Skip to main content
v2026.11,986 entries · CC-BY 4.0

Freeze-Pump-Thaw Degassing: Steps, Cycles and Hazards

Freeze-pump-thaw degassing: the steps, how many cycles, how four lab procedures differ and the glassware hazards to control.

Written and maintained by CASRAI Editorial Board

Published

On this page

Freeze-pump-thaw is a degassing method for liquids: the liquid is frozen in a sealed, thick-walled vessel, the headspace above the frozen solid is evacuated, the vessel is closed and the liquid is warmed until it just melts so dissolved gas escapes into the headspace, and the cycle is repeated, usually three times, before the vessel is backfilled with inert gas. It is the method that Jeffrey Bode’s Not Voodoo guide calls the most effective way to degas a solvent, and its main hazard, according to the Oregon State University Department of Chemistry safe operating procedure, is the glass vessel shattering.

What freeze-pump-thaw degassing is and who uses it

Dissolved oxygen and other gases in a solvent or solution can ruin an air-sensitive reaction. Bode’s guide points out that for some chemistry degassing is essential: reactions heated above 120 °C, many reactions held above 80 °C for a prolonged period, organometallic reactions, radical and photochemical reactions, and substrates that contain thiols, thioethers, phosphines or electron-rich aromatic groups. Freeze-pump-thaw is the technique chemists reach for when sparging is not rigorous enough or when the liquid is too valuable or too volatile to lose to bubbling.

The users are synthetic and organometallic chemists, chemists running radical reactions, which Bode says must be degassed unless oxidation is the desired outcome (controlled radical methods such as RAFT polymerisation and atom transfer radical polymerisation are examples), and anyone preparing deuterated or other expensive solvents for air-sensitive work. The operation is performed on a Schlenk line with a flask of the kind described in what is a Schlenk tube. It is a bench technique for trained laboratory staff, not something to improvise: every procedure cited here treats it as a pressure-and-vacuum operation on glassware with an implosion risk.

Equipment and set-up

The Oregon State procedure and the University of Washington handout describe the same basic set-up. The sources agree on the parts below, and differ only in the cooling bath and the details of the thaw.

  • A thick-walled Schlenk flask with a closable stopcock. Oregon State says to use only thick-walled glass designed to resist pressure changes, typically a thick-walled Schlenk flask. Bode also allows a heavy-wall sealed tube. Fill the flask no more than 50% full: the University of Washington handout warns that overfilled flasks frequently shatter during the process.
  • A Schlenk line. Oregon State specifies a dedicated gas and vacuum double manifold with exit bubblers, with at least one bubbler-protected vent open at all times to prevent overpressurisation. Both Oregon State and the University of Washington say to leave the connecting hose on vacuum throughout.
  • A freezing bath. Liquid nitrogen is the usual choice in the Oregon State, University of Washington and Bode procedures. The LibreTexts Schlenk Line Survival Guide uses a −78 °C dry ice and acetone bath by default and reserves liquid nitrogen for valuable reagents and deuterated solvents.
  • A tepid water bath for the thaw. Oregon State and the University of Washington both specify a tepid water bath; the LibreTexts guide says not to use a warm water bath at all. This difference is covered below.
  • A fume hood with a blast shield. Oregon State says all manipulation of sealed vessels should happen in a well-vented hood behind a blast shield, with the hood cleared of unnecessary equipment and chemicals.

This page does not cover vacuum concentration of samples, which is a different operation with different equipment; see SpeedVac for that.

Step-by-step procedure

The sequence below follows the Oregon State safe operating procedure (version 1.0, Christopher M. Beaudry, Department of Chemistry), cross-checked against the University of Washington handout (Mark Lonergan, 2003). Where another source gives a different value, the next section says so rather than blending them.

  1. Load the flask. Place the solvent or solution in a Schlenk flask with the stopcock closed. Fill it to no more than 50% of its volume.
  2. Connect to the line. Attach the flask to the Schlenk line and leave the hose on vacuum throughout the procedure.
  3. Freeze. Freeze the liquid completely, preferably in liquid nitrogen. Before freezing, make sure the environment in the flask is free of oxygen so that liquid oxygen does not condense.
  4. Pump. When the liquid is fully frozen, open the stopcock to vacuum. Oregon State and Washington give 10 to 30 minutes, or until the headspace is completely evacuated.
  5. Seal. Close the stopcock so the flask holds a static vacuum.
  6. Thaw. Warm the liquid in a tepid water bath until it just melts. Gas bubbles will evolve; try not to disturb the liquid.
  7. Repeat. Remove the water bath, refreeze, pump and thaw again. Oregon State requires at least three cycles in total; Washington repeats until gas no longer evolves on thawing, with a minimum of three.
  8. Backfill. After the last cycle fill the flask with inert gas and seal it. The solvent is ready to use.

Where the sources disagree, and what that means at the bench

Four readable procedures describe the same technique with different numbers, and none of them is wrong on its own terms. The data table below lists each source’s parameters side by side. The differences that matter in practice are the following.

Pump time per cycle

Bode’s guide pumps for 2 to 3 minutes under high vacuum with the flask still in liquid nitrogen. The LibreTexts guide pumps for 3 to 5 minutes with the flask in its cooling bath. Oregon State and the University of Washington both pump for 10 to 30 minutes. The spread is roughly an order of magnitude. The sources do not state why they differ, so the sensible reading is that the shorter times belong to small volumes on a good high-vacuum line and the Oregon State and Washington times are written as conservative defaults for a teaching and safety document; that reading is an interpretation, not something any of the sources says. Follow the procedure your own laboratory has approved for your line and flask size.

Liquid nitrogen or dry ice and acetone

Bode, Oregon State and Washington freeze in liquid nitrogen. LibreTexts defaults to a −78 °C dry ice and acetone bath because it limits solvent loss under vacuum, and says liquid nitrogen fully freezes the contents but is reserved for valuable reagents or deuterated solvents. LibreTexts also ties liquid nitrogen to a specific hazard: a leak under static vacuum can condense liquid oxygen. Oregon State and Washington handle the same hazard by requiring an oxygen-free flask before freezing.

Tepid water or no warm bath

This is the one point where the sources give opposite instructions. Oregon State and Washington thaw in a tepid water bath, until the liquid just melts. LibreTexts says to remove the flask from the bath and let it warm slowly, and says not to use a warm water bath or a heat gun. Every source agrees on the point that matters: no direct heat. Oregon State says a heat gun may cause the vessel to explode violently.

How the end of the procedure is decided

Washington frames the stopping rule as an observation: repeat until you no longer see gas evolve as the solution thaws, with a minimum of three cycles. Oregon State, LibreTexts and Bode state a fixed count of three (Bode says usually three). The two framings are compatible: three is the floor, and bubbling on the last thaw is the sign that another cycle is needed.

How many cycles are enough

Three cycles is the figure given by Oregon State, LibreTexts, Washington and Bode. LibreTexts says the three cycles ensure that all dissolved gas and gas trapped within the frozen solid is removed. Bode adds that strict procedures sometimes prescribe 10 cycles for the use of sensitive catalysts. A reasonable working rule that follows from the Washington handout is to treat three as the minimum and to continue while gas is still evolving on thaw. The sources do not give an oxygen concentration target, so none is stated here.

Thawing without breaking the flask

The thaw is where glassware most often fails, because the sealed vessel is under static vacuum while the liquid warms and releases gas. The sources identify four ways to get it wrong: letting the frozen solvent thaw on its own, using a container of water that melts only the bottom of the frozen solvent (both from Washington and Oregon State), applying direct heat such as a heat gun (Oregon State, LibreTexts), and overfilling the flask (Washington, Oregon State). Oregon State adds that the thaw should not be left unattended. The goal in every source is a gentle, even melt that releases bubbles without disturbing the liquid.

Hazards and the controls the sources require

The Oregon State SOP names one main hazard, shattering of the glass vessel, which can cause an explosion or implosion. Its controls are specific enough to use as a pre-run checklist.

Control What Oregon State requires
People Never work alone; another person must be present.
Eyes and face Chemical splash goggles or safety glasses meeting ANSI Z87.1; a face shield over them if there is a risk of explosion.
Hands and body Thick gloves when handling sealed vessels; a fully buttoned, knee-length lab coat; fully enclosed shoes.
Location Sealed vessels handled in a well-vented fume hood behind a blast shield, with the hood cleared of unneeded equipment and chemicals.
Manifold Dedicated double manifold with bubblers; at least one bubbler-protected vent open at all times.
Emergency equipment Eyewash and safety shower within ten seconds of travel; shower inspected within the last 12 months; know the nearest fire extinguisher and alarm.

For emergencies, Oregon State says that if a vessel shatters behind the blast shield the response is to turn off the vacuum line and clean the area, and that a hazardous liquid leaving the hood should be reported to environmental health and safety as a chemical spill.

Common mistakes and why they are dangerous

Mistake Consequence given in the source Source
Opening the flask to inert gas while it is frozen in liquid nitrogen Pitfall listed; backfilling nitrogen or argon while frozen and then sealing can cause an explosion on warming as the condensed gas expands LibreTexts
Freezing in liquid nitrogen with a leak under static vacuum Liquid oxygen can condense LibreTexts; Oregon State
Heating the flask with a heat gun or warm bath Flask may crack, implode or explode violently Oregon State; LibreTexts
Overfilling above 50% of flask volume Overfilled flasks frequently shatter Washington; Oregon State
Thawing unevenly or leaving the thaw unattended Vessel may break Washington; Oregon State
Pumping a liquid with no cold bath Solvent evaporates under vacuum LibreTexts

Freeze-pump-thaw compared with other degassing methods

Freeze-pump-thaw is not always the right method. The sources describe three alternatives and rank them differently, which is worth knowing before committing a flask and an hour of line time.

Method How the sources describe it Source
Freeze-pump-thaw Described as the most effective method; sealed flask, three cycles; limits solvent loss because the liquid is cold under vacuum Bode; LibreTexts
Sonication under light vacuum Rough degassing: 0.5 to 1 minute under house vacuum, atmosphere replaced with inert gas, 5 to 10 cycles; adequate for HPLC solvents and some reactions Bode
Purging (sparging) Bubble nitrogen or argon for 30 minutes to 1 hour; Bode calls it the least effective but acceptable for large volumes; take care to prevent evaporation and water condensation Bode
Sparging and boil-degas LibreTexts says either can be as effective as freeze-pump-thaw but with significant solvent loss, so they suit bulk organic solvents only, not expensive deuterated solvents or volatile liquid reagents LibreTexts

Notice that Bode and LibreTexts do not agree on how well sparging performs: Bode ranks it lowest, while LibreTexts says it can match freeze-pump-thaw. The two statements are reconcilable if you read Bode as comparing methods for typical use and LibreTexts as describing what is achievable with a long enough treatment, but that is an interpretation. What both agree on is the practical split: sparging for large volumes of ordinary solvent, freeze-pump-thaw for small, valuable or volatile liquids.

A pre-run checklist

Built only from the controls and pitfalls above, a short check before starting a freeze-pump-thaw run is: is the flask thick-walled and no more than half full; is a second person present; is the blast shield in place and the hood cleared; is at least one bubbler vent open on the manifold; are goggles, face shield, thick gloves and lab coat on; is the flask oxygen-free before it goes into liquid nitrogen; is the thaw bath ready and will someone stay with the flask; and do you know the backfill gas and when it is introduced (only after the flask is back at room temperature, per LibreTexts). If any answer is no, stop and fix it before the flask goes into the bath.

Backfilling and using the degassed liquid

After the final thaw, the flask is backfilled with inert gas. LibreTexts says this should be done once the flask has returned to room temperature, and that a reactive gas such as hydrogen, carbon monoxide or carbon dioxide can be used instead when the chemistry calls for it. Oregon State and Washington backfill and seal, after which the solvent is ready to use. The procedures cited here do not state how long degassed solvent stays oxygen-free, so check your own storage practice and the requirements of the reaction rather than assuming a shelf life.

Sources

Data: Freeze-pump-thaw parameters as stated in four laboratory procedures

Side-by-side comparison of the freezing bath, pump time per cycle, thaw method, number of cycles and stopping rule given in four openly readable freeze-pump-thaw procedures: an Oregon State SOP, the LibreTexts Schlenk Line Survival Guide, a University of Washington handout and Jeffrey Bode's Not Voodoo guide. Values are copied as stated by each source; no source gives a vacuum pressure.

Freeze-pump-thaw parameters as stated in four laboratory procedures
SourceFreezing bathPump time per cycle (minutes)Thaw methodCyclesStopping rule
Oregon State SOP (Beaudry)Liquid nitrogen (preferred)10-30, or until headspace evacuatedTepid water bath until it just meltsAt least 3Minimum of three cycles, then backfill with inert gas
LibreTexts Schlenk Line Survival GuideDry ice/acetone at -78 C; liquid nitrogen for valuable liquids3-5, flask in cooling bathRemove from bath and warm slowly; no warm water bath or heat gun3Three cycles in total
University of Washington (Lonergan)Liquid nitrogen (usually best)10-30Tepid water bath until it just meltsMinimum 3Repeat until no gas evolves on thawing, minimum three cycles
Not Voodoo (Bode)Liquid nitrogen2-3, high vacuum, flask in liquid nitrogenWarm until completely meltedUsually 3Three cycles usually; strict procedures may prescribe 10 for sensitive catalysts

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · full access

Ask about Freeze-Pump-Thaw Degassing: Steps, Cycles and Hazards

Full access: large monthly usage, up to 10 source passages behind each answer, and MCP and API access for Claude Code, Codex and other CLIs. Cancel any time.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →