filter; gas inlet filters and tubing; and, on units that have one, a UV lamp and a water jacket. The Thermo preventive maintenance table treats the blower wheel and scroll, the top duct and the side ducts as parts to be decontaminated, which is why they are removable. For how these parts differ between incubator types, see how to choose a CO2 incubator and, for a non-CO2 benchtop unit, what is a bacterial incubator.
Where can I find spare parts for my incubator machine?
Start with the manufacturer. The Thermo operating manual states that the manufacturer supplies spare and replacement parts, technical troubleshooting and on-site service, and its sensor-failure instruction is to contact the local distributor or Technical Services. Replacing a fuse is a good example of why you should use the manual’s parts list: the manual has a fuse table (Table 4-1) and says to replace a fuse with one of the same specification. For consumables with a fixed interval, such as the 6-monthly air sample and access port filters or Esco’s door gasket and UV lamp, order the part named in your model’s manual rather than a generic equivalent.
Before sending an incubator out for repair
If a technician needs to take the unit away, or work on it on site, it must be decontaminated first. The University of Nebraska-Lincoln EHS safe operating procedure covers incubators being relocated, decommissioned or sent for authorised repair. Its steps: remove biological and liquid materials; turn the unit off and unplug it; clean up any spills; apply an agent-specific disinfectant to the top, bottom and sides of the shelves (or autoclave shelves that tolerate 121 °C or higher); wipe all interior surfaces including the inside of the door and the gaskets at the disinfectant’s labelled contact time; wipe the exterior; remove biohazard stickers once decontamination is done. It then says not to use the incubator again after decontamination, to post a sign showing the date of decontamination, and to tape the door shut if necessary for transport. Equipment authorised for radioactive material has to be decontaminated by the radiation safety staff before the biological decontamination.
Common mistakes
- Cleaning with bleach. Thermo, Esco, Utah and Nebraska all warn against chlorine-based cleaners on stainless steel interiors.
- Spraying disinfectant at sensors or the control panel. Thermo and Esco both say to avoid it; use a soaked cloth.
- Powering on while the alcohol is still wet. Thermo says alcohol is flammable and the unit must be dry first.
- Returning cultures immediately after cleaning. Thermo’s manual calls for running the unit empty for 24 hours.
- Using tap water in the humidity pan. Esco says chlorinated tap water deteriorates stainless steel and leaves scale.
- Leaving the door open. Esco lists overheating, a fire hazard and contamination as the consequences.
- Letting items touch the side walls or placing the unit near an autoclave, radiator or oven, or somewhere the power plug is not reachable (Esco).
- Attempting internal adjustments. Thermo restricts these to qualified service personnel.
Safe use and handling basics
Disconnect the power cord before cleaning (Thermo). Esco adds that CO2 and N2 can suffocate in high concentration, that the gas tank should have a two-stage regulator, and that the incubator should be indoors, off the floor, in ventilated space but not in air currents, with unobstructed access to the power plug. Thermo recommends not using powdered gloves for tissue cultures. Wear the protective equipment your laboratory requires for the disinfectant in use, following its label and safety data sheet (Nebraska).
How to assemble an egg incubator?
This page is about laboratory incubators, so it does not cover assembling poultry or egg incubators. Those are consumer or agricultural products that ship with their own manufacturer instructions, which are the right source for assembly.
Frequently asked questions
How do I clean my incubator?
Power it off and unplug it, remove shelves, ducts, filters and the humidity pan, wash with a stainless-steel-compatible disinfectant, rinse, wipe with 70% ethanol and dry fully. Never use chlorine-based disinfectants, per Thermo and Esco. Thermo then calls for running the unit empty for 24 hours before it goes back into service.
Where can I find spare parts for my incubator machine?
From the manufacturer or its qualified service organisation. Thermo’s manual says the manufacturer supplies spare and replacement parts, technical troubleshooting and on-site service. Order the part that your model’s manual specifies.
What are the parts of an incubator?
For a CO2 incubator: shelves, ducts, blower wheel and scroll, humidity pan, inner glass door and gasket, temperature and CO2 sensors, filters, gas tubing, and on some models a UV lamp and water jacket.
How often should an incubator be serviced?
The sources do not give one interval. Esco lists a yearly inspection and calibration; Thermo ties cleaning and calibration intervals to use, environment and accuracy required; Utah gives monthly-to-quarterly CO2 calibration depending on door traffic. See the data table.
Can I repair an incubator myself?
Routine cleaning, water, filter replacement and inspection are operator tasks. Thermo states that all internal adjustments and maintenance must be performed by qualified service personnel.
How to assemble an egg incubator?
That is outside this guide, which covers laboratory incubators. Use the instructions supplied by the maker of the egg incubator.
Sources
- Thermo Scientific: Operating and Maintenance Manual 7013010 Rev. 6, Models 3010/3020 water jacketed CO2 incubators, checked 2026-11-21
- Esco Lifesciences: CO2 Incubator Maintenance Tips and Procedures, V.B Mar2021, checked 2026-11-21
- University of Utah Institutional Biosafety Committee: Best Practices for CO2 Incubator Maintenance, checked 2026-11-21
- University of Nebraska-Lincoln EHS: Biological Decontamination of Laboratory Equipment (Safe Operating Procedure), checked 2026-11-21
Data: CO2 incubator maintenance intervals compared across a manufacturer manual, a manufacturer guide and a biosafety office
Compares how three sources schedule routine CO2 incubator maintenance tasks: the Thermo Scientific 3010/3020 operating manual, Esco Lifesciences' maintenance guide and the University of Utah Institutional Biosafety Committee. Values are transcribed as each source states them; where a source gives no value the cell says so.
Period covered: 2021/2026.
| Task | Thermo Scientific manual 7013010 Rev. 6 | Esco Lifesciences guide V.B Mar2021 | University of Utah IBC |
|---|---|---|---|
| Humidity pan water | Check level; sterile, distilled or demineralised water, fill to within 1/2 inch of top | Refill with sterile distilled water once a week | Sterile distilled water, pH 7-9; change every two weeks; keep pan full |
| Interior cleaning | Complete decontamination between experiments; more often depending on use and conditions | Weekly interior and exterior cleaning; inner chamber decontamination as needed | Clean 1-2 times per month depending on number of users; top every 2 weeks |
| Air, HEPA or ULPA filters | Air sample and access port filters every 6 months recommended; HEPA and gas filters or as needed | ULPA filter (170 L/240 L) yearly or when visibly dirty | HEPA filters every 6 months to a year depending on use |
| Gas inlet filters | Change as needed (preventive maintenance table) | At least yearly or when dirty | 6 months to 1 year |
| Calibration | Verify and document CO2 and temperature calibration; interval depends on use, environment and accuracy | Yearly for temperature, CO2 and humidity sensors; oxygen sensor yearly and as needed | CO2 calibration monthly to quarterly depending on door traffic |
| Door gasket | Inspect door gasket seal | Replace outer magnetic gasket as needed (brittleness, corrosion, wear, damage) | Not covered |
| Heat sterilisation cycle | Not covered | Run decontamination or sterilisation cycle if fitted | Once a month to once every six months |
| Gas tank | Check CO2 tank levels | Daily; regulator gauge not below 15 psig | Not covered |
Sources
- Thermo Scientific Operating and Maintenance Manual 7013010 Rev. 6 (accessed 2026-11-21)
- Esco Lifesciences CO2 Incubator Maintenance Tips and Procedures V.B Mar2021 (accessed 2026-11-21)
- University of Utah IBC Best Practices for CO2 Incubator Maintenance (accessed 2026-11-21)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.