Centrifuge bottles are large-volume, capped containers, typically 250 mL to 1 L, that hold a sample while a centrifuge spins it. They are made from a rigid plastic such as polycarbonate (PC) or polypropylene copolymer (PPCO), and each model carries a manufacturer-stated maximum relative centrifugal force (RCF, in × g) that depends on the rotor it is run in. They are the next size up from centrifuge tubes, and they are built to be filled, capped, balanced and spun in high-capacity rotors.
What centrifuge bottles are
A centrifuge bottle is a screw-capped, wide-bodied vessel sized for a centrifuge rotor that takes bottles instead of small tubes. Manufacturer product records describe the class by material and closure: a clear polycarbonate bottle with a polypropylene screw cap, a polypropylene copolymer bottle with a sealing closure, or a high-performance bottle with an aluminum cap. Search results also use the phrases “bottle centrifuge” and “centrifuge bottle” for the same thing. A “bottle centrifuge” in everyday use is simply a centrifuge whose rotors accept bottles, and a “centrifuge bottle” is the container itself. This page uses the second term.
Who uses them: lab staff who spin volumes of hundreds of millilitres in one run. If your protocol spins volumes of a few millilitres, you are in tube territory (see below). To place bottles in the wider picture of instruments, start with CASRAI’s guide to what a centrifuge is and its explanation of the centrifuge rotor, because a bottle is only ever as capable as the rotor it sits in.
What is the centrifuge tube, and how is it different from a bottle?
A centrifuge tube is a small capped container, usually a few millilitres to tens of millilitres, designed to hold a sample in a centrifuge. The most common form in biology labs is the conical tube, covered in CASRAI’s guide to the conical centrifuge tube. A bottle is the larger-volume counterpart: the bottle records reviewed for this page range from 250 mL to 1,000 mL, while the Nalgene Oak Ridge polypropylene copolymer tube used here for contrast holds 16 mL.
The size difference changes the numbers. The 16 mL Oak Ridge tube is rated by its manufacturer to 50,000 × g, higher than any of the bottles discussed below. The same record states that the tube must be filled to 80 percent, and that a sealing cap gives extra leak protection when spinning above 10,000 × g. Bottles and tubes therefore share rules (fill level, sealing closures) but not ratings, and one should never be substituted for the other on the assumption that they are equivalent.
What are the bottles in a lab called?
Within centrifugation, the container for spinning volumes in the hundreds of millilitres is called a centrifuge bottle, and manufacturers name specific products by material, size and closure. Examples in the records read for this page are a “Nalgene polycarbonate centrifuge bottle” (500 mL), a “Nalgene PPCO bottle with sealing closure” (250 mL) and a “Sorvall high performance centrifuge bottle” (1 L). Other lab bottles, such as media or reagent storage bottles, are not designed for centrifuge rotors, and nothing on a general storage bottle’s label implies a safe spin speed. Always use a bottle that the manufacturer lists for centrifugation, in a rotor that the manufacturer lists for that bottle.
Materials: polycarbonate versus polypropylene copolymer
Two plastics appear across the manufacturer records. Their differences are the first decision point when choosing a bottle, and the table below keeps only differences that a manufacturer record states directly.
| Property | Polycarbonate (PC) | Polypropylene copolymer (PPCO) |
|---|---|---|
| Chemical compatibility, as stated for tubes | Described, on the Oak Ridge tube record, as compatible with fewer reagents than PPCO; the 500 mL bottle is clear polycarbonate | Described as compatible with a wider variety of reagents than polycarbonate |
| Autoclaving | The 500 mL bottle is listed as autoclavable; the 1 L record warns that autoclaving weakens PC bottles and shortens their life | The 250 mL bottle is listed as autoclavable |
| Radiation sterilization | Named as the alternative where PPCO is not recommended for radiation | Not recommended for sterilization by radiation |
| Rated speed in the records read | 13,700 × g (500 mL bottle) | 27,500 × g (250 mL bottle with sealing closure) |
Two cautions come with this comparison. First, the two rated speeds belong to bottles of different sizes in different rotors, so they are not a head-to-head test of the two plastics. Second, the compatibility statement is made on the tube record, not on the bottle records; for a specific solvent or reagent, use the manufacturer’s chemical resistance information and CASRAI’s reference on chemical resistance of plastics as a starting point, not as a substitute for a compatibility test under your own conditions.
Closures, sealing caps and what “leakproof” means
Closures matter as much as the bottle body. The 500 mL polycarbonate bottle ships with a polypropylene screw cap, and the manufacturer lists optional polypropylene sealing cap assemblies for the 250 mL and 1,000 mL sizes. These assemblies are described as providing leakproof performance at maximum speeds or when hazardous materials are spun. The 250 mL PPCO bottle’s sealing closure uses a silicone gasket and is described as giving “leakproof assurance at maximum speeds”, including with hazardous materials.
The 1 L Sorvall high-performance bottle uses an aluminum cap instead. That cap tolerates 121 °C for 20 minutes or 134 °C for 3 minutes, and the record sets a service life of 1,000 cycles or 5 years for the cap.
The 500 mL record also defines what “leakproof” means in practice. It refers to the manufacturer’s own test criteria, run with water, and states that it may not hold for other liquids; the record tells users to test under their own conditions. For a lab working with a viscous, volatile or biohazardous sample, this is the sentence to read twice. A leakproof claim is a test result for a specified liquid, not a property of the bottle in general.
Independent biosafety guidance points the same way for infectious material. Minnesota’s biosafety guidance says to use sealed safety cups, safety buckets or sealed rotors with O-rings as secondary containment, to cap tubes before centrifugation and to replace worn, cracked or missing O-rings. Cornell’s centrifuge guidance recommends O-ring gasketed screw-top tubes and describes sealed cups or buckets as secondary containment that increases safety if something goes wrong inside the machine.
Rated speeds and why they depend on the rotor
A centrifuge bottle’s rating is not a single number. The maximum RCF is stated for a bottle in a named rotor, and the same bottle can have several ratings. The Sorvall 1 L bottle record is the clearest example: it lists three ratings for one bottle family, one per rotor.
- 20,584 × g in the F10-4×1000 LEX rotor in a LYNX 6000 or LYNX 4000 centrifuge.
- 17,568 × g in the F9-6×1000 LEX rotor in a LYNX 6000.
- 9,333 × g in the F5-10×1000 LEX rotor in an RC BIOS 10.
The bottle did not change between these three lines; the rotor did, and so did the rating, by more than a factor of two. The rating belongs to the bottle, rotor and centrifuge together. For the wider matching decision, see CASRAI’s guide to choosing a centrifuge.
The data table below this page’s text sets out the ratings from the manufacturer records side by side, including the 500 mL polycarbonate bottle at 13,700 × g, the 250 mL PPCO bottle at 27,500 × g and the 16 mL Oak Ridge tube at 50,000 × g. Two further compatibility notes appear in the records: the 250 mL bottle used in a Sorvall GSA rotor needs adapter DS3125-0250, and the 1 L record gives a maximum sample density of 1.2 g/mL. Both are the type of detail that a catalogue headline hides and a rotor manual confirms.
A worked example: reading a bottle record before a run
- Find the bottle’s maximum RCF for your rotor, not the largest number on the page. For the 1 L bottle, that means choosing among 20,584, 17,568 and 9,333 × g according to the rotor model.
- Compare it with the RCF your protocol needs. If the protocol is written in RPM, convert it using your rotor’s radius, as explained in CASRAI’s centrifuge guide. A protocol value above the bottle’s rated figure for that rotor is not a spin you can run in that bottle.
- Check the sample density against the stated limit (1.2 g/mL for the 1 L bottle).
- Choose the closure the record names for your use (sealing cap assembly for hazardous material or maximum-speed runs).
- Confirm whether an adapter is needed for the bottle size in your rotor.
Fill levels, autoclaving and inspection
How full to fill a bottle
The manufacturer records agree on a fill rule. The 500 mL polycarbonate bottle record says to fill to at least 80 percent of capacity; the 250 mL PPCO bottle record says to fill to 80 percent; the 1 L record gives the working range of 800 to 1,000 mL. This is a bottle-specific instruction. General biosafety guidance for tubes runs the other way: the University of Minnesota warns never to overfill centrifuge tubes and not to exceed three-quarters full. The two statements concern different containers and different purposes (the bottle records give a fill requirement, the tube guidance a limit on overfilling), so follow the instructions for the specific container in front of you. When in doubt, the manufacturer’s instructions for that bottle take precedence over a general rule written for tubes.
Autoclaving
The 1 L Sorvall record gives the most specific handling statement in the set: do not autoclave with closures engaged, because bottles can collapse on cooling. The same record notes that autoclaving weakens polycarbonate bottles and shortens their life. CASRAI’s guide to which plastics are autoclavable covers the general logic; for centrifuge bottles, the instruction to follow is the bottle’s own.
Inspection and retirement
The 1 L record sets replacement intervals: the aluminum cap at 1,000 cycles or 5 years, and the PPCO or PC bottles and plugs at 50 cycles. It also lists the signs for removing a bottle from service: wear, cracks, crazing, discoloration, yellowing, deformation, scratches, abrasion or chemical attack. Minnesota’s guidance adds a check before every use for cracks and deformities in tubes, bottles and rotors. Keeping a cycle log for each bottle is the practical way to apply a 50-cycle interval; the interval is the manufacturer’s, and the log is the lab’s own way of keeping to it.
After the run
Cornell’s guidance says not to open the centrifuge until it has fully stopped and to allow a rest of at least 30 minutes so most aerosols settle. Minnesota’s guidance says to wait at least 10 minutes before opening when the material is infectious. These are two institutions’ guidance, not a universal number, and a lab’s own biosafety officer sets the rule that applies on site. Both sources also say to balance loads carefully before the run and to stop immediately if noise or vibration begins.
What are three types of centrifuges?
CASRAI’s centrifuge guide sorts lab centrifuges by the maximum force they produce and the sample formats they take. It names microcentrifuges (for microtubes), general-purpose clinical or benchtop centrifuges (conical tubes, blood-collection tubes, microplates), high-speed centrifuges (floor-standing units with substantially higher RCF) and ultracentrifuges (the highest-force class). Any three of these answer the question, but the pairing for a bottle is set by the bottle record and the rotor: the 1 L records above name specific high-performance centrifuges and rotors.
How much does a lab centrifuge cost?
Cost depends on the centrifuge class, the rotors, the adapters and the bottles, and this page does not give prices because none of the manufacturer or biosafety sources used here state them. For the cost drivers behind a purchase, CASRAI’s guide to choosing a centrifuge is the right place.
Common mistakes
- Taking the largest rating on the page as the rating for your setup. The 1 L bottle’s rating varies from 9,333 to 20,584 × g by rotor.
- Treating “leakproof” as universal. The manufacturer’s test uses water and may not hold for other liquids.
- Autoclaving with the cap tightened. The 1 L record warns that bottles can collapse on cooling.
- Applying a tube rule to a bottle (or the reverse). Fill levels and ratings are set per container.
- Running a bottle past its service interval. The 1 L record sets 50 cycles for the bottles and plugs and lists visible signs, such as crazing or yellowing, that call for removal.
- Using radiation sterilization on PPCO. The records say it is not recommended.
- Skipping the adapter. A 250 mL bottle in a Sorvall GSA rotor needs adapter DS3125-0250.
Frequently asked questions
What are centrifuge bottles?
They are capped, large-volume plastic containers, 250 mL to 1 L in the records read, made for spinning samples in a centrifuge rotor that accepts bottles. Each model carries a manufacturer-stated maximum RCF that depends on the rotor.
What is the centrifuge tube?
It is a small capped container for spinning a sample, typically a few millilitres to tens of millilitres. The conical tube is the most common form; a 16 mL Oak Ridge PPCO tube is rated to 50,000 × g by its manufacturer.
What are the bottles in a lab called?
Bottles designed for spinning are called centrifuge bottles, and manufacturers name them by material, volume and closure. Ordinary storage bottles are not made for centrifuge rotors.
How much does a lab centrifuge cost?
It varies with centrifuge class, rotor and accessories, and no source used for this page states a price, so none is given.
What are three types of centrifuges?
General-purpose benchtop, high-speed and ultracentrifuges are three classes in CASRAI’s centrifuge guide, which also describes microcentrifuges for microtubes.
How full should a centrifuge bottle be?
The records read say to fill to at least 80 percent (500 mL polycarbonate bottle), to 80 percent (250 mL PPCO bottle), or to 800 to 1,000 mL (1 L bottle). Follow the instructions for the specific bottle.
Can centrifuge bottles be autoclaved?
The 500 mL polycarbonate and 250 mL PPCO bottles are listed as autoclavable. The 1 L record says not to autoclave with closures engaged and that autoclaving weakens polycarbonate bottles.
When should a centrifuge bottle be replaced?
The 1 L record says to replace bottles and plugs by 50 cycles and to remove any bottle showing wear, cracks, crazing, discoloration, yellowing, deformation, scratches, abrasion or chemical attack.
Sources
- Thermo Fisher Scientific: Nalgene polycarbonate centrifuge bottle, 500 mL (3122-0500), product record, checked 2026-11-20
- Thermo Fisher Scientific: Nalgene PPCO centrifuge bottle with sealing closure, 250 mL (3141-0250), product record, checked 2026-11-20
- Thermo Fisher Scientific: Sorvall high performance centrifuge bottle, 1 L (010-1456), product record, checked 2026-11-20
- Thermo Fisher Scientific: Nalgene Oak Ridge PPCO centrifuge tube, 16 mL (3139-0016), product record, checked 2026-11-20
- University of Minnesota, Health Safety and Risk Management: Centrifuges, biosafety guidance, checked 2026-11-20
- Cornell University Environment, Health and Safety: Centrifuges, biosafety guidance, checked 2026-11-20
Data: Maximum rated RCF and handling notes for Thermo Fisher centrifuge bottles and a reference tube
Compares the manufacturer-stated maximum relative centrifugal force, material, nominal volume and rotor condition for centrifuge bottles of 250 mL to 1 L and one 16 mL tube, compiled from four Thermo Fisher Scientific product records.
Period covered: 2026.
| Product | Material | Nominal volume (mL) | Maximum RCF (x g) | Rotor or condition stated | Autoclave note |
|---|---|---|---|---|---|
| Nalgene PC centrifuge bottle 3122-0500 | Polycarbonate | 500 | 13700 | Maximum speed stated for the bottle; PP screw cap included | Autoclavable |
| Nalgene PPCO bottle with sealing closure 3141-0250 | Polypropylene copolymer | 250 | 27500 | Maximum speed stated for the bottle; sealing cap assembly with silicone gasket | Autoclavable; not recommended for radiation sterilization |
| Sorvall high performance bottle 010-1456 (PPCO) or 010-1459 (PC) | PPCO or PC | 1000 | 20584 | F10-4x1000 LEX rotor in LYNX 6000 or LYNX 4000 | Do not autoclave with closures engaged; aluminum cap 121 C 20 min or 134 C 3 min |
| Sorvall high performance bottle 010-1456 (PPCO) or 010-1459 (PC) | PPCO or PC | 1000 | 17568 | F9-6x1000 LEX rotor in LYNX 6000 | Do not autoclave with closures engaged; aluminum cap 121 C 20 min or 134 C 3 min |
| Sorvall high performance bottle 010-1456 (PPCO) or 010-1459 (PC) | PPCO or PC | 1000 | 9333 | F5-10x1000 LEX rotor in RC BIOS 10 | Do not autoclave with closures engaged; aluminum cap 121 C 20 min or 134 C 3 min |
| Nalgene Oak Ridge tube 3139-0016 (tube, for contrast) | Polypropylene copolymer | 16 | 50000 | Maximum speed stated for the tube; sealing cap advised above 10,000 x g | Not stated in the record read |
Sources
- Nalgene polycarbonate centrifuge bottle 500 mL, 3122-0500 (accessed 2026-11-20)
- Nalgene PPCO bottle with sealing closure 250 mL, 3141-0250 (accessed 2026-11-20)
- Sorvall high performance centrifuge bottle 1 L, 010-1456 (accessed 2026-11-20)
- Nalgene Oak Ridge PPCO tube 16 mL, 3139-0016 (accessed 2026-11-20)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.