Skip to main content
v2026.11,986 entries · CC-BY 4.0

NanoDrop 2000: Specifications, Protein Measurement and Status

NanoDrop 2000 explained: how pedestal measurement works, protein A280 vs colorimetric ranges, 2000 vs 2000c, pedestal care and the 2028 service date.

Written and maintained by CASRAI Editorial Board

Published

On this page

The NanoDrop 2000 is a Thermo Scientific microvolume UV-Vis spectrophotometer that measures nucleic acid and protein concentration on a 0.5 to 2 µL droplet held by surface tension between two optical fibers, with no cuvette needed on the standard model. Thermo Fisher stopped selling the 2000 and 2000c on 24 September 2021 and has committed to servicing them through 23 September 2028, so labs meet it mostly as an installed instrument rather than a new purchase.

This page is a research-practice reference for people who use, document or inventory the instrument: what it measures, how its protein modes differ, where the 2000 and 2000c differ, what the manual says about care, and what the end-of-sale notice means for a lab. It is general instrument information, not a replacement for the user manual or your own SOP.

What the NanoDrop 2000 is and how it measures

The NanoDrop 2000 and the cuvette-capable 2000c are covered by one Thermo user manual. According to that manual, the instrument combines sample retention technology with a traditional cuvette option on the 2000c. In pedestal mode, a 1 to 2 µL sample is pipetted onto a measurement pedestal that has a fiber optic cable (the receiving fiber) embedded in it. A second fiber (the source fiber) is brought into contact with the liquid so that it bridges the gap between the two fiber ends. A pulsed xenon flash lamp provides the light, and a spectrometer with a linear CCD array analyses the light that passes through the sample. The instrument is run from software on a connected PC, and results are stored in workbook files.

The manual lists a 2048-element linear silicon CCD detector and a 190 to 840 nm wavelength range. Pedestal pathlength is 1 mm, auto-ranging down to 0.05 mm for concentrated samples. Because the pathlength is that short, the manual states the system can measure samples up to 200 times more concentrated than a standard cuvette allows without dilution. The software reports absorbance normalised to a 10 mm (1 cm) path, so numbers can be compared with conventional spectrophotometer readings. The sample retention technology is covered by US patents 6,628,382 and 6,809,826, per the same manual.

Specifications stated in the manual

  • Minimum sample size 0.5 µL; 1 to 2 µL is the normal working volume.
  • Absorbance precision 0.002 (1 mm path); absorbance range 0.02 to 300 on a 10 mm equivalent basis.
  • dsDNA detection limit 2 ng/µL on the pedestal, with a maximum of 15,000 ng/µL.
  • Spectral resolution below 1.8 nm (FWHM at the 253.7 nm mercury line).
  • Pedestal material: 303 stainless steel and quartz fiber. Footprint 14 cm × 20 cm, weight 2.0 kg, 12 VDC supply.

The Thermo product page lists the same sample volume (0.5 to 2.0 µL), wavelength range (190 to 840 nm) and dsDNA range (2 to 15,000 ng/µL), which confirms the manual figures still match the manufacturer’s current description of the product.

NanoDrop 2000 vs 2000c: pedestal and cuvette mode

The two models share the pedestal. The difference is that only the 2000c accepts cuvettes: the manual states the 2000c takes 10 mm cuvettes up to 48 mm tall and adds a cuvette mode with heating and stirring. Several software options, such as Use Cuvette, are labelled in the manual as 2000c only. The table below sets the two measurement modes side by side, using only values printed in the manual.

Item Pedestal mode (2000 and 2000c) Cuvette mode (2000c only)
Sample volume 0.5 to 2 µL (1 to 2 µL typical) Enough to cover the light path; the manual refers to the cuvette maker’s volume recommendations
Pathlength 1 mm, auto-ranging to 0.05 mm 10, 5, 2 or 1 mm
dsDNA detection limit 2 ng/µL 0.4 ng/µL
dsDNA maximum 15,000 ng/µL 750 ng/µL
Measurement time Under 5 seconds Under 3 seconds
Extras None Cuvette holder heating to 37 ± 0.5 °C; stirrer 150 to 850 RPM; beam 8.5 mm above the cuvette base

For a lab deciding which mode suits a sample, the pattern in these numbers is simple: the pedestal handles high-concentration, low-volume samples, while the cuvette mode reaches lower concentrations because the light path is longer. The manual also advises quartz cuvettes for analytical wavelengths below 340 nm, because most plastic and glass cuvettes block UV, and recommends masked cuvettes when using micro-volume cuvettes so that all detected light has passed through the sample.

Measuring protein concentration on the NanoDrop 2000

For a protein concentration question, the manual offers several applications, and choosing the right one is the main decision. The Protein A280 application reads the protein’s own absorbance at 280 nm. The manual says it applies to purified proteins that contain Trp or Tyr residues, or Cys-Cys disulphide bonds, and that it needs no standard curve. The default setting is 1 Abs = 1 mg/mL, based on a 0.1% protein solution giving an absorbance of 1.0 at 280 nm over a 10 mm path. Other reference options use mass extinction coefficients for a 1% (10 mg/mL) solution: 6.7 for BSA, 13.7 for IgG and 26.4 for lysozyme, with a user-entered option for any other protein.

The manual is explicit about the limit of this method: colorimetric assays such as BCA, Pierce 660 nm, Bradford and Lowry are more commonly used for uncharacterised protein solutions and cell lysates. The NanoDrop software includes pre-programmed colorimetric methods for labs that already use those assays. Unlike A280, the colorimetric applications require a standard curve to be generated before sample concentrations can be read, and the manual recommends following the assay manufacturer’s guidance for building that curve. The Proteins and Labels application extends the A280 approach to fluorescent dye concentration for protein array conjugates and can also be used to assess metalloprotein purity.

The data table below compares the protein measurement applications on the quantitative ranges the manual prints for them. Ranges are expressed against the manual’s reference standards (purified BSA for the UV methods), so a real sample will behave differently if its extinction coefficient or composition departs from the reference.

Small-volume option and blanking

For concentrated samples, the manual describes a small sample volume option in the Nucleic Acid and Protein A280 applications that allows measurements on just 0.5 µL when the 10 mm equivalent absorbance is 3.0 or higher (for example above 150 ng/µL of dsDNA). The blank should be the same buffer as the sample, at similar pH and ionic strength, and the manual recommends taking a new blank about every 30 minutes when measuring many samples. A University of Maryland instrument SOP for the NanoDrop 2000 adds, from field experience, that 1 µL is enough for aqueous nucleic acid solutions while 2 µL is used for purified protein and for Bradford, BCA, Lowry or Pierce 660 nm assays.

Nucleic acid readings and purity ratios

For nucleic acids the instrument applies the Beer-Lambert relationship with the manual’s constants: 50 ng-cm/µL for double-stranded DNA, 33 for single-stranded DNA and 40 for RNA. The software reports the 260/280 ratio, which the manual says is generally accepted as about 1.8 for pure DNA and about 2.0 for pure RNA. An appreciably lower value may indicate protein, phenol or other contaminants that absorb strongly at or near 280 nm. The 260/230 ratio is a secondary purity measure, commonly in the range 1.8 to 2.2 for pure nucleic acid, where a lower value may indicate co-purified contaminants. A default baseline correction at 340 nm is available.

These ratios are indicators, not a verdict on sample quality. Absorbance cannot tell DNA from RNA or from free nucleotides, which is why absorbance readings are normally paired with other checks when downstream work is sensitive. If absorbance is the wrong tool for the question, for example for dilute or contaminated samples where a dye-based method would be more specific, see the CASRAI comparison Qubit vs NanoDrop and the guides to the Qubit fluorometer and the PicoGreen assay.

Where it matters for labs: care, records and lifecycle

Pedestal care

Between samples, the manual says to raise the arm and wipe the sample from both the upper and lower pedestals with a dry, lint-free laboratory wipe, and states that simple wiping prevents carryover between samples varying by more than 1000-fold in concentration. It also warns that solutions and reagents containing surfactants may “un-condition” the pedestal surfaces so that the liquid column does not form. When columns break during measurement, the manual points to the NanoDrop Pedestal Reconditioning Compound (PR-1). Because additives such as protein, DNA, RNA, buffer salts and detergent-like molecules lower surface tension, the manual suggests increasing the sample to 2 µL for such samples and recommends a precision 0 to 2 µL pipettor with precision tips to deliver 1 to 2 µL accurately.

Lifecycle and asset registers

Thermo’s notice states that sales of the NanoDrop 2000/2000c ended on 24 September 2021, with service committed globally through 23 September 2028, and names the NanoDrop One/OneC as the instrument that can take over quantification and assessment of DNA, RNA and proteins. The current Thermo product page for the 2000 marks new sales as discontinued and points to the NanoDrop Ultra as the newest model. The notice also lists a NanoDrop One/OneC wavelength range of 190 to 850 nm and a dsDNA range of 2.0 to 27,500 ng/µL, against 190 to 840 nm and 2.0 to 15,000 ng/µL for the 2000.

For a lab manager, the practical consequence is editorial rather than regulatory: a service end date is a natural field in an equipment record, and the instrument’s PC software requirements matter. The manual lists Windows XP and Vista (32-bit), while the product page lists Windows 7, 8 and 10 Pro, so the control PC is part of the asset. For regulated work, Thermo’s notice says its optional Security Suite for the One/OneC can help meet Title 21 CFR Part 11 requirements for electronic records; the notice makes no equivalent statement for the 2000.

Common misunderstandings

  • The NanoDrop 2000 takes cuvettes. Only the 2000c does, per the manual; the standard 2000 is pedestal-only.
  • A280 works for any protein sample. The manual limits it to purified proteins with Trp, Tyr or Cys-Cys, and points to colorimetric assays for uncharacterised solutions and cell lysates.
  • A good 260/280 means a clean sample. The manual treats the ratios as indicators of possible contamination, with reference values of about 1.8 for DNA and about 2.0 for RNA.
  • Any blank will do. The blank should match the sample buffer in pH and ionic strength, and a fresh one is recommended about every 30 minutes in long runs.
  • Droplet problems mean a faulty instrument. Per the manual, surfactant-containing solutions can un-condition the pedestals; reconditioning is a documented step.
  • A discontinued instrument is an unsupported one. Thermo has committed to servicing the 2000 and 2000c through 23 September 2028.

For the broader instrument class, see what a microvolume spectrophotometer is and choosing a spectrophotometer for DNA concentration.

Sources

Data: NanoDrop 2000 protein measurement applications: quantitative ranges and intended samples

Compares the protein quantitation applications of the Thermo NanoDrop 2000/2000c software on lower and upper limits, typical reproducibility and intended sample type, compiled from the measurement-ranges table and application chapters of the Thermo user manual V1.0.

Period covered: Manual revision March 2009.

NanoDrop 2000 protein measurement applications: quantitative ranges and intended samples
ApplicationLower limit (mg/mL)Upper limit (mg/mL)Typical reproducibilityStandard curve neededIntended sample (manual)
Protein A280 (pedestal)0.14000.10-10 mg/mL: ±0.10 mg/mL; above 10 mg/mL: ±2%NoPurified proteins with Trp, Tyr or Cys-Cys; limits stated as purified BSA
Proteins and Labels (pedestal)0.1200.10-10 mg/mL: ±0.10 mg/mLNoProtein A280 plus fluorescent dye concentration for protein array conjugates
BCA (20:1 reagent to sample)0.282% over entire rangeYesColorimetric assay for uncharacterised solutions and cell lysates
Modified Lowry0.242% over entire rangeYesColorimetric assay for uncharacterised solutions and cell lysates
Bradford (50:1 reagent to sample)0.18100-500 µg/mL: ±25 µg/mL; 500-8000 µg/mL: ±5%YesColorimetric assay for uncharacterised solutions and cell lysates

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about NanoDrop 2000: Specifications, Protein Measurement and Status

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →