Oxygen Regulator: What It Is, Types and General Use

What an oxygen regulator does, how Thorpe, Bourdon and dial flow controls differ, and the WHO-UNICEF pressure and flow figures behind them.

Written and maintained by CASRAI Editorial Board

Last updated

An oxygen regulator is the device between a compressed oxygen cylinder and the equipment that delivers oxygen to a patient. It lowers the high cylinder pressure to a low, usable outlet pressure and, on many models, also sets or shows the flow. This page describes products and general use; it is not medical advice.

Who uses it: clinic and hospital staff who set up cylinder oxygen, biomedical and procurement teams who specify equipment, and students learning medical gas systems. The figures below come from the WHO-UNICEF technical specifications for oxygen therapy devices and from the scope of the ISO 10524-1 standard, both linked in the Sources section at the end.

What an oxygen regulator is

An oxygen regulator (also called a pressure regulator or pressure-reducing valve) is the pressure-control component of a cylinder oxygen set-up. A full cylinder holds oxygen at very high pressure; the regulator reduces that to a fixed, much lower pressure that downstream tubing and flow devices are built to handle. ISO 10524-1:2019, the international standard for these devices, covers pressure regulators and pressure regulators with flow-metering devices intended for administering medical gases and mixtures such as oxygen, medical air and oxygen/nitrous oxide mixtures.

Some cylinders have a built-in integral regulator, so no separate regulator is needed.

How an oxygen regulator is built

The WHO-UNICEF specification for the cylinder pressure regulator assembly describes these parts and features:

  • Inlet connection: the regulator must be compatible with the cylinder’s primary valve, either a pin index or a bullnose type.
  • Body and valve materials: casing of steel, plated brass or aluminium, with a brass valve.
  • Pressure gauge: an integrated manometer reading 0 to 20 000 kPa, so the user can see cylinder pressure.
  • Outlet: available with a DISS connector or a 6 mm barbed outlet. DISS connectors are threaded, with a diameter unique to each gas.
  • Flow control (optional): a basic open/close model, or one with an integrated Thorpe tube or Bourdon flowmeter.

Versions exist for oxygen and for medical air, and in standard and MRI-compatible forms.

The Pin Index Safety System is a key safeguard: the arrangement of the pins is unique for each gas, to prevent the use of the wrong gas. For the cylinder side of the connection, see CASRAI’s entries on CGA fittings and compressed gas cylinder sizes.

Types and variants: how the flow control differs

The main difference between regulator models is how flow is set and shown. WHO-UNICEF compares three flowmeter types (its Table 4.1), and the choice decides whether a separate pressure-reducing valve is needed at all.

Flowmeter type How it works Pressure-reducing valve Orientation and accuracy notes
Thorpe tube Variable orifice; a float moves in a clear upright tube Requires a separate pressure-reducing valve Must be kept vertical; works from a 345 kPa source only
Bourdon gauge Fixed orifice Integrates a pressure-reducing valve Unaffected by gravity and usable at any angle (emergency and transport use); not back-pressure compensated; less accurate at low flows
Dial / click-type Fixed orifice Integrates a pressure-reducing valve Regulates flow but does not measure it; accurate only at its rated pressure (345 kPa)

WHO-UNICEF also notes that a preset regulator with one low-pressure gauge is recommended as the first choice in low-resource settings, while an adjustable regulator with a double gauge (high and low pressure) gives more options but adds complexity.

Specifications that matter

Specification WHO-UNICEF value
Nominal inlet pressure 13 700 kPa (137 bar, 1987 psi)
Maximum inlet pressure 20 000 kPa (200 bar, 2901 psi)
Outlet pressure 345 kPa (3.45 bar, 50 psi)
Integrated gauge range 0 to 20 000 kPa
Standard cylinder nominal pressure 13 700 kPa
Integral-valve cylinder pressure 23 000 to 30 000 kPa
Low flow range (oxygen only) 0 to 3 or 4 L/min, accuracy 10%
Standard flow ranges 0 to 7/8 and 0 to 15/16 L/min (Thorpe or Bourdon), accuracy 10%
High flow range At least 0 to 25 L/min, accuracy 10%
Dial-style ranges Low 0 to 3/4, standard 0 to 15, high 0 to 25 L/min, accuracy 10%

How to choose: criteria stated in the specification

WHO-UNICEF frames selection around items to specify, not a ranking of products:

  • Gas and connector: specify the compatible connector when procuring. DISS connectors are gas-specific, and the cylinder valve must match the regulator inlet (pin index or bullnose).
  • Flow device: decide whether a basic regulator, or one with a Thorpe or Bourdon flowmeter, fits the use. Thorpe tubes need a separate pressure-reducing valve and a vertical position; Bourdon gauges can be used at any angle.
  • Flow range: pick the low, standard or high range tier from the table above.
  • Standard compliance: the specification refers to ISO 10524 for pressure regulators used with medical gases.
  • Warranty: WHO-UNICEF recommends a 3-year warranty on regulators and valves, and at least 2 years.
  • Total cost: the pressure-reducing valve is a required accessory for some set-ups and belongs in the cost of the whole system.

Safe use and handling basics

  • Protect the cylinder and all other fittings from contamination by oil and grease.
  • Never use lubricants, because they are flammable.
  • Keep a Thorpe-tube flowmeter vertical. Read the float at the centre when it is a ball, and from the top when it is a bobbin.
  • Disconnect the device before cleaning, and clean it according to the manufacturer’s instructions and your infection prevention and control protocol.
  • Use only connectors made for oxygen; the pin index and DISS systems exist so the wrong gas cannot be connected.

For cylinder handling, storage and identification, see compressed gas cylinder safety and compressed gas cylinder colour codes.

Common mistakes

  • Lubricating a fitting. The specification warns against any lubricant on oxygen equipment.
  • Using a Thorpe tube off-vertical. It must be vertical to read correctly.
  • Assuming a dial or click-type device measures flow. It regulates flow but does not measure it, and is accurate only at its rated 345 kPa.
  • Buying a Thorpe flowmeter without a pressure-reducing valve. It requires one.
  • Not specifying the connector. DISS threads differ for each gas, so state the compatible connector when procuring.

Frequently asked questions

What does an oxygen regulator do?

It reduces the high pressure inside an oxygen cylinder to a low outlet pressure (345 kPa, or 50 psi, in the WHO-UNICEF specification), and some models also control or show the flow.

Is a regulator the same as a flowmeter?

No. A regulator controls pressure; a flowmeter sets or shows flow. A Thorpe-tube flowmeter needs a separate pressure-reducing valve, while Bourdon and dial types integrate one. Some regulators come with a flowmeter attached.

What standard covers oxygen regulators?

BS EN ISO 10524-1:2019, published 28 February 2019, covers design, construction, type testing and marking requirements for pressure regulators for medical gases, including those with integral flow-metering devices.

Why can’t a regulator from another gas be used on oxygen?

Connections are gas-specific. The pin index arrangement is unique for each gas, and DISS connectors have a unique diameter per gas, which prevents the wrong gas being connected.

Where do regulators fit with other oxygen supplies?

The regulator feeds a delivery device such as a nasal cannula; see the nasal cannula flow rate and sizing guide. Oxygen masks are listed in the CPAP and oxygen masks category.

Sources

  • WHO and UNICEF, Technical specifications and guidance for oxygen therapy devices, WHO Medical device technical series, 14 January 2019, ISBN 9789241516914: who.int/publications/i/item/9789241516914 (checked 29 October 2026).
  • BSI, BS EN ISO 10524-1:2019, Pressure regulators for use with medical gases. Pressure regulators and pressure regulators with flow-metering devices: BSI Knowledge listing (checked 29 October 2026).

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Oxygen Regulator: What It Is, Types and General Use

Ask your first 2 questions free below. Subscribers get 150 a day for $29 a month.

An AI assistant specialized in research administration. It cites the sources behind every answer, labels web answers and says when it can't answer.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays free to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →

Ask CASRAI · Regulatory Radar

Research-admin question? Get an answer that links its sources.

An AI assistant specialized in research administration. Every answer links its sources to check before you act. 2 questions free, no account. $29/month after.

  • Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.
  • Every answer numbers its sources and links each one, so you can check the source yourself.