Written and maintained by CASRAI Editorial Board
Published · Last updated
On this page
General standards information for laboratory, clinic and supply staff. It is not medical advice and does not describe how any device is used on a patient.
Short answer
ISO 11070 is the international standard for sterile, single-use intravascular introducers, dilators and guidewires. It sets requirements for introducer needles, introducer catheters, sheath introducers, guidewires and dilators supplied sterile, and the scope also covers kits that combine these devices. The edition most often cited in Europe is BS EN ISO 11070:2014+A1:2018, which BSI lists as current (checked 1 November 2026). A claim that names only “ISO 11070:2014” leaves out the amendment.
What ISO 11070 covers
The NSAI Standards listing gives the title as Sterile single-use intravascular introducers, dilators and guidewires (ISO 11070:2014). Its abstract says the standard specifies requirements for “introducer needles, introducer catheters, sheath introducers, guidewires, and dilators” that are “intended for single use in conjunction with intravascular catheters specified in ISO 10555-1” (NSAI listing, checked 1 November 2026).
BSI Knowledge describes the same scope: the five device types supplied sterile, plus kits that combine them. BSI classifies the standard under committee CH/84 and ICS 11.040.25.
In short, ISO 11070 is a device standard for the accessory devices used alongside intravascular catheters. It is not a catheter standard in itself; the catheters it refers to are specified in ISO 10555-1.
Edition and amendment status
Several records describe one document, which is why a conformity claim can look inconsistent. The table below lists only what the two standards bodies state.
| Document | Status in the record | Source |
|---|---|---|
| I.S. EN ISO 11070:1999 | Superseded by the 2014 edition | NSAI |
| I.S. EN ISO 11070:2014 (published 1 December 2014) | Superseded on 28 January 2020 by the edition with the amendment | NSAI |
| I.S. EN ISO 11070:2014&A1:2018 | Replaced the plain 2014 edition (a corrected version, &LC:2018, is also listed) | NSAI |
| BS EN ISO 11070:2014+A1:2018 (published 30 November 2018) | Current | BSI Knowledge |
BSI states that the +A1:2018 text is identical to EN ISO 11070:2014/A1:2018 and to ISO 11070:2014/Amd 1:2018. The amendment is therefore an international change as well as a European one. The practical point is that “2014” and “2014+A1:2018” are not the same citation, and in the Irish national adoption the plain 2014 edition is recorded as superseded.
Device types in scope
The scope names five device types. CASRAI covers three of them in separate guides, so this page does not repeat their content.
- Introducer needles. Named in the scope; no separate CASRAI guide.
- Introducer catheters. Named in the scope; no separate CASRAI guide.
- Sheath introducers. See introducer sheath: FDA class and product codes.
- Guidewires. See guide wire types and sizes.
- Dilators. See dilator sizes.
Kits that bundle these devices fall within the scope as well. For one common kit format, see the central line kit guide, and for the catheter these accessories are used with, see what is a central venous catheter. A related product-type comparison is Cordis catheter vs central line.
Reading a conformity claim against the published record
This checklist is an editorial reading of the facts above, not a requirement written in the standard. Each line can be checked against the BSI and NSAI records.
| What the claim should show | Why the record makes it checkable |
|---|---|
| The edition, including the amendment (for example 2014+A1:2018) | NSAI records the plain 2014 edition as superseded; BSI lists the +A1:2018 edition as current |
| Which regional adoption is cited (ISO, EN ISO, BS EN ISO, I.S. EN ISO) | The same technical text is published under several designations |
| Which of the five device types the claim is about | The scope lists introducer needles, introducer catheters, sheath introducers, guidewires and dilators |
| Whether a kit is covered as a kit | BSI states the scope includes kits combining these devices |
| That the device is supplied sterile and for single use | Both words appear in the standard’s title and scope |
A standard number on a data sheet is a statement by the manufacturer. The standard text itself and any conformity assessment documents are what substantiate it; this page summarises the published edition records only.
Common misunderstandings
“ISO 11070 is the catheter standard.”
It covers introducers, dilators and guidewires used with intravascular catheters. The catheters themselves are specified in ISO 10555-1, which the abstract names.
“ISO 11070:2014 is the current edition.”
For the European and Irish adoptions, the amended edition (2014+A1:2018) is the one the records show as current, and the plain 2014 edition is shown as superseded.
“It applies only to loose components.”
BSI’s scope statement also covers kits that combine these devices.
FAQ
What is ISO 11070?
It is the standard for sterile single-use intravascular introducers, dilators and guidewires, specifying requirements for introducer needles, introducer catheters, sheath introducers, guidewires and dilators.
Which edition is current?
BSI Knowledge lists BS EN ISO 11070:2014+A1:2018, published 30 November 2018, as current (checked 1 November 2026).
Is the 2014 edition without the amendment still in force?
NSAI records I.S. EN ISO 11070:2014 as superseded on 28 January 2020 by the edition that includes the 2018 amendment.
Sources
- BSI Knowledge, Sterile single-use intravascular introducers, dilators and guidewires (BS EN ISO 11070:2014+A1:2018), checked 1 November 2026.
- NSAI Standards, I.S. EN ISO 11070:2014 listing, checked 1 November 2026.