Skip to main content
v2026.11,986 entries · CC-BY 4.0

EPA Method 1633A: 40 PFAS in Water, Soil, Biosolids and Tissue

What EPA Method 1633A measures, holding times by matrix, how it differs from Method 1633, and its CWA regulatory status.

Written and maintained by CASRAI Editorial Board

Published

On this page

EPA Method 1633A is the U.S. Environmental Protection Agency’s liquid chromatography tandem mass spectrometry (LC-MS/MS) method for measuring 40 per- and polyfluoroalkyl substances (PFAS) in water, soil, sediment, biosolids and fish and shellfish tissue. Revision A, published in December 2024 as EPA 820-R-24-007, supersedes all earlier versions. As of the EPA pages last updated 25 February 2026, it had been proposed, but not yet promulgated, for Clean Water Act compliance monitoring.

Definition and scope

The full title is “Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous, Solid, Biosolids, and Tissue Samples by LC-MS/MS”, issued by the EPA Office of Water. Section 1.1 states that it is for use under the Clean Water Act (CWA) and covers the PFAS listed in its Table 1 in aqueous samples, solid samples (soil, biosolids, sediment) and tissue. EPA describes it as currently the only PFAS method validated in multiple laboratories for aqueous matrices that include wastewater, surface water, groundwater and landfill leachate, as well as for soil, sediment, biosolids, and fish and shellfish tissue.

The method has 40 target analytes spread across nine compound classes. Multi-laboratory validation data show the method is suitable for all 40 in all the matrices studied, but performance differs between classes. EPA therefore flags in Table 1 the analytes that did not perform as well in some matrices; the flagged analytes still reached 50 to 150 percent accuracy in most validation analyses, and the method tells data users and laboratories to account for that during project planning. The 40 target analytes are listed in Table 1 of the method itself, which is the authority for the exact list.

Quantification relies on isotopically labelled standards. Where linear and branched isomers are present and standards containing both are commercially available, the method reports one result calculated from the combined responses of the isomers. By default, CWA uses of the method report analytes in their acid or neutral forms.

How the method is used in practice

All matrices follow the same pattern: spike isotopically labelled standards into the sample, extract, clean up, then analyse by LC-MS/MS in multiple reaction monitoring (MRM) mode. Concentrations come from isotope dilution or extracted internal standard quantification, which corrects for losses during preparation and can address matrix effects. What differs is the extraction:

  • Aqueous samples: extracted with solid-phase extraction (SPE) cartridges, then cleaned up with carbon.
  • Solid samples: extracted into basic methanol, then cleaned up with carbon and SPE cartridges.
  • Tissue samples: extracted in potassium hydroxide and acetonitrile, followed by basic methanol, then cleaned up with carbon and SPE cartridges.

The method states that the instrumental portion is for analysts experienced with LC-MS/MS, or working under close supervision of such analysts, and that each laboratory must demonstrate acceptable performance using the procedure in Section 9.2 before using it.

Sample-container effects

Section 1.4 notes that PFAS readily adhere to the walls of sample containers and can stratify inside them. The method therefore requires the handling procedures in Section 11 and Appendices A and B. One consequence for project teams: since the 2024 revision, subsampling a sample requires client approval (Section 11.2, repeated as a warning in Appendix B), so the sampling plan and the laboratory contract should agree on this before bottles are filled.

Holding times

Section 8.5 sets holding times by matrix and storage temperature, and Revision A adds a summary in Section 8.5.6. The data table below compares them. For NPDES permit work, laboratories must meet the holding times for the analytes the permit lists. Tissue frozen within 48 hours of collection starts its holding-time clock when it is further processed (Section 8.5.3). Biosolids are held up to 90 days, and EPA recommends freezing if they are stored more than a few days. Extracts may be held 90 days at 0 to 6 °C or at or below -20 °C in the dark, with an issue noted for 11Cl-PF3OUdS and 9Cl-PF3ONS after 28 days.

Version history and what Revision A changed

The method went through several public drafts before it became a final method. This cross-walk is the quickest way to tell which version a laboratory report or contract is citing.

Date Version
August 2021 First draft
April 2022 Second draft
December 2022 Third draft
July 2023 Fourth draft
January 2024 Method 1633 (no longer a draft)
March 2024 Errata sheet for Method 1633
December 2024 Method 1633 Revision A (1633A)

The method’s own change table lists these differences in Revision A:

  • Section 1.6: laboratories may add other PFAS target analytes or alternative labelled standards without prior EPA approval, provided they meet the Section 9 requirements for the new analytes.
  • Sections 8.2.1 and 11.0: “percent solids” is replaced by “total suspended solids” for aqueous samples.
  • Section 8.5.6: new summary of holding times.
  • Section 11.2: client approval is needed before subsampling.
  • Section 15.1.3: six analytes are identified for which the ion-ratio criterion does not apply, because PFPeA has low signal-to-noise.
  • Section 15.2: the percent-solids term is used for dry-weight reporting.
  • Section 15.3.2: a sentence prohibiting reporting for compliance when extracted internal standard recoveries are below 5 percent was deleted.
  • Table 11: new rows for the laboratory duplicate and the field duplicate.
  • Table 10: m/z values rounded to unit mass resolution.

Regulatory status: performance-based, but not yet promulgated

Two separate ideas are easy to confuse here. First, Section 1.5 calls the method “performance-based”: laboratories may modify it without additional EPA review to improve performance, provided every performance criterion in the method is still met. For CWA uses, further flexibility is described at 40 CFR 136.6, Method modifications and analytical requirements. Changes outside 136.6 and Section 9.0 may, after promulgation, require review under the Alternate Test Procedure program at 40 CFR 136.4 and 136.5. Our guide to 40 CFR Part 136 explains that framework.

Second, approval is a separate matter. EPA proposed adding Method 1633A at 40 CFR 136.3 in December 2024 (docket EPA-HQ-OW-2024-0328), in a Methods Update Rule signed on 6 December 2024 that also proposed Method 1621 (adsorbable organic fluorine) and Method 1628. The same proposal included ASTM D8421, which covers the same 40 PFAS as 1633A, as a voluntary consensus standard. EPA’s PFAS methods page, last updated 25 February 2026, states that the method is not nationally required for CWA compliance monitoring until EPA has promulgated it through rulemaking. Check the Federal Register and EPA’s Methods Update Rules page for any change since that date.

How 1633A relates to other EPA methods

Method numbers answer different questions, and the questions below come up often alongside 1633A.

What is the EPA 537.1 method used for?

Method 537.1 is an EPA drinking-water method for a defined list of PFAS by solid-phase extraction and LC-MS/MS. It is a drinking-water method, whereas 1633A is aimed at CWA matrices such as wastewater, soil and tissue. See our guide to EPA Method 537.1, and the guide to EPA Method 533, for the drinking-water comparison.

What is the EPA 200.8 test method used for?

Method 200.8 measures trace elements (metals) in water by inductively coupled plasma mass spectrometry. It is an elemental method and has nothing to do with PFAS. Our comparison of 200.8, 200.7 and 6020 covers it.

What are EPA methodologies?

They are the published analytical procedures that EPA, mainly through its Office of Water for water and the SW-846 compendium for waste, specifies for measuring a pollutant: sample collection and preservation, preparation, instrument analysis and quality control. A method number such as 1633A identifies one procedure and one revision; the “A” is a revision letter, which is why the version cited on a report matters.

Do all pesticides have an EPA number?

No. An EPA registration number is issued to pesticide products registered under the Federal Insecticide, Fungicide, and Rodenticide Act (FIFRA); it is a product identifier and unrelated to analytical method numbers such as 1633A. EPA exempts certain “minimum risk pesticides” from the requirement to be registered (source: EPA, Minimum Risk Pesticides, read 21 November 2026).

Where it matters for laboratories and project teams

  • Contracts and reports: state whether the work is to Method 1633, 1633A or a draft, because the January 2024 and December 2024 texts differ in the items listed above.
  • Analyte list: because laboratories may add targets under Section 1.6, confirm which analytes beyond the 40 in Table 1 are reported and that the Section 9 requirements were met for them.
  • Short holding times: if NMeFOSE, NEtFOSE, NMeFOSAA or NEtFOSAA are target analytes, aqueous samples have a 7-day limit unless stored at or below -20 °C; if NFDHA is a target in soil, sediment or tissue, the limit is 3 days.
  • Permit language: for NPDES monitoring, whether 1633A is required depends on permit language and rulemaking status, not on the method alone.

Common misunderstandings

  • “1633A is an approved CWA method.” It was proposed for approval; EPA’s page says it is not nationally required until promulgated.
  • “Performance-based means anything goes.” Modifications are allowed only when all the method’s performance criteria are met and the changes are documented.
  • “It only covers water.” It also covers soil, sediment, biosolids and tissue.
  • “The A revision is only editorial.” The change table includes new analyte flexibility, a subsampling approval requirement and changes to QC and reporting rules.

FAQ

What does Method 1633A measure?

Forty PFAS in nine compound classes, in aqueous, solid, biosolids and tissue samples, by LC-MS/MS with isotope dilution or extracted internal standard quantification.

Is Method 1633A the same as Method 1633?

No. Method 1633 was published in January 2024 after four drafts; Revision A (December 2024) supersedes all previous versions.

Can a laboratory add PFAS to the target list?

Yes, under Section 1.6 and 40 CFR 136.6, without prior EPA approval, if the Section 9 requirements are completed for the new analytes.

Sources

Data: EPA Method 1633A holding times by sample type and storage condition

Holding times in days from Section 8.5 and the 8.5.6 summary of Method 1633A (December 2024), compiled by sample type and storage temperature, with the shorter limit that applies when specific analytes are targets.

Period covered: 2024. Coverage: United States.

EPA Method 1633A holding times by sample type and storage condition
Sample typeStorage conditionStandard holding time (days)Shorter holding time (days)Condition for the shorter limit
Aqueous, including leachates0 to 6 °C, dark287NMeFOSE, NEtFOSE, NMeFOSAA or NEtFOSAA is a target analyte (or store at or below -20 °C)
Aqueous, including leachatesAt or below -20 °C90noneNot applicable
Soil and sedimentEither storage temperature903NFDHA is an analyte of interest
BiosolidsFrozen preferred; EPA recommends freezing if stored more than a few days90noneNot applicable
Tissue0 to 6 °C903NFDHA is an analyte of interest
Extracts0 to 6 °C or at or below -20 °C, dark90noneIssue noted for 11Cl-PF3OUdS and 9Cl-PF3ONS after 28 days

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about EPA Method 1633A: 40 PFAS in Water, Soil, Biosolids and Tissue

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →