Respironics EverFlo: Technical Specs & Guide

A technical guide to the Respironics EverFlo oxygen concentrator: 5 LPM continuous flow, PSA molecular sieve mechanics, FDA standards, and facility planning.

Written and maintained by CASRAI Editorial Board

Published

On this page

Medical device notice: This technical guide is intended for clinical researchers, biomedical engineers, laboratory personnel, and healthcare procurement specialists. It provides equipment specifications, regulatory classifications, and facility engineering guidelines. It does not provide personal medical advice, clinical diagnoses, or patient-specific treatment dosing.

Short Answer

The Respironics EverFlo is a stationary medical oxygen concentrator delivering continuous-flow oxygen from 0.5 to 5.0 liters per minute (LPM) at 93% (±3%) purity. Manufactured by Philips Respironics, the 31-pound unit uses dual-bed pressure swing adsorption (PSA) molecular sieves to extract oxygen from ambient air, drawing an average of 350 watts of electrical power.

Definition and Scope

The Respironics EverFlo is regulated by the U.S. Food and Drug Administration (FDA) as a Class II medical device under 21 CFR § 868.5440 (Portable Oxygen Generator / Oxygen Concentrator, Product Code CAW), cleared under 510(k) premarket notification K061261. In clinical reimbursement frameworks, including Medicare Part B Durable Medical Equipment (DME), the device is billed under Healthcare Common Procedure Coding System (HCPCS) code E1390, defined as a stationary oxygen concentrator with a single delivery port providing 85% or greater oxygen purity at prescribed operational flow rates.

In healthcare and research facilities, the EverFlo operates as a dedicated point-of-use gas separation appliance. Rather than depending on central gas pipelines or high-pressure cylinders, it supplies continuous oxygen for recovery suites, outpatient clinics, veterinary surgery, and vivarium animal research. While hospital facilities depend on central infrastructure—as detailed in our analysis of hospital oxygen plant architectures—compact stationary units supply point-of-use gas without cryogenic storage hazards or cylinder logistics.

How It Works and How It Is Used in Practice

The EverFlo operates on Pressure Swing Adsorption (PSA), physically isolating oxygen from ambient room air without chemical consumables. Ambient air contains roughly 78% nitrogen, 21% oxygen, and 1% argon and trace gases. The EverFlo separates these gases through an engineered cycle:

  1. Intake Filtration: Room air passes through an acoustic intake resonator and an intake filter, capturing airborne particulates.
  2. Compression: An oil-free compressor pressurizes air to 20–30 PSI and feeds a motorized four-way rotary valve.
  3. Zeolite Sieve Adsorption: Air enters one of two aluminum canisters containing synthetic zeolite molecular sieve beads. Under pressure, the zeolite selectively adsorbs nitrogen molecules, allowing oxygen and argon to flow into an internal surge tank.
  4. Regeneration Purge: Before the active bed saturates, the valve diverts flow to the second sieve canister. The first canister depressurizes to atmospheric pressure, venting adsorbed nitrogen through a rear exhaust port while receiving a small reverse purge of concentrated oxygen.
  5. Delivery Metering: Concentrated gas passes through a bacterial micro-disk filter, a pressure regulator delivering 5.5 PSI (38 kPa), and a front-panel flowmeter calibrated from 0.5 to 5.0 LPM.

The EverFlo consumes an average of 350 watts at 120 VAC / 60 Hz. Facility managers calculating utility loads can review our guide on how much electricity an oxygen concentrator uses across multi-unit deployments. Models equipped with an Oxygen Percentage Indicator (OPI) utilize an ultrasonic sensor: purity above 82% illuminates green, levels between 70% and 82% illuminate yellow for service, and purity below 70% triggers an audible alarm and red warning light.

Understanding where the EverFlo fits within respiratory supply chains requires comparing stationary concentrators with high-flow devices, portable units, and compressed cylinders:

Equipment Category Flow Range Purity Weight Power Draw Primary Clinical & Lab Role
Standard EverFlo 0.5–5.0 LPM 93% ±3% 31 lbs (14 kg) ~350 W Routine continuous oxygen therapy, bench research, animal care
EverFlo Q 0.5–5.0 LPM 93% ±3% 31 lbs (14 kg) ~350 W Acoustically damped sleep labs and noise-sensitive clinics (40 dBA)
High-Flow Units (M10) 1.0–10.0 LPM 92% ±4% 53 lbs (24 kg) ~600 W High-demand clinical suites and multi-station laboratory manifolds
Portable Concentrators 1–5 Pulse Settings 90% ±3% 8.5–10 lbs Battery / 150 W Ambulatory transport, field research, mobile emergency kits
Compressed Gas Tanks 0.5–15+ LPM 99.5%+ USP 15–115+ lbs 0 W Emergency grid backup, anesthesia delivery, zero-power settings

When modeling equipment life-cycle costs, supply officers consult our comparison of oxygen concentrators versus oxygen tanks to evaluate capital investment against recurring cylinder replenishment. For clinical applications requiring airway moisture, clinicians attach an external bubble bottle via the cabinet strap, detailed in our guide on oxygen concentrator humidifier bottles and care.

Where It Matters for Labs, Clinics, and Supply Decisions

Biomedical engineers, research managers, and supply officers evaluate several criteria when procuring and deploying EverFlo units:

  • Electrical and Thermal Dissipation: With a continuous 350-watt draw, each operating unit rejects roughly 1,194 BTU/hr of heat. Branch circuits must prevent breaker trips in multi-unit rooms, and HVAC systems must maintain ambient temperatures between 55°F and 90°F (12°C to 32°C).
  • Room Air Ventilation: Concentrators exhaust nitrogen while capturing oxygen. Operating in enclosed, airtight storage spaces alters room gas balances; continuous room air turnover is required.
  • Elevation Limits: The EverFlo operating envelope covers 0 to 7,500 feet (0 to 2,286 meters). At higher altitudes, reduced barometric pressure decreases compressor mass intake, lowering oxygen concentration unless high-altitude compensation is utilized.
  • Preventive Maintenance: The EverFlo features an internal compressor intake filter with a 2-year service interval, eliminating weekly washing of external foam filters common on older equipment.
  • Emergency Power Redundancy: Because concentrators cease delivery during power outages, facilities housing oxygen-dependent subjects must maintain backup high-pressure cylinders. For general equipment principles, see our explainer on what an oxygen concentrator is.

Common Misunderstandings

Addressing operational misconceptions prevents procurement and clinical handling mistakes:

  • Concentrator Purity vs. Cryogenic Gas: Concentrators do not produce 100% pure oxygen. PSA systems concentrate atmospheric argon alongside oxygen, producing 93% (±3%) oxygen and 4% to 5% argon. This meets FDA and USP 93 Percent Oxygen standards, but differs from 99.5%+ cryogenic liquid oxygen.
  • Operating in Confined Closets: Placing units in unventilated closets to muffle sound causes heat accumulation and compressor thermal tripping. Units require at least 6 to 12 inches of perimeter clearance.
  • External Filter Washing: The EverFlo houses an internal intake filter behind a rear panel for two-year technician replacement, eliminating weekly cabinet washing.
  • Flowmeter Ball Alignment: To set gas flow accurately, operators align the center of the float ball with the desired LPM calibration line, rather than the top or bottom edge.

Frequently Asked Questions

What is the HCPCS billing code for the Respironics EverFlo?

The Respironics EverFlo is billed under HCPCS code E1390, designating a stationary oxygen concentrator with a single delivery port providing 85% or greater oxygen purity at prescribed flow rates.

What flow rate and oxygen purity does the Respironics EverFlo deliver?

The system provides continuous-flow oxygen from 0.5 to 5.0 LPM at an oxygen purity of 93% (±3%) across all flow settings under standard operating conditions.

How much electrical power does the Respironics EverFlo use?

The unit draws an average of 350 watts during standard continuous operation on a 120 VAC / 60 Hz electrical circuit.

What does the Oxygen Percentage Indicator (OPI) signify?

The OPI continuously monitors purity: a green LED confirms normal output (>82% purity), a yellow LED indicates low purity (70%–82%) requiring service, and a red LED with an audible alarm indicates critically low purity (<70%).

What are the dimensions and weight of the Respironics EverFlo?

The EverFlo measures 23 in high, 15 in wide, and 9.5 in deep (584 mm × 381 mm × 241 mm) and weighs 31 pounds (14 kg), with a molded handle and four casters.

What is the difference between the standard EverFlo and the EverFlo Q?

The standard EverFlo operates at approximately 45 dBA, while the EverFlo Q incorporates acoustic baffling to reduce sound output to 40 dBA for noise-sensitive clinics and sleep labs.

Sources

  1. U.S. Food and Drug Administration (FDA). 510(k) Premarket Notification K061261: L4 Oxygen Concentrator (EverFlo). Applicant: Respironics, Inc. Cleared June 28, 2006. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K061261 (accessed 2026-11-09).
  2. U.S. Code of Federal Regulations. Title 21 § 868.5440: Portable Oxygen Generator / Oxygen Concentrator (Class II, Product Code CAW). Legal Information Institute, Cornell Law School. https://www.law.cornell.edu/cfr/text/21/868.5440 (accessed 2026-11-09).
  3. Centers for Medicare & Medicaid Services (CMS). Durable Medical Equipment, Prosthetics, Orthotics, and Supplies (DMEPOS) Fee Schedule: HCPCS Code E1390. https://www.cms.gov/medicare/payment/fee-schedules/dmepos-fee-schedule (accessed 2026-11-09).
  4. Philips Respironics. EverFlo and EverFlo Q User and Service Reference Manual. Murrysville, PA: Philips Respironics. https://www.documents.philips.com (accessed 2026-11-09).
  5. International Organization for Standardization (ISO). ISO 80601-2-69:2020: Medical electrical equipment — Part 2-69: Particular requirements for the basic safety and essential performance of oxygen concentrator equipment. https://www.iso.org/standard/72552.html (accessed 2026-11-09).

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Respironics EverFlo: Technical Specs & Guide

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo