Skip to main content
v2026.11,986 entries · CC-BY 4.0
LAC HealthSponsored

Shop paper point supplies

  • FDA registered
  • ISO 13485
  • BBB Accredited A+

Shop at LAC Health

Paper Point: What It Is and How the FDA Regulates It

A paper point is an absorbent endodontic device, Class I under 21 CFR 872.3830. See its FDA code, and what the withdrawn ISO 7551 means for labels.

Written and maintained by CASRAI Editorial Board

Published · Last updated

On this page

General product and regulatory information only; it is not medical or dental advice and does not describe a clinical procedure.

Short answer

A paper point is an absorbent paper device used in endodontic (root canal) therapy to dry, or to apply medication to, the root canal of a tooth. In the United States the FDA lists it as an “endodontic paper point” under 21 CFR 872.3830, a Class I device. A product label’s “ISO” claim needs reading with care, because the one ISO standard written for dental absorbent points has been withdrawn.

Definition and scope

The federal regulation defines the device in one sentence: an endodontic paper point is “a device made of paper intended for use during endodontic therapy to dry, or apply medication to, the root canal of a tooth” (21 CFR 872.3830(a), eCFR, checked 2026-10-31). The device sits in the dental devices part of the FDA’s rules, alongside other endodontic items such as the endodontic file.

The FDA product code for the device is EKN, “Point, Paper, Endodontic”. It is a different product from a gutta-percha or other obturating point, and from general-purpose test papers such as nitrazine paper.

How it is used

The regulation names two purposes: drying the root canal and applying medication to it. A manufacturer record in the FDA’s AccessGUDID database describes one product as an absorbent paper point used during endodontic therapy to dry, and to remove foreign material from, the root canal (record 08806383516839, NLM AccessGUDID, checked 2026-10-31). Rinsing in endodontic care is a separate topic; see the guide to the dental irrigation syringe. This page does not cover clinical technique. For another endodontic-adjacent product, see the guide to Endo-Ice.

How the FDA regulates it

Item What the source says Source
Regulation 21 CFR 872.3830, “Endodontic paper point” eCFR, checked 2026-10-31
Classification Class I (general controls) eCFR 872.3830(b)
Premarket notification Exempt from 510(k), subject to the limitations in 21 CFR 872.9 eCFR 872.3830(b)
Product code EKN, Point, Paper, Endodontic AccessGUDID record, checked 2026-10-31

“Exempt from 510(k)” is not the same as unregulated: general controls still apply, and the exemption is limited by 21 CFR 872.9.

What one manufacturer record shows

AccessGUDID entries are written by each manufacturer, so one record is an example and not a rule for the category. The record cited above states that the product is packaged sterile, does not require sterilization before use, is single-use, and is prescription-labelled rather than over-the-counter. The same record lists product code EKN. A different product may declare different packaging or sterility attributes, which is why the record for the exact item matters.

Standards: what an “ISO” claim can and cannot mean

Standard Scope as listed by BSI Status as listed by BSI (checked 2026-10-31)
BS EN ISO 7551:1998, Dental absorbent points Requirements and test methods for nonmedicated absorbent points used in endodontic procedures, including standard and taper sizes. Does not cover sterility or freedom from biological hazard. Published 15 Aug 1997; withdrawn 17 May 2023
BS EN ISO 6877:2021, Endodontic obturating materials Dimensions and numerical and colour-coding systems for obturating points (metal, coated metal, polymeric, thermoplastic). Sterility excluded. The listing does not mention paper points. Published 30 Sep 2021; listed as withdrawn effective 14 Aug 2025

Read together, the two entries mean that a listing which says only “ISO” has not told you which standard, which edition, or whether that edition is still current. They also show that a standard’s scope and a product’s sterility are separate questions: the absorbent-points standard excludes sterility from its scope.

What to read on a listing

  • The regulation or product code: 21 CFR 872.3830 and EKN identify the device category.
  • The standard cited, with its number and year, and whether that edition is current.
  • Whether the item is declared packaged sterile, and whether it is declared single-use (both are separate attributes in a GUDID record).
  • Size and taper markings as printed by the manufacturer; the absorbent-points standard covers standard and taper sizes.

Common misunderstandings

Is a paper point the same as a gutta-percha point?

No. The regulation defines the paper point as a paper device for drying or applying medication. Obturating points are covered by a separate standard, ISO 6877, which does not list paper points.

Does “ISO standard” on a label guarantee a sterile product?

No. ISO 7551 states that sterility is outside its scope. Sterile packaging is declared by the manufacturer, for example in a GUDID record.

Does 510(k) exemption mean no FDA oversight?

No. The device is Class I and subject to general controls; the exemption is limited by 21 CFR 872.9.

FAQ

What is a paper point used for?

Per 21 CFR 872.3830, to dry, or apply medication to, the root canal of a tooth during endodontic therapy.

What class of device is a paper point?

Class I (general controls), exempt from 510(k) subject to the limitations in 21 CFR 872.9.

What is the FDA product code?

EKN, “Point, Paper, Endodontic”.

Which standard covers dental absorbent points?

BS EN ISO 7551:1998, which BSI lists as withdrawn on 17 May 2023.

Sources

  • eCFR, 21 CFR 872.3830, Endodontic paper point: https://www.ecfr.gov/current/title-21/section-872.3830 (checked 2026-10-31)
  • NLM AccessGUDID record 08806383516839: https://accessgudid.nlm.nih.gov/devices/08806383516839 (checked 2026-10-31)
  • BSI Knowledge, BS EN ISO 7551:1998 Dental absorbent points: https://knowledge.bsigroup.com/products/dental-absorbent-points (checked 2026-10-31)
  • BSI Knowledge, BS EN ISO 6877:2021 Endodontic obturating materials: https://knowledge.bsigroup.com/products/dentistry-endodontic-obturating-materials (checked 2026-10-31)

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Paper Point: What It Is and How the FDA Regulates It

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →