What Is a Kendrick Extrication Device (KED)? Parts, Specs and Safe Use

A Kendrick Extrication Device (KED) is a vest-style short spinal device. See its components, Ferno Model 125 specifications and load limit, and how it differs from a long spine board.

Written and maintained by CASRAI Editorial Board

Published · Last updated

This page describes products and general use; it is not medical advice.

On this page

A Kendrick Extrication Device (KED) is a semi-rigid, vest-style emergency patient-handling device. Ferno’s Model 125 users’ manual describes it as designed to aid in the immobilization and short transfer movement of patients with suspected spinal or cervical injuries or fracture, and for use by a minimum of two trained operators. This page covers what it is made of, its published dimensions and load limit, and how the manufacturer says to clean and dry it.

What a KED is

The KED is a short spinal device: it wraps the head, neck and torso of a seated patient so the patient can be moved onto a long spine board. It does not replace the board. Ferno’s manual directs that once the KED is applied and the patient has been transferred, the patient is secured to the long spine board with restraints. The manual also lists radiographic translucence as a feature.

Who uses it: emergency medical services and rescue personnel trained on the device, and the instructors, students and service buyers who work with it. The manual states that only trained personnel should apply the KED, and that trainees should read the manual, practice, be tested and have their training recorded.

How it is built

According to the Model 125 users’ manual (Ferno pub. 234-1754-02, April 2001 edition), the device has color-coded sewn-in securing straps, wraparound head flaps, wraparound torso flaps, head straps, an Adjusta-Pad for padding, integrated carrying handles and a carrying case. The listed components are:

  • 3 torso straps (3 buckles) and 2 leg straps (2 buckles)
  • 2 head straps, 2 head flaps and 2 torso flaps
  • 1 center and 2 side lifting handles
  • 10 hook-and-loop strips, the Adjusta-Pad and a carrying case

Specifications (Ferno Model 125)

Values below are from section 3.2 of Ferno’s Model 125 users’ manual, April 2001 edition, rounded as the manual gives them. Ferno states it may change specifications without notice, so check the manual that ships with the unit you actually have.

Specification Open flat Rolled, in case
Length 33 in (83 cm) 35 in (90 cm)
Width 32 in (80 cm) 11 in (27 cm)
Thickness 1 in (3 cm) 6 in (14 cm)
Weight (with restraints, pad, head straps and case) 7 lb 11 oz (3 kg)
Load limit 500 lb (227 kg); the manual says to inspect the KED if the limit has been exceeded

KED and long spine board: how the roles differ

Device Role as described in the Ferno manual
KED Immobilization and short transfer movement of a patient with suspected spinal or cervical injury; designed for use by a minimum of two trained operators
Long spine board Where the manual says the patient is secured with restraints after the KED is in place and the patient transferred

For the wider family of boards, stretchers and related straps, see CASRAI’s backboards and stretchers guide, the EMS equipment overview the guide to stretcher straps and what is a cervical collar.

Safe use and handling basics

General points from the manufacturer’s manual, not a procedure:

  • The manual says to use the device only as it describes and to permit only trained personnel to apply it.
  • It tells users to follow their medical director’s guidelines and local protocols.
  • The manual includes adaptations for particular patients. These depend on the patient’s size and condition and fall under local protocol and training, so they are not reproduced here.

Cleaning and drying

Ferno’s manual (section 8.2) describes this sequence: rinse with warm water; lightly scrub the KED and the Adjusta-Pad with a mild soap solution and a medium-bristle brush; wash the head straps with soap and a clean sponge; rinse; disinfect with a solution of 2% to 5% chlorine bleach in water; rinse; and dry completely before returning the device to its case. The manual gives a drying time of 2 to 36 hours because moisture enters through the sewing holes and wets the battens inside, and warns that storing a damp or wet KED in its carrying case encourages mold growth. The manual also has a maintenance schedule table covering cleaning, disinfecting and inspecting; follow that table for your edition.

Common handling mistakes

  • Putting it away damp. See the drying note above.
  • Ignoring the load limit. The manual says to inspect the device if the 500 lb limit has been exceeded.
  • Using an old manual. Specifications and instructions can change by edition and model.

FAQ

Is a KED a spine board?

No. A KED is a short device for immobilization and short transfer; the manual has the patient secured to a long spine board afterwards.

What is the load limit of a KED?

The Ferno Model 125 manual (April 2001 edition) gives 500 lb (227 kg) and says to inspect the device if that limit has been exceeded.

How is a KED cleaned?

Mild soap, rinse, a 2% to 5% chlorine bleach solution, rinse, then complete drying, which the manual says can take 2 to 36 hours.

Sources

  • Ferno, Model 125 KED Users’ Manual, pub. 234-1754-02, April 2001 edition (superseded edition), checked 2026-11-01: manual PDF.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about What Is a Kendrick Extrication Device (KED)? Parts, Specs and Safe Use

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →