The PicoGreen assay is an ultrasensitive fluorescence-based analytical protocol designed to quantify double-stranded DNA (dsDNA) in aqueous solution across a dynamic range spanning 25 pg/mL to 1,000 ng/mL. Developed around an unsymmetrical cyanine fluorophore, the assay exhibits an increase in fluorescence emission exceeding 1,000-fold upon selective binding to dsDNA, while remaining virtually non-fluorescent in its unbound state. This photophysical selectivity enables accurate, high-throughput nucleic acid quantification in microplate formats even in the presence of equimolar single-stranded DNA, RNA, and common laboratory extraction reagents that compromise classical ultraviolet absorbance measurements.
Definition, Core Mechanism, and Optical Properties
The PicoGreen assay relies on Quant-iT PicoGreen dsDNA reagent, an unsymmetrical cyanine dye originally synthesized and characterized by Molecular Probes (now Thermo Fisher Scientific) in the late 1990s. In free aqueous solution at physiological or neutral pH, the PicoGreen molecule exhibits low intrinsic quantum yield because rotational freedom around its central methine bridge allows rapid non-radiative dissipation of absorbed photon energy. When the fluorophore interacts with double-stranded DNA, it intercalates between adjacent base pairs and binds along the minor groove of the double helix. This binding event restricts torsional rotation, locking the chromophore into a rigid conformation that channels absorbed energy into radiative fluorescence emission, yielding an enhancement of more than 1,000-fold over free dye background.
The optical excitation maximum of DNA-bound PicoGreen occurs at approximately 480 to 502 nm, and its fluorescence emission peaks between 520 and 523 nm. These spectral coordinates align closely with the optical filter and dichroic mirror configurations standardly used for fluorescein isothiocyanate (FITC), allowing laboratories to execute the assay on conventional fluorescence microplate readers, filter fluorometers, spectrofluorometers, and even standard real-time quantitative PCR instruments without requiring specialized laser lines. Typically, microplate readers are configured with an excitation filter of 485 ± 10 nm and an emission filter of 528 ± 10 nm or 535 ± 15 nm.
A primary chemical advantage of PicoGreen over older fluorophores such as Hoechst 33258 (a bisbenzimide dye) is its relative insensitivity to base-pair composition. While Hoechst 33258 exhibits pronounced preferential binding toward adenine-thymine (AT)-rich tracts, resulting in significant calibration errors when standardizing mammalian, bacterial, or GC-rich templates against a single reference, PicoGreen exhibits negligible base selectivity across varying genomic compositions. Consequently, researchers can accurately quantify diverse genomic DNAs, PCR amplicons, and synthetic constructs using a single standardized reference, such as purified bacteriophage lambda DNA. When quantifying complex mixtures or comparing circular constructs to linearized libraries, understanding topological behavior is essential, as detailed in our guide to plasmid DNA vs genomic DNA isolation and structural characteristics.
Dynamic Range, Standard Curves, and Microplate Workflows
The PicoGreen assay provides a linear dynamic range spanning more than four orders of magnitude, traditionally divided into two distinct operating protocols depending on anticipated analyte concentrations: the high-range assay and the low-range assay. Selecting the appropriate assay range optimizes dye economy, prevents fluorophore saturation, and maintains precision across both concentrated purification yields and trace forensic or liquid biopsy isolates.
- High-Range Protocol (1 ng/mL to 1,000 ng/mL): The working dye reagent is prepared by diluting the concentrated DMSO stock solution 200-fold into 1X TE buffer (10 mM Tris-HCl, 1 mM EDTA, pH 7.5). The standard curve is constructed using bacteriophage lambda DNA diluted to concentrations of 1,000 ng/mL, 500 ng/mL, 100 ng/mL, 10 ng/mL, 1 ng/mL, and a 0 ng/mL reagent blank. In a standard 96-well microplate, 100 µL of each standard or unknown sample is combined with 100 µL of the 200-fold diluted PicoGreen working solution to yield a final reaction volume of 200 µL per well.
- Low-Range Protocol (25 pg/mL to 25 ng/mL): For ultra-low input samples, such as single-cell sequencing inputs, laser capture microdissections, or ancient DNA extracts, the concentrated stock is diluted 2,000-fold in 1X TE buffer. Standard series points typically span 25 ng/mL, 2.5 ng/mL, 250 pg/mL, 25 pg/mL, and a blank. This format enables reproducible quantification down to absolute quantities as low as 2.5 to 5 pg of dsDNA in a 200 µL reaction well.
Executing the assay requires systematic attention to liquid handling, plate geometry, and light exposure. Unlike colorimetric benchmarks discussed in our overview of protein assays, where absorbance values stabilize over extended periods, fluorescent cyanine dyes are sensitive to photobleaching and surface adsorption. The standard bench procedure follows sequential execution stages:
- Buffer and Reagent Preparation: Prepare fresh 1X TE working buffer from molecular-biology-grade stocks using nuclease-free deionized water. Ensure the pH is calibrated between 7.5 and 8.0, as acidic pH below 7.0 reduces fluorescence intensity. Equilibrate the PicoGreen DMSO stock to room temperature before opening to avoid condensation, then dilute into plastic tubes.
- Standard Dilution Series: Dilute the reference bacteriophage lambda DNA stock serially in 1X TE buffer across clean polypropylene microcentrifuge tubes or a dedicated deep-well dilution plate. Never perform serial dilutions directly inside the final assay plate to avoid cumulative pipetting meniscus errors.
- Plating Samples and Standards: Dispense 100 µL of standards in triplicate and unknown test samples in duplicate or triplicate into black-walled, flat-bottom 96-well or 384-well microplates. Black plates are mandatory to suppress well-to-well optical crosstalk and autofluorescent background scatter.
- Reagent Addition and Incubation: Add 100 µL of diluted PicoGreen working solution to each well using a calibrated multichannel pipette or automated liquid handler. Mix gently by orbital plate shaking for 20 to 30 seconds or gentle pipette aspiration. Incubate the microplate for 2 to 5 minutes at room temperature (20 °C to 25 °C), completely shielded from ambient light using aluminum foil or a light-tight drawer.
- Fluorescence Measurement: Read the plate on a fluorescence reader with excitation set near 485 nm and emission near 528 nm. Acquire data using consistent photomultiplier tube (PMT) gain settings or automated dynamic gain optimization calibrated against the top standard.
- Data Reduction: Subtract the average fluorescence value of the reagent blank wells from all standard and sample measurements. Plot net relative fluorescence units (RFU) versus DNA concentration. For the high-range assay, linear regression analysis typically yields a coefficient of determination (R²) exceeding 0.995. Unknown concentrations are calculated directly from the regression slope and intercept, accounting for any pre-dilution factors applied before loading.
Contaminant Tolerance and Selectivity Thresholds
In routine molecular workflows, DNA samples rarely exist as pure isolates in plain water; they contain residual salts, chaotropes, organic solvents, detergents, and proteins carried over from lysis and silica-column purification procedures. One of the defining technical strengths of the PicoGreen assay is its substantial resilience against common chemical interferents that severely invalidate optical absorbance readings on spectrophotometers.
Nevertheless, specific reagents alter dye-DNA binding kinetics or directly quench fluorophore emission when present above defined critical concentrations. The table below delineates the empirical tolerance limits of the PicoGreen assay against common laboratory reagents, showing the highest concentration at which the measured fluorescence signal deviates by less than 10% from an uncontaminated control:
| Chemical Reagent / Interferent | Class or Application | Maximum Tolerated Concentration | Mechanism of Interference Above Threshold |
|---|---|---|---|
| Sodium Chloride (NaCl) | Neutral salt / ionic strength | 200 mM | High ionic strength competitively destabilizes electrostatic binding between dye and DNA phosphate backbone. |
| Magnesium Chloride (MgCl₂) | Divalent cation / enzyme cofactor | 10 mM | Divalent cations compete for minor groove binding sites and alter DNA secondary conformation at elevated concentrations. |
| Urea | Protein denaturant | 1.0 M | Denatures double-stranded DNA into single strands, extinguishing the specific intercalation signal. |
| Ethanol | Precipitation solvent | 10% (v/v) | Alters dielectric constant of solvent; concentrations above 10% can induce partial DNA compaction and dye precipitation. |
| Phenol | Extraction solvent | 0.1% (v/v) | Direct chemical quenching and optical absorption; residual phenol causes substantial false absorbance on UV instruments. |
| Bovine Serum Albumin (BSA) | Carrier protein / blocking agent | 100 µg/mL (10-fold excess) | Minimal binding below 100 µg/mL; massive protein excess causes non-specific hydrophobic dye sequestering. |
| Sodium Dodecyl Sulfate (SDS) | Anionic detergent | 0.01% (w/v) | Critical inhibitor; SDS forms micellar structures that encapsulate the hydrophobic cyanine dye, causing severe fluorescence loss. |
| Triton X-100 | Non-ionic detergent | 0.1% (v/v) | Well-tolerated at typical lysis levels; excessive non-ionic surfactant perturbs dye solvation. |
| Agarose | Electrophoresis matrix | 0.1% (w/v) | Solubilized molten agarose introduces light scatter and viscous restriction of diffusion. |
| Free dNTPs | Enzyme substrates | 100 µM | Unincorporated mononucleotides do not provide an organized helical duplex structure and yield negligible signal. |
When preparing samples downstream of harsh extraction chemistries, understanding these thresholds prevents severe quantification errors. For example, if an extraction protocol utilizes a high-SDS lysis buffer, thorough column washing or magnetic bead rinsing is critical to prevent detergent carryover from reaching the 0.01% critical threshold in the final assay well, as explored in our foundational review of DNA extraction protocol basics.
Regarding nucleic acid specificity, PicoGreen exhibits marked preference for double-stranded DNA over single-stranded nucleic acids. Equimolar amounts of single-stranded oligonucleotides, denatured DNA, or ribosomal RNA yield fluorescence signals that are generally less than 10% of the signal generated by duplex DNA. However, in cellular extracts where total RNA mass exceeds genomic DNA mass by 10-fold to 50-fold, that residual 10% RNA cross-reactivity can produce a 50% to 100% overestimation of genuine dsDNA content. In such applications, treating the sample with DNase-free RNase A (10 to 20 µg/mL for 15 minutes at 37 °C) prior to PicoGreen addition eliminates RNA-derived background and restores absolute quantification accuracy.
Types and Related Concepts: PicoGreen, RiboGreen, and Optical Platforms
Fluorescent nucleic acid quantitation extends beyond dsDNA into specialized sister chemistries and competing instrumental platforms. Understanding how PicoGreen relates to companion dyes, such as RiboGreen, and instrument formats provides essential clarity for experimental design.
The RiboGreen RNA Assay Variant
While PicoGreen targets duplex DNA, its sister reagent, Quant-iT RiboGreen (characterized by Singer, Jones, and colleagues in 1998), is engineered specifically for solution-based RNA quantification. RiboGreen is also an unsymmetrical cyanine dye with excitation and emission maxima (~500 nm excitation, ~525 nm emission) that match fluorescein optical channels. Like PicoGreen, RiboGreen exhibits dramatic fluorescence enhancement upon nucleic acid binding, enabling RNA detection down to 1 ng/mL in microplate formats (spanning 1 ng/mL to 1,000 ng/mL across low- and high-range setups).
However, an essential biochemical difference exists between the two assays: while PicoGreen is highly selective for dsDNA over RNA, RiboGreen binds both RNA and double-stranded DNA with comparable affinity. Consequently, when quantifying RNA in extracts that contain residual genomic DNA, the sample must be pre-treated with RNase-free DNase I to degrade DNA fragments before running the RiboGreen assay. Alternatively, subtracting a companion PicoGreen measurement from a total nucleic acid RiboGreen readout allows mathematical deconvolution of mixed samples.
Instrument Platforms: Microplates, Qubit, and Spectrophotometry
Laboratories quantifying nucleic acids choose between three predominant technological paradigms: microplate-based fluorometry (PicoGreen), single-tube dedicated fluorometry (such as the Invitrogen Qubit platform), and microvolume ultraviolet-visible spectrophotometry (such as the Thermo Scientific NanoDrop instrument). Each format presents distinct trade-offs across sensitivity, dynamic range, specificity, and throughput:
| Operational Parameter | PicoGreen Microplate Assay | Qubit Fluorometer (HS/BR) | NanoDrop Spectrophotometer | Hoechst 33258 Dye Assay |
|---|---|---|---|---|
| Primary Detection Principle | Fluorescence intercalation (dye binding) | Fluorescence intercalation (dedicated tube) | UV absorbance at 260 nm (Beer-Lambert law) | Minor groove bisbenzimide fluorescence |
| Lower Limit of Detection | 25 pg/mL (cuvette) / 250 pg/mL (plate) | 10 pg/µL (10 ng/mL in assay tube) | ~2,000 pg/µL (2 ng/µL, highly noisy) | 10,000 pg/mL (10 ng/mL) |
| Upper Linear Limit | 1,000 ng/mL (high range) | 100 ng/µL (HS) / 1,000 ng/µL (BR) | 15,000 to 27,500 ng/µL | 15,000 ng/mL |
| Analyte Specificity | High (specific to dsDNA; under 10% RNA/ssDNA cross-reactivity) | High (specific to target assay kit) | None (absorbs all nucleic acids, dNTPs, phenol) | Moderate (dsDNA-specific, high AT bias) |
| Sample Consumption | 1 to 10 µL of sample (in 100 µL dilution) | 1 to 20 µL loaded into 200 µL tube | 1 to 2 µL neat sample (non-destructive) | 10 to 100 µL sample volume |
| Throughput Capacity | High (96 or 384 samples in 5 minutes) | Low to Moderate (1 tube per read, 10–20/run) | Low (1 pedestal read per minute) | Moderate (cuvette or microplate) |
| Instrumentation Cost | Standard fluorescence plate reader | Dedicated benchtop instrument | Dedicated microvolume spectrophotometer | Fluorescence plate reader or fluorometer |
For an in-depth breakdown of how dedicated fluorometers compare to pedestal absorbance devices in purchase decisions and core lab operations, consult our detailed analysis of Qubit vs NanoDrop instrumentation. For foundational principles of UV-Vis absorbance calculation and pathlength physics, refer to our overview of UV-Vis spectrophotometer basics.
The companion data table below compares the photophysical properties, detection thresholds, and dynamic ranges across fluorometric nucleic acid reagents and optical benchmarks.
Where It Matters: High-Throughput Genomics, Biobanking, and Quality Control
The operational value of the PicoGreen assay is most evident in applications where high sample numbers intersect with low analyte concentrations and stringent downstream quality requirements. Core genomics laboratories, automated biobanks, and clinical diagnostic discovery pipelines rely extensively on PicoGreen for critical operational checkpoints.
Next-Generation Sequencing (NGS) Library Preparation and Normalization
In high-throughput sequencing pipelines (such as Illumina, PacBio, and Oxford Nanopore workflows), accurate library quantification is the single most critical factor governing sequencing data quality. Over-clustering on Illumina flow cells results in overlapping fluorescent clusters, mixed signal deconvolution failures, and catastrophic drops in Q30 quality scores. Conversely, under-clustering wastes expensive flow cell real estate and yields insufficient reads to meet experimental coverage targets.
Because enzymatic fragmentation, adapter ligation, and PCR enrichment yield minute quantities of dual-indexed libraries—often in 96-well or 384-well plate configurations at concentrations below 5 ng/µL—absorbance-based spectrophotometry is entirely unsuited for quantification. PicoGreen enables automated liquid handlers to quantify an entire 384-well plate in parallel within minutes, feeding concentration data into automated normalization scripts that pool equimolar amounts of hundreds of libraries into a single sequencing run.
Liquid Biopsies and Cell-Free DNA (cfDNA) Assessment
Non-invasive prenatal testing (NIPT) and oncology liquid biopsies isolate cell-free circulating DNA from peripheral blood plasma. Total cfDNA yields frequently range from 1 to 10 ng per milliliter of plasma, eluting into final volumes of 20 to 50 µL with concentrations often falling below 0.2 ng/µL (200 pg/µL). At these low concentrations, UV absorbance cannot generate a detectable signal above detector noise. The low-range PicoGreen protocol reliably quantifies these trace fractions, ensuring that input masses meet enzymatic library conversion thresholds before committing valuable patient samples to expensive sequencing runs.
Biobanking and Forensic Repository Operations
Large-scale biorepositories storing hundreds of thousands of genomic DNA specimens require rapid, automated quality control upon sample intake. Operating automated fluorometric plate readers equipped with PicoGreen allows biobanks to establish standardized concentration registries while consuming less than 1% of the archived sample volume. Furthermore, pairing PicoGreen with automated robotic liquid handling reduces technician exposure to repeated strain injuries and eliminates manual pipetting variance.
Laboratory Consumables and Vessel Handling Requirements
A critical operational consideration when procuring supplies for PicoGreen assays involves vessel material composition. Untreated borosilicate glass surfaces possess fixed negative surface charges (silanol groups) that exhibit strong non-specific affinity for cationic and hydrophobic cyanine dyes. Preparing working solutions or executing assays in glass test tubes or glass-bottom plates results in rapid dye depletion onto vessel walls, causing signal suppression of up to 40% within 10 minutes. Consequently, all reagent preparation, standard dilutions, and assay reads must be conducted exclusively in virgin polypropylene tubes or polystyrene microplates. Furthermore, black microplates must feature solid black wells or black walls with optical-grade clear flat bottoms if bottom-reading optics are utilized.
Common Misunderstandings and Pitfalls to Avoid
While the PicoGreen assay is robust, routine laboratory execution can suffer from several well-documented technical pitfalls. Recognizing these common misunderstandings ensures consistent assay reproducibility and high analytical rigor.
- Misunderstanding 1: Assuming Absolute Immunity to RNA Interference. While PicoGreen is often described as dsDNA-specific, its 10% cross-reactivity with RNA becomes significant when analyzing tissues rich in total RNA, such as liver, pancreas, or cell culture lysates. If a sample contains 1 ng/µL dsDNA and 15 ng/µL co-extracted RNA, the RNA contributes an apparent 1.5 ng/µL of fluorescence, causing a 150% overestimation of DNA content. Always perform an RNase digestion step when working with unpurified total nucleic acid lysates.
- Misunderstanding 2: Overlooking Rapid Photobleaching Under Ambient Light. Cyanine fluorophores undergo irreversible photodegradation when exposed to bright laboratory fluorescent lights or direct sunlight. Leaving loaded 96-well plates uncovered on the laboratory bench for 20 minutes can reduce fluorescence by 8% to 15%, with outer wells bleaching faster than inner wells. Always shield microplates with aluminum foil or opaque black covers during the 5-minute incubation period.
- Misunderstanding 3: Extrapolating Beyond the Dynamic Range Ceiling. At DNA concentrations exceeding 1,000 ng/mL in the high-range assay, the standard curve deviates from linearity due to inner filter effects (self-absorption of excitation light) and complete depletion of free fluorophore. Calculating concentrations of concentrated samples that yield RFU values above the highest standard produces substantial errors. Samples exceeding 1,000 ng/mL must be diluted in 1X TE buffer and re-assayed.
- Misunderstanding 4: Ignoring Microplate Edge Effects and Meniscus Distortions. Evaporation across peripheral wells (rows A and H, columns 1 and 12) during extended incubations creates optical meniscus curvature that alters fluorescence excitation paths. To mitigate edge effects, ensure plates are read immediately following the 5-minute room-temperature incubation and avoid extended shaking steps at elevated temperatures.
- Misunderstanding 5: Using Incompatible Detergent Concentrations. Sodium dodecyl sulfate (SDS) carried over from lysis buffers is a potent inhibitor of PicoGreen binding. An SDS concentration as low as 0.05% completely eliminates fluorescence enhancement by partitioning the dye into detergent micelles. If SDS cannot be avoided in sample extraction, ensure samples are diluted at least 10-fold to 50-fold into 1X TE buffer so that final well SDS concentrations remain strictly below 0.01%.
The troubleshooting table below highlights key observational symptoms encountered during microplate fluorometry, their root causes, and corrective actions:
| Observed Symptom | Likely Root Cause | Recommended Corrective Action |
|---|---|---|
| High background fluorescence in reagent blank wells | Contaminated TE buffer, dirty microplate wells, or dye precipitation | Prepare fresh 1X TE buffer using nuclease-free water; ensure microplates are dust-free; centrifuge diluted dye briefly if particulates are suspected. |
| Severe fluorescence loss across all sample wells | SDS contamination (exceeding 0.01%), acidic buffer pH (below 7.0), or photobleaching | Confirm TE buffer pH is 7.5 to 8.0; dilute samples to lower residual detergent below 0.01%; verify plate was shielded from light during incubation. |
| Non-linear standard curve (R² below 0.99) | Serial dilution pipetting errors or dye depletion at top concentrations | Prepare fresh standard curve using calibrated pipettes; verify lambda DNA stock concentration; ensure highest standard does not exceed 1,000 ng/mL. |
| Significant duplicate well variance (exceeding 10% CV) | Incomplete mixing, air bubbles in wells, or dye binding to glass tips | Centrifuge microplate at 500 × g for 1 minute to pop bubbles; use virgin polypropylene tips; mix plate for 30 seconds on orbital shaker prior to read. |
| Fluorescence signal drifting downward over time | Continuous excitation light exposure during repeated reads | Minimize reading cycles; configure plate reader with minimal flash counts per well; maintain consistent read timing across experimental plates. |
Frequently asked questions
Can the PicoGreen assay distinguish between fragmented and intact genomic DNA?
No. PicoGreen binds double-stranded DNA by intercalation and minor groove interaction regardless of overall fragment length, provided the fragments are at least 20 to 30 base pairs in length. Consequently, a solution containing 10 ng/mL of highly fragmented 150-base-pair apoptotic or sonicated DNA produces approximately the same fluorescence intensity as 10 ng/mL of high-molecular-weight intact genomic DNA. To assess fragment size distribution, laboratories must complement PicoGreen quantification with automated capillary electrophoresis or agarose gel electrophoresis.
How does PicoGreen compare to NanoDrop absorbance at 260 nm?
PicoGreen fluorometry offers roughly 400-fold to 1,000-fold greater sensitivity than microvolume UV-Vis spectrophotometers such as the NanoDrop. NanoDrop instruments measure total absorbance at 260 nm according to the Beer-Lambert law, which cannot distinguish dsDNA from single-stranded DNA, RNA, free dNTPs, or chemical contaminants like phenol and salts. PicoGreen selectively measures dsDNA and remains accurate down to 25 to 250 pg/mL, whereas NanoDrop measurements below 5 to 10 ng/µL carry high coefficients of variation.
Why must PicoGreen assays be prepared in plastic containers rather than glass?
PicoGreen is an unsymmetrical cyanine dye with hydrophobic aromatic ring systems and a positive charge. Untreated borosilicate glass surfaces contain negatively charged silanol groups that electrostatically attract and adsorb the dye molecules out of solution. Storing working reagent or conducting assays in glass tubes can deplete active fluorophore by up to 40% within minutes, causing severe underestimation of DNA concentration. Always utilize virgin polypropylene tubes and polystyrene microplates.
What is the difference between PicoGreen and RiboGreen assays?
PicoGreen is optimized specifically for quantifying double-stranded DNA and exhibits less than 10% cross-reactivity with RNA. In contrast, RiboGreen is optimized for quantifying RNA and detects single-stranded and total RNA down to 1 ng/mL. However, RiboGreen also binds double-stranded DNA with high affinity. Therefore, while PicoGreen can measure dsDNA directly in the presence of moderate RNA, RiboGreen requires enzymatic DNase digestion to eliminate DNA before accurate RNA quantification can occur.
How does detergent carryover such as SDS affect PicoGreen fluorescence?
Sodium dodecyl sulfate (SDS) is exceptionally detrimental to the PicoGreen assay. At concentrations above 0.01% (w/v), SDS molecules form hydrophobic micelles that competitively sequester the free PicoGreen cyanine dye away from DNA binding sites. This micellar encapsulation extinguishes the fluorescence enhancement mechanism, resulting in false-negative or drastically suppressed concentration readings. Samples extracted using SDS must be washed thoroughly or diluted extensively into 1X TE buffer before assaying.
Can PicoGreen be measured on a standard qPCR machine?
Yes. Because the excitation (~480 nm to 502 nm) and emission (~520 nm to 523 nm) spectra of PicoGreen closely match those of FAM and SYBR Green I, any standard real-time PCR thermal cycler can read PicoGreen fluorescence. Users can set up an isothermal plate-read protocol (e.g., 25 °C for 1 minute with optical acquisition on the FAM channel) to measure standard curves and sample plates when a dedicated fluorescence microplate reader is unavailable.
Sources
- Thermo Fisher Scientific. (2020). Quant-iT PicoGreen dsDNA Reagent and Kits User Guide (Pub. No. MAN0001931, Rev. A.0 / MP 07581). Checked 2026-11-14. https://assets.thermofisher.com/TFS-Assets/LSG/manuals/mp07581.pdf
- Singer, V. L., Jones, L. J., Yue, S. T., & Haugland, R. P. (1997). Characterization of PicoGreen reagent and development of a fluorescence-based solution assay for double-stranded DNA quantitation. Analytical Biochemistry, 249(2), 228–238. Checked 2026-11-14. https://doi.org/10.1006/abio.1997.2177
- Thermo Fisher Scientific. (2021). Quant-iT RiboGreen RNA Reagent and Kit User Guide (Pub. No. MAN0001977, Rev. A.0 / MP 11490). Checked 2026-11-14. https://assets.thermofisher.com/TFS-Assets/LSG/manuals/mp11490.pdf
- Jones, L. J., Yue, S. T., Cheung, C. Y., & Singer, V. L. (1998). RNA quantitation by fluorescence-based solution assay: RiboGreen reagent characterization. Analytical Biochemistry, 265(2), 368–374. Checked 2026-11-14. https://doi.org/10.1006/abio.1998.2914
- Molecular Devices. (2022). DNA and RNA absorbance and fluorescence measurements using SpectraMax microplate readers (Application Note 012-0056). Checked 2026-11-14. https://www.moleculardevices.com/applications/nucleic-acid-quantitation
Data: Nucleic Acid Quantification Fluorophores and Optical Benchmarks
Comparative specifications of fluorometric dyes and optical methods for nucleic acid quantification compiled from manufacturer manuals and peer-reviewed characterization studies.
Period covered: 1997/2026. Coverage: Global.
| Reagent or Method | Target Nucleic Acid | Excitation Peak (nm) | Emission Peak (nm) | Lower Detection Limit (pg/mL) | Upper Linear Range (ng/mL) | Fluorescence Enhancement (-fold) |
|---|---|---|---|---|---|---|
| Quant-iT PicoGreen dsDNA (Low Range) | Double-stranded DNA (dsDNA) | 502 | 523 | 25 | 25 | 1000 |
| Quant-iT PicoGreen dsDNA (High Range) | Double-stranded DNA (dsDNA) | 502 | 523 | 250 | 1000 | 1000 |
| Quant-iT RiboGreen RNA (Low Range) | Total / Single-stranded RNA | 500 | 525 | 1000 | 50 | 1000 |
| Quant-iT RiboGreen RNA (High Range) | Total / Single-stranded RNA | 500 | 525 | 20000 | 1000 | 1000 |
| Hoechst 33258 Bisbenzimide Dye | Double-stranded DNA (AT-rich) | 352 | 461 | 10000 | 15000 | 30 |
Sources
- Thermo Fisher Scientific Quant-iT PicoGreen dsDNA Reagent User Guide MAN0001931 (accessed 2026-11-14)
- Singer et al. Characterization of PicoGreen reagent, Anal Biochem 1997 (accessed 2026-11-14)