Microscale Thermophoresis: Working Principle, Kd Measurement, and Lab Protocol

Microscale thermophoresis (MST) measures biomolecular interactions in free solution by tracking molecular migration along a microscopic temperature gradient.

Written and maintained by CASRAI Editorial Board

Published

On this page

Microscale thermophoresis (MST) is an immobilization-free biophysical technique that quantifies biomolecular binding interactions in free solution by monitoring the directed migration of fluorescent molecules through an infrared-laser-induced microscopic temperature gradient. By measuring changes in molecular size, electric charge, and hydration shell upon complex formation, MST delivers precise equilibrium dissociation constants (Kd) from sub-nanomolar to millimolar ranges while consuming as little as four microliters of sample per capillary.

Definition, physical principles, and the Ludwig-Soret effect

Microscale thermophoresis operates on the principle of thermophoresis—often termed the Ludwig-Soret effect or thermodiffusion—in which particles, macromolecules, and ions experience directed physical transport along an applied spatial temperature gradient. While classical macroscopic thermodiffusion was documented in the nineteenth century by Carl Ludwig and Charles Soret, modern microscale thermophoresis adapts this transport phenomenon into a microvolume analytical platform by focusing an infrared (IR) laser beam (typically emitting at 1470 nm or 1480 nm, coinciding with a water absorption band) into a thin fused-silica glass capillary. This optical excitation induces an immediate, localized temperature elevation of 1 K to 6 K within a microscopic radius of 25 μm to 50 μm.

In a continuous solvent phase, the net mass flux density (J) of a solute molecule driven by both thermophoretic drift and opposing Brownian mass diffusion is expressed by the transport equation:

J = -c D_T grad(T) – D grad(c)

In this governing equation, c denotes solute concentration, grad(T) is the spatial temperature gradient, D_T represents the thermophoretic mobility coefficient, D is the translational Brownian diffusion coefficient, and grad(c) is the concentration gradient. When the infrared laser creates a steady-state thermal profile, the opposing fluxes establish a balance (J = 0). Rearranging this condition yields the spatial concentration distribution:

grad(c) / c = -(D_T / D) grad(T) = -S_T grad(T)

Here, the ratio S_T = D_T / D defines the Soret coefficient, expressed in units of inverse kelvin (K^-1). Integrating across the applied temperature increment (ΔT = T_hot – T_cold) produces the steady-state concentration ratio between the heated laser focal zone and the surrounding ambient capillary fluid:

c_hot / c_cold = exp(-S_T ΔT) ≈ 1 – S_T ΔT

Because the laser-induced temperature jump ΔT remains modest (usually 2 K to 5 K), the linear approximation holds with high experimental precision. Thermophoretic movement is not solely governed by molecular mass; rather, S_T is fundamentally determined by the molecule’s hydrodynamic radius (r_h), its net surface charge density (σ), its tertiary or quaternary conformation, and the entropy of its solvent hydration shell (Δs_hyd). When an unlabeled ligand binds to a fluorescently labeled macromolecule, even a minor alteration in surface charge, localized conformational flexibility, or water-shell ordering alters S_T, resulting in a measurable shift in the steady-state concentration profile.

Anatomy of an MST trace: T-jump versus thermophoresis

A standard microscale thermophoresis experiment records fluorescence emission over time across three distinct temporal phases: initial cold baseline, infrared laser activation, and laser deactivation recovery. Understanding these individual phases is essential for resolving true binding signals from photophysical artifacts.

During the initial phase (typically seconds 0 to 5), the excitation light source measures the cold baseline fluorescence (F_initial or F_cold) at ambient capillary temperature before the heating laser is engaged. When the infrared laser switches on, the observed signal displays a two-step response profile:

  • The Temperature Jump (T-jump phase, 0 to 1 second): Within the first several hundred milliseconds of laser irradiation, the local solution heats rapidly. Before significant molecular migration can take place, the fluorescence quantum yield of the fluorophore shifts instantaneously due to temperature dependence. This photophysical phenomenon, termed Temperature-Related Intensity Change (TRIC), is governed by the partial derivative of fluorescence with respect to temperature (∂F/∂T). Because TRIC is exceptionally sensitive to the immediate chemical microenvironment of the fluorophore—including local dielectric constants, nearby quenching residues like tryptophan, and hydrogen bonding—binding events frequently produce a pronounced T-jump change even before net diffusion occurs.
  • The Thermophoresis Phase (1 to 20 seconds): Following the initial thermal relaxation, molecules migrate out of (or into) the heated beam waist along the temperature gradient. For the vast majority of biological macromolecules in standard aqueous buffers, S_T is positive, driving solute depletion away from the hot center toward colder surrounding fluid. This steady mass depletion continues until thermophoretic efflux exactly counterbalances Fickian diffusion influx, establishing a flat steady-state plateau (F_hot).
  • The Back-Diffusion Phase (laser deactivation, 20 to 30 seconds): When the IR laser is turned off, the microscopic temperature gradient collapses within milliseconds. Deprived of the driving thermal force, the concentrated molecules in the surrounding cold fluid diffuse back into the focal area purely via Brownian motion, returning the fluorescence signal toward the original baseline.

To quantify the interaction independently of initial sample loading variations, the software computes the normalized fluorescence (F_norm), defined as the ratio of fluorescence during the heated window to the initial cold baseline:

F_norm = F_hot / F_cold = 1 + (∂F/∂T – S_T) ΔT

Plotting F_norm (conventionally reported in per mille, ‰) as a function of titrated ligand concentration yields a classic sigmoidal saturation binding isotherm. This equilibrium signal combines contributions from both TRIC and thermophoresis, maximizing the sensitivity of detection across diverse molecular classes.

Step-by-step laboratory protocol for Kd determination

Executing a reliable microscale thermophoresis titration requires systematic assay design, careful pipetting techniques, and rigorous equilibration. The following procedural workflow outlines the execution of an equilibrium binding assay for an unlabeled ligand titrated against a fluorescently tagged target protein.

When preparing sensitive microvolume titration series, researchers frequently utilize specialized technique adjustments. Incorporating reverse pipetting protocols helps minimize volume variations and meniscus clinging when transferring viscous reaction buffers or detergent-containing solutions across multiple tubes.

  1. Target Fluorophore Optimization: Prepare the fluorescently labeled target molecule at a constant, low concentration. As a general rule, target concentration should reside at or below the expected Kd (typically between 5 nM and 50 nM for standard Monolith detectors, or 10 pM to 1 nM for high-sensitivity pico detectors). Ensure baseline fluorescence counts fall within the linear range of the optical system (typically 200 to 1,500 counts for standard blue/red channels). Target proteins can be labeled using amine-reactive NHS-ester dyes, thiol-reactive maleimide dyes, His-tag affinity fluorophores (such as RED-tris-NTA), or expressed as GFP/YFP fusion constructs in downstream protein engineering workflows.
  2. Serial Dilution Series Preparation: Prepare a 16-point 1:2 serial dilution series of the unlabeled ligand in microcentrifuge tubes or a low-binding microtiter plate. Set the highest ligand concentration (tube 1) at least 20-fold to 50-fold above the estimated dissociation constant to guarantee complete curve saturation at the upper plateau. Dilute the ligand across tubes 2 through 16 using assay buffer, leaving tube 16 as an interaction-free negative control containing buffer only. Ensure the total ligand preparation volume exceeds 15 μL to 20 μL per tube to avoid evaporative loss during handling.
  3. Equimolar Mixing and Incubation: Add an equal volume of the constant fluorescent target solution to each of the 16 dilution tubes (e.g., 10 μL of 20 nM target mixed with 10 μL of ligand dilution, yielding 10 nM target across all 16 points). Mix gently by pipetting without creating bubbles or foam. Incubate the reaction mixtures at ambient temperature (or controlled assay temperature) for 10 to 30 minutes to allow the binding equilibrium to stabilize fully.
  4. Capillary Loading and Capillary Scan: Immerse the tip of a glass capillary into each reaction tube; the 4 μL to 6 μL volume aspirates spontaneously via capillary action. Place the 16 capillaries sequentially onto the instrument capillary tray. Initiate the automated pre-run capillary scan. The instrument moves the tray across the optical detector to measure initial fluorescence profiles across all 16 positions, verifying uniform loading, sample concentration consistency, and the absence of meniscus bubbles.
  5. Infrared Laser Measurement and Data Acquisition: Set the excitation LED power (typically 20% to 80%) to achieve optimal baseline counts, and select the IR laser heating power (typically low, medium, or high, corresponding to approximately 20%, 40%, or 80% laser output, producing ΔT values between 2 K and 6 K). Run the measurement protocol. The automated system scans through capillaries 1 to 16, exposing each capillary to the heating sequence and recording fluorescence traces in real time.
  6. Mathematical Curve Fitting: Export the recorded traces into analysis software. The software extracts F_norm from a chosen analysis window (e.g., T-jump at 0.5 s to 1.5 s, or thermophoresis at 4.0 s to 5.0 s) and plots F_norm against the logarithm of total ligand concentration [L]_total. To account for potential target depletion when the constant target concentration [T]_total approaches the true Kd, data are fitted using the quadratic binding model derived from the law of mass action:

[TL] = ([T]_total + [L]_total + Kd – sqrt(([T]_total + [L]_total + Kd)^2 – 4 [T]_total [L]_total)) / 2

F_norm([L]_total) = F_unbound + (F_bound – F_unbound) * ([TL] / [T]_total)

In this equation, [TL] represents the concentration of target-ligand complex at equilibrium, F_unbound is the normalized fluorescence of the free target, and F_bound is the normalized fluorescence of the fully saturated complex. Fitting extracts both Kd and the corresponding 95% confidence intervals.

Worked calculation: small-molecule inhibitor binding to a protein kinase

To illustrate how experimental MST raw data translates into an equilibrium dissociation constant, consider a representative binding titration between a fluorescently labeled serine/threonine protein kinase (catalytic domain, molecular weight 42 kDa) and a low-molecular-weight kinase inhibitor (molecular weight 380 Da).

In this assay setup, the labeled kinase is held constant at [T]_total = 20 nM across all 16 capillaries. The inhibitor is prepared in a 1:2 serial dilution series starting at 10,000 nM (10 μM) down to 0.305 nM in capillary 15, with capillary 16 serving as the ligand-free reference control ([L]_total = 0 nM). The assay buffer contains 20 mM HEPES (pH 7.5), 150 mM NaCl, 5 mM MgCl2, 1 mM DTT, and 0.05% Tween-20 to suppress capillary surface adsorption. The instrument records F_cold between 0.0 s and 1.0 s, engages the IR laser at 40% power (ΔT ≈ 3.5 K) from 1.0 s to 21.0 s, and integrates F_hot between 4.0 s and 5.0 s.

The resulting baseline fluorescence is F_cold = 820 counts for the free kinase. At 40% IR laser power, thermophoretic depletion lowers the focal fluorescence to F_hot = 746 counts. The baseline normalized fluorescence for the unbound target is therefore:

F_unbound = (746 / 820) * 1000 = 909.76 ‰

At high inhibitor concentrations (e.g., capillaries 1 and 2, where [L]_total reaches 10,000 nM and 5,000 nM), the kinase active site is fully saturated. Because inhibitor binding alters the surface hydration entropy and local electrostatic charge of the ATP-binding pocket, thermophoretic migration increases, resulting in greater focal depletion: F_hot drops to 728 counts for the same initial F_cold = 820 counts. The normalized fluorescence of the fully bound complex is:

F_bound = (728 / 820) * 1000 = 887.80 ‰

The total experimental amplitude (dynamic response window) is:

ΔF_norm = F_bound – F_unbound = 887.80 ‰ – 909.76 ‰ = -21.96 ‰

An amplitude of ~22 ‰ represents an excellent signal-to-noise ratio in MST (standard instrument noise is typically ≤ 0.5 ‰, yielding a signal-to-noise ratio > 40). At capillary 8, where the total inhibitor concentration is [L]_total = 78.1 nM, the observed normalized fluorescence is 898.78 ‰. The fraction of bound kinase (θ) at this titration point is calculated as:

θ = (F_observed – F_unbound) / (F_bound – F_unbound) = (898.78 – 909.76) / (887.80 – 909.76) = -10.98 / -21.96 = 0.500

Because exactly 50% of the target kinase is bound (θ = 0.50), the concentration of complex is [TL] = 0.50 * 20 nM = 10 nM. The free target concentration is [T]_free = 10 nM, and the free inhibitor concentration is [L]_free = [L]_total – [TL] = 78.1 nM – 10 nM = 68.1 nM. Applying the equilibrium mass action relationship:

Kd = ([T]_free * [L]_free) / [TL] = (10 nM * 68.1 nM) / 10 nM = 68.1 nM

Nonlinear least-squares fitting across all 16 capillaries using the quadratic binding equation yields an identical fitted value of Kd = 68.2 ± 3.4 nM. This worked example illustrates why fitting with the quadratic model is necessary: when [T]_total (20 nM) is within an order of magnitude of the Kd (68 nM), assuming [L]_free ≈ [L]_total introduces a ~15% systematic error in the calculated affinity.

Comparing MST with SPR, ITC, and BLI

Selecting an appropriate biophysical method for interaction analysis requires balancing sample consumption, sensitivity, throughput, and physical constraints such as surface immobilization. The structured table in the dataset below compares technical specifications across these technologies, while the discussion here highlights practical selection tradeoffs encountered in the laboratory.

In structural biology and functional proteomics, researchers often corroborate interactions identified in computational databases like the STRING protein-protein interaction database or validated biochemically via Western blotting and ELISA assays using biophysical titration methods. Here is how MST compares to the three primary alternatives:

  • Surface Plasmon Resonance (SPR): SPR is the historical gold standard for kinetic analysis, delivering real-time association rate constants (k_on) and dissociation rate constants (k_off) alongside affinity constants. However, SPR requires immobilizing one binding partner onto a gold sensor chip via amine coupling, biotin-streptavidin capture, or nickel-NTA coordination. Immobilization can introduce steric hindrance, alter native protein conformation, or mask critical binding epitopes. In contrast, MST operates entirely in free solution inside inert glass capillaries, eliminating surface artifacts. Furthermore, SPR microfluidic flow cells are highly susceptible to clogging when exposed to crude cell lysates, liposomes, or viscous samples, whereas MST capillaries easily accommodate unpurified biological matrices.
  • Isothermal Titration Calorimetry (ITC): ITC is the definitive reference method for universal label-free thermodynamic profiling, measuring reaction enthalpy (ΔH), entropy (ΔS), and binding stoichiometry (n) directly from the heat absorbed or released upon titration. However, ITC suffers from high sample consumption (requiring 200 μL to 1,400 μL of protein at micromolar concentrations per titration run) and low throughput (typically 1 to 4 runs per day). MST requires less than 5% of the protein mass consumed in a single ITC experiment and operates at nanomolar concentrations, making MST practical when working with scarce or difficult-to-express proteins.
  • Bio-Layer Interferometry (BLI): BLI utilizes optical fiber biosensors in a fluidics-free “dip-and-read” format, eliminating microfluidic clogs while measuring binding kinetics. Like SPR, however, BLI relies on surface immobilization of the target molecule on the biosensor tip. In addition, BLI exhibits lower sensitivity when screening low-molecular-weight chemical fragments (< 200 Da) binding to large proteins due to the physical limit of optical thickness changes. Because MST responds to changes in hydration shell and surface charge rather than added mass alone, MST readily measures small molecules, peptides, and ions binding to large complexes.

When quantifying target concentrations before setting up MST titrations, laboratories commonly employ fluorescence-based quantification such as the PicoGreen assay protocol for double-stranded DNA interactants or fluorometric dye-binding assays for proteins to ensure accurate stock concentrations.

Capillary selection, artifacts, and the SD-test

Because microscale thermophoresis relies on optical tracking within narrow capillaries, interactions between the sample and the capillary inner wall represent the primary source of experimental noise and artifacts. Implementing systematic quality control steps ensures data integrity.

Modern MST systems offer three distinct capillary surface chemistries:

  • Standard Treated Capillaries: Manufactured from high-purity fused silica with a standard surface treatment. Standard capillaries serve as the economical default for well-behaved, soluble proteins with neutral or moderately acidic isoelectric points.
  • Hydrophobic Capillaries: Engineered with specialized inner coatings to alter surface wetting properties. These capillaries are utilized for non-polar molecules, organic co-solvent systems, or specific membrane-mimetic environments where standard glass interactions occur.
  • Premium Coated (Hydrophilic) Capillaries: Passivated with a densely grafted, covalently attached hydrophilic polymer layer that effectively shields surface silanol groups. Premium capillaries are mandatory when working with basic proteins (pI > 8.5), highly sticky intrinsically disordered proteins, polycationic peptides, or detergent-sensitive membrane proteins that adhere non-specifically to bare silica walls.

To identify experimental anomalies before acquiring thermophoresis data, operators must inspect the automated capillary scan profile. A healthy capillary scan displays 16 distinct, symmetric, Gaussian-like fluorescence peaks of uniform height. Three common anomalies indicate data-compromising artifacts:

  1. Split Peaks or Shoulders: When a capillary scan reveals double peaks, distorted peak crests, or asymmetric shoulders, sample molecules are adsorbing non-specifically to the inner capillary walls (sticking). The fix requires switching from Standard to Premium Coated capillaries, adding 0.005% to 0.05% non-ionic surfactant (Tween-20 or Pluronic F-127), or supplementing the buffer with 0.1 mg/mL bovine serum albumin (BSA).
  2. Ligand-Dependent Initial Fluorescence Shifts: If initial cold fluorescence (F_initial) varies across the 16 capillaries by more than ±10% in a concentration-dependent manner, this shift may indicate either a genuine binding-induced conformational change around the fluorophore (local quenching/enhancement) or compound-induced aggregation, loss of solubility, or autofluorescence. To differentiate between these mechanisms, researchers perform an SD-test (SDS Denaturation test). In the SD-test, aliquots from the highest ligand capillary and the ligand-free control capillary are treated with 4% SDS and 40 mM DTT and heated to 95 °C for 5 minutes to fully denature all proteins. If the fluorescence difference disappears after denaturation, the original shift reflected a specific binding-induced microenvironment change around the intact native protein, confirming that the fluorescence data reflects valid binding. If the difference persists after complete denaturation, the ligand itself is either quenching the fluorophore directly, fluorescing at the detection wavelength, or causing irreversible target precipitation.
  3. Irregular Bumps and Spikes in MST Traces: Jagged spikes, erratic baseline jumps, or non-exponential relaxation curves during heating indicate macroscopic aggregation, dust particles, or micro-bubbles passing through the laser beam waist. Centrifuging all stock solutions and diluted mixtures at 15,000 × g for 10 minutes prior to loading capillaries effectively removes pre-formed aggregates and particulates.

Common misunderstandings and practical pitfalls

Despite the operational simplicity of microscale thermophoresis, several conceptual and technical misconceptions frequently compromise assay interpretation:

  • Misconception: MST measures mass diffusion like dynamic light scattering. While thermophoresis involves particle movement, it is fundamentally driven by the thermal diffusion coefficient D_T, not mass diffusion D alone. Because S_T = D_T / D depends strongly on surface charge density and hydration entropy, MST can easily detect binding of a 100 Da ion or chemical fragment to a 150 kDa antibody if binding alters local surface hydration or charges, whereas mass-dependent techniques like SPR or light scattering struggle with tiny mass changes.
  • Misconception: A flat MST response proves an interaction does not occur. In rare cases, a ligand binds to a target without altering the hydrodynamic radius, net surface charge, or hydration shell, resulting in ΔS_T ≈ 0. If TRIC also produces no shift (∂F/∂T unchanged), the net amplitude ΔF_norm will be zero despite robust binding. When an expected interaction yields a flat line, researchers should re-test using an alternative fluorophore conjugated to a different residue, switch from N-terminal to C-terminal labeling, or perform a label-free assay tracking intrinsic tryptophan fluorescence.
  • Pitfall: Operating with excessive target concentrations. Using a fluorescent target concentration substantially higher than the true Kd forces the titration into a stoichiometric titration regime rather than an equilibrium binding regime. When [T]_total >> Kd, the binding isotherm becomes linear rather than sigmoidal, preventing accurate mathematical extraction of sub-nanomolar dissociation constants. Always calibrate LED power to allow the lowest workable target concentration.
  • Pitfall: Neglecting DMSO and co-solvent matching. Small-molecule inhibitor screening often introduces dimethyl sulfoxide (DMSO) from compound stock solutions. Because thermophoresis is exceptionally sensitive to solvent viscosity and thermal conductivity, even a 0.5% mismatch in DMSO concentration between dilution tubes introduces severe baseline drift that distorts Kd calculations. All serial dilution steps must maintain an exact, constant concentration of vehicle solvent (e.g., exactly 1.0% DMSO across all 16 tubes).
  • Pitfall: Air bubble entrapment during capillary filling. Immersing capillaries too quickly or touching the meniscus at an angle can trap micro-bubbles inside the capillary bore. Air bubbles diffract the excitation light and distort the temperature field, leading to failed capillary scans and erratic traces. Inspect filled capillaries against a dark background before loading onto the tray.

Frequently asked questions

What is microscale thermophoresis used for?

Microscale thermophoresis is primarily used in academic and pharmaceutical laboratories to quantify biomolecular interactions in free solution. Common applications include measuring binding affinities (dissociation constants, Kd) between proteins and small-molecule drugs, protein-protein interactions, antibody-antigen binding, protein-nucleic acid complexes, and protein-lipid interactions. It is especially valued for analyzing interactions in complex biological liquids like cell lysates and blood serum without target purification.

How much sample is required for an MST experiment?

A standard MST assay consumes remarkably low sample volumes. Each measurement capillary requires only 4 μL to 6 μL of reaction mixture. A complete 16-point serial titration series requires approximately 100 μL to 150 μL of total sample volume per binding partner, with target macromolecule concentrations typically ranging from 5 nM to 50 nM (or down to picomolar concentrations on specialized high-sensitivity instruments).

Can MST be performed without fluorescent labeling?

Yes. Label-free microscale thermophoresis instruments monitor intrinsic protein fluorescence from native tryptophan and tyrosine residues using specialized ultraviolet (UV) optics (excitation at 280 nm, emission at 360 nm). Label-free MST eliminates the chemical modification step, ensuring proteins remain entirely in their native, unmodified state, provided the target contains at least one solvent-exposed tryptophan residue.

What is the difference between TRIC and thermophoresis?

TRIC (Temperature-Related Intensity Change) is an instantaneous photophysical change in fluorophore brightness caused purely by the temperature shift (ΔT) within the first second of heating, reflecting the immediate chemical microenvironment of the fluorophore. Thermophoresis is the subsequent physical mass transport of molecules out of the heated laser focal zone over several seconds, governed by the Soret coefficient (S_T). MST analysis software can evaluate TRIC alone, thermophoresis alone, or a combined window.

How does MST compare to surface plasmon resonance (SPR)?

MST measures binding entirely in free solution inside capillaries without immobilizing either partner on a sensor surface, avoiding surface-induced denaturation, steric hindrance, or microfluidic clogging. SPR requires immobilizing one partner on a sensor chip and excels at resolving real-time association (k_on) and dissociation (k_off) kinetic rates, whereas MST primarily measures equilibrium dissociation constants (Kd).

What does the Soret coefficient (S_T) measure?

The Soret coefficient (S_T = D_T / D) quantifies the ratio of thermophoretic mobility to Brownian diffusion, describing how strongly a solute concentrates or depletes along a temperature gradient. In microscale thermophoresis, S_T is determined by a molecule’s hydrodynamic radius, net surface charge, conformation, and hydration entropy.

Why is an SD-test performed in microscale thermophoresis?

The SDS Denaturation test (SD-test) is a diagnostic control performed when initial cold fluorescence varies significantly across ligand titration points. By denaturing the samples with SDS and DTT at 95 °C, the test determines whether the fluorescence change was caused by a specific native binding interaction (which disappears upon unfolding) or by non-specific compound aggregation, quenching, or capillary sticking (which persists after denaturation).

Sources

  • Wienken, C. J., Baaske, P., Rothbauer, U., Braun, D., & Duhr, S. (2010). Protein-binding assays in biological liquids using microscale thermophoresis. Nature Communications, 1, 100. https://www.nature.com/articles/ncomms1093 (accessed 2026-11-15)
  • Jerabek-Willemsen, M., André, T., Wanner, R., Roth, H. M., Duhr, S., Baaske, P., & Breitsprecher, D. (2014). MicroScale Thermophoresis: Interaction analysis and beyond. Journal of Molecular Structure, 1077, 101-113. https://doi.org/10.1016/j.molstruc.2014.03.009 (accessed 2026-11-15)
  • Seidel, S. A. I., Dijkman, P. M., Lea, W. A., van den Bogaart, G., Jerabek-Willemsen, M., Lazic, A., Schmitz, J. S., Marte, N. J., Duhr, S., Baaske, P., & Breitsprecher, D. (2013). Microscale thermophoresis quantifies biomolecular interactions under previously challenging conditions. Methods, 59(3), 301-315. https://doi.org/10.1016/j.ymeth.2012.12.005 (accessed 2026-11-15)
  • Sparks, R. P., Becker, A., Arango, A. S., Tajkhorshid, E., & Fratti, R. A. (2020). Use of Microscale Thermophoresis to Measure Protein-Lipid Interactions. Journal of Visualized Experiments, 160, e60607. https://www.jove.com/t/60607 (accessed 2026-11-15)
  • NanoTemper Technologies. (2024). Monolith NT.115 and Monolith Pico Technical Specifications and Assay Optimization Guide. https://nanotempertech.com/monolith/ (accessed 2026-11-15)

Data: Biophysical Interaction Method Specifications and Operational Parameters

Comparative specifications, sample consumption thresholds, dynamic affinity ranges, and immobilization requirements across solution and surface biophysical binding methods compiled from peer-reviewed literature and instrument technical specifications.

Period covered: 2010/2026. Coverage: International.

Biophysical Interaction Method Specifications and Operational Parameters
Biophysical MethodPrimary Readout MechanismSample EnvironmentImmobilization RequiredMinimum Sample Volume (microliters)Typical Dynamic Affinity Range (molar)Kinetic Rate ResolutionMatrix Tolerance
Microscale Thermophoresis (MST)Fluorescence change in microscale thermal gradientFree solution in glass capillaryNo41 pM to 1 mMLimited (primarily equilibrium Kd)High (lysates, serum, plasma)
Surface Plasmon Resonance (SPR)Refractive index shift on optical sensor surfaceSolid surface interface / microfluidic flowYes5010 pM to 1 mMHigh (real-time kon and koff)Low to Moderate (requires clean analyte)
Isothermal Titration Calorimetry (ITC)Heat of reaction (enthalpy change) upon titrationFree solution in sample cellNo20010 nM to 100 uMNo (thermodynamic equilibrium only)Low (susceptible to buffer mismatch heat)
Bio-Layer Interferometry (BLI)Optical interference pattern shift at fiber-optic tipSolid surface interface / dip-and-readYes4010 pM to 1 mMHigh (real-time kon and koff)Moderate to High (fluidics-free tip format)
Mass Photometry (MP)Interferometric light scattering of single particlesLanding on glass coverslip interfaceNo1010 nM to 1 uMLimited (single-particle equilibrium mass)Low (requires low background scattering)

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Microscale Thermophoresis: Working Principle, Kd Measurement, and Lab Protocol

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →