Written and maintained by CASRAI Editorial Board
Published
On this page
Pipette repair for a piston-operated micropipette comes down to one split: users can clean the pipette and replace a short list of wear parts (seal, O-ring, tip-holder, tip-ejector, push-button and connecting nut), while anything that involves the piston assembly, readjustment or calibration belongs with the manufacturer’s authorised service centre. That is how the Gilson PIPETMAN Classic user manual draws the line, and Eppendorf’s own guidance says readjusting a pipette should be the very last option when its volume is in doubt. The manual for your exact model always governs; this guide shows what two manufacturers’ documents say and how to triage a fault before you open the pipette.
It is written for researchers, lab technicians and lab managers who keep air-displacement pipettes running and need to decide quickly whether a fault is a bench fix, a calibration question or a shipment to a service centre. Glass and serological pipettes are a different instrument class; see what a pipette is for the overview.
Signs a pipette needs repair
The faults that the Gilson PIPETMAN Classic manual lists in its troubleshooting chapter are sample leaking, failure to aspirate, inaccurate volumes, poor precision, and tips that fall off or do not fit. Eppendorf adds two readings worth knowing before you disassemble anything: a delivered volume that is too low is often an infallible sign of leakage, and a pipette that has become stiff after frequent use with volatile organic liquids may have a swollen piston seal rather than a mechanical failure.
A fault that looks like a broken pipette is sometimes a handling or measurement problem. Eppendorf lists the conditions to rule out before concluding that a pipette has drifted: whether the tip is the right one for the system, the temperature of the sample, the pipette and the air, and the conversion from milligrams to microlitres. Only when those checks remove all doubt should you assume that something such as replacing several key components has altered the adjustment. The section on handling errors below covers the common causes.
What you can replace and what needs a service centre
For the PIPETMAN Classic, Gilson limits user maintenance to cleaning or autoclaving specified parts and to replacing the push-button, connecting nut, tip-ejector, tip-holder, seal and O-ring. The table sorts the parts that the Gilson manual mentions by who should handle them.
| Part or task | Who handles it (Gilson PIPETMAN Classic manual) |
|---|---|
| Push-button, connecting nut, tip-ejector, tip-holder | User may replace; loose connecting nuts are tightened |
| Seal and O-ring | User may replace; replace the two together when worn or damaged; clean or replace the seal first, then the O-ring |
| Piston assembly | May be removed for cleaning only; if it is changed, the pipette must be adjusted and calibrated at an authorised Gilson Service Center |
| Damaged or corroded piston | Return the pipette to the supplier |
| Readjustment after a part replacement | Contact an authorised Service Center |
| P2 and P10 models | Should not be disassembled; only the push-button, tip-ejector, dual position tip-ejector and adapter are replaceable |
Three points from the same manual shape the decision. First, the O-ring and seal sit on the piston and must not be autoclaved. Second, liquid must never enter the upper part of the pipette (the handle). Third, after replacing any part, verify performance with the manufacturer’s verification procedure before putting the pipette back into use. The data table at the end of this page matches each Gilson symptom to its listed causes and the action the manual assigns.
The five tests that tell you what is wrong
Eppendorf’s inspection guidance separates five checks that are often confused. A leak test does not determine performance, and a calibration is the only check that separates random from systematic error.
| Test | What it checks | Notes from Eppendorf Userguide No. 24 |
|---|---|---|
| Functional test | Broken parts, visible contamination | Done by looking and handling |
| Leak test | Airtightness of the tip fit and the piston seal | Does not determine performance |
| Volumetric check | Delivered volume | Two-fold determination; for example when you change to a new batch of alternative tips |
| Quick check | Delivered volume, intermediate check | Four-fold determination between calibrations |
| Calibration | Random and systematic error | Ten-fold determination |
How the Gilson leak test works
The PIPETMAN Classic manual describes the leak test in chapter 10. You fit a PIPETMAN DIAMOND tip, set the maximum volume listed in the specifications, pre-rinse the tip, aspirate that volume of distilled water and hold the pipette vertically for 20 seconds. A droplet forming at the tip means a leak. For P2 to P200 models you can also re-immerse the tip and watch the liquid level inside it: a falling level indicates a leak. If the pipette fails, the manual says to replace the O-ring and seal, check that everything is reassembled correctly and repeat the test. Run the test after maintenance and after decontamination.
Because a leak test says nothing about accuracy, a pipette that passes it can still deliver the wrong volume. Calibration, covered below, is the check that answers that question.
Pipette maintenance: cleaning, autoclaving and drying limits
Most pipette maintenance is cleaning, and the manufacturers draw different limits depending on the model. Always read the limits for your own pipette; the two documents below are examples of how much they can differ.
| Topic | Gilson PIPETMAN Classic | Eppendorf (Userguide No. 24) |
|---|---|---|
| Parts that may be autoclaved | Tip-ejector, tip-holder and connecting nut only; the body and the piston assembly are not autoclavable | Eppendorf mechanical pipettes are fully autoclavable; for electronic pipettes only the lower part; filter tips and dispenser tips should not be autoclaved |
| Autoclave conditions | 20 minutes at 121 °C, 0.1 MPa | 20 minutes at 121 °C, 1 bar overpressure |
| Cleaning | Soap solution recommended; only the connecting nut, tip-ejector, tip-holder, piston assembly, seal and O-ring may be immersed; ultrasonic bath for 20 minutes at 50 °C, or soft cloth or brushes; rinse with distilled water and dry | Dry cleaned parts at a maximum of 60 °C |
| Decontamination chemicals | Check compatibility with stainless steel and PA, PBT, PC, POM and PVDF | Sodium hypochlorite at a maximum of 4% for nucleic acid decontamination, then rinse with distilled water |
| Lubrication after autoclaving | Not stated in the sections read | Not needed for Eppendorf pipettes |
The Gilson manual also sets a usage window of +4 °C to +40 °C and says not to pipette liquids above 70 °C or below 4 °C. After pipetting liquids that give off corrosive vapour, remove the tip-holder and rinse the piston, O-ring and seal with distilled water.
If liquid is accidentally drawn into the instrument, Eppendorf’s advice is that it must not dry: clean the cone and the piston and cylinder immediately and lubricate the piston and cylinder if needed. For dust, aerosols and routine upkeep before a service decision, see the Gilson PIPETMAN guide for model-specific care.
When the liquid is the cause of the damage
Eppendorf describes two liquid-driven failures that look like mechanical wear. Frequent use of volatile organic liquids can make the piston seal, even a so-called maintenance-free one, swell, so the pipette becomes stiff. The remedy it gives is to air the pipette out under a hood overnight, shorten the inspection interval and replace the piston seal more often. Saturated solutions can crystallise and destroy the seal, so Eppendorf recommends regular cleaning, greasing of the piston and cylinder and seal inspection. The same guide notes that leakage may occur when piston grease is etched by aggressive liquids pipetted frequently, and that this has to be detected in good time.
The practical reading is that the repair interval depends on what you pipette. Eppendorf lists the factors that set the inspection frequency: how often the pipette is used, the number of users, how aggressive the liquids are, the error limits the user accepts and the laboratory’s own regulations.
Decontaminate before you ship a pipette
Both manufacturers require a clean instrument at the service centre. Gilson says a pipette returned to a Gilson Service Center must be completely free of chemical, biological or radioactive contamination and returned in the safety bag included with it. Eppendorf says an instrument must be decontaminated before repair, calibration or maintenance and that a signed decontamination form must accompany the shipment, or the calibration laboratory will not process it.
Eppendorf ties the depth of cleaning to the biosafety level of the work. For biosafety levels 1 and 2, cleaning is mandatory and further steps are recommended. For levels 3 and 4, cleaning, decontamination, disinfection and sterilisation are mandatory. For chemical and radioactive contamination, cleaning is mandatory and users must be able to prove the contamination has been removed.
Handling errors that mimic a faulty pipette
Many repair requests are really technique problems. Eppendorf’s Userguide No. 25 tabulates sources of error for air-cushion pipettes with the maximum effect each can have, quoted as a percentage of the nominal volume. The effects are reference values, not guarantees.
| Handling or equipment error | Effect reported by Eppendorf | Remedy it gives |
|---|---|---|
| Not wiping the tip on the vessel wall | Up to 3.0% | Wipe the tip on the vessel wall |
| Not pre-wetting the tip | Up to 2% | Pre-wet the tip |
| Vapour pressure differing from water | Up to 2.0% | Sufficient pre-wetting; a dripping tip is the sign |
| Uneven rhythm and timing | Up to 1.5% | Even pipetting technique |
| Immersion depth and holding angle | Up to 1.0% | Hold the pipette vertically |
| Liquid density differing from water | Up to 1.0% | Readjust the pipette, observing the user information |
| Uneven piston movement | Up to 0.5% | Smooth operation; clean and lubricate the piston |
| Leaky pipette tips | 0.5% up to 50% | Use original or recommended pipette tips |
Two further points from Eppendorf’s Userguide No. 20 apply directly. For air-cushion pipettes, choose a tip so that the air cushion is as small as possible, immerse the tip only a few millimetres, hold the pipette vertically, aspirate slowly and evenly and allow a wait of 1 to 3 seconds for the liquid to rise. The guide’s immersion table gives 1 mm for 0.1 to 1 µl, 2 to 3 mm for 1 to 100 µl, 2 to 4 mm for 101 to 1,000 µl and 3 to 6 mm for 1,001 to 10,000 µl. A pipette held 45° from vertical adds 2.9 µl (0.29%) with a 1,000 µl pipette, and a 60° angle adds 0.53 µl (0.53%) with a 100 µl pipette.
Gilson’s troubleshooting table lists incorrect operator technique as a cause of poor precision, with operator training as the action, and says its specifications are guaranteed only with genuine PIPETMAN DIAMOND tips. Gilson also lists low-quality tips as a cause of tips falling off or not fitting. Eppendorf adds that contamination of the cone usually arrives through handling errors, such as using a tip too small for the set volume or pressing the operating button too quickly. Storing air-cushion pipettes upright, ideally in a carousel, is Eppendorf’s storage advice, and filter tips or positive-displacement systems protect the piston and cylinder.
A repair walk-through: from low volume to back in service
The sequence below is an illustrative composite built from the two manufacturers’ documents, not a record of any one laboratory’s case. It shows how the tests, parts and service rules fit together when a pipette starts delivering less than the set volume.
- Rule out the conditions first. Eppendorf asks you to recheck the tip system, the temperatures of sample, pipette and air, and the mass-to-volume conversion before suspecting the instrument, and notes that a volume that is too low is often a sign of leakage.
- Do a functional test and a leak test. Look for broken parts and visible contamination, then run the leak test described above. Eppendorf is explicit that the leak test checks airtightness only, not performance.
- If it leaks, work through the cheap causes. For the PIPETMAN Classic the manual lists a damaged tip-holder (replace it) and a worn O-ring or seal (replace both). If the model is a P2 or P10, do not disassemble it; a damaged tip-holder on those models may mean a damaged piston.
- Repeat the leak test, then verify performance. Gilson says to check the pipette’s performance after replacing any part. A quick check (four-fold) is Eppendorf’s intermediate test between calibrations.
- If performance is still off, stop opening the pipette. A damaged or corroded piston, a changed piston assembly or a need for readjustment goes to the supplier or an authorised service centre. Clean the instrument first, complete the decontamination form and ship it in the supplied safety bag.
- Calibrate on return. A pipette that has had its piston assembly changed has to be adjusted and calibrated at an authorised Gilson Service Center according to the manual; either way, put the next calibration on the annual cycle.
The order matters because each step is cheaper and less invasive than the next, and because the manuals make several of the later steps the manufacturer’s job rather than the user’s.
Calibration and adjustment after a repair
Replacing a seal or O-ring is not the end of a repair: Gilson says to verify performance afterwards, and Eppendorf says compliance with the error limits in ISO 8655 must be checked at least once a year, with shorter intervals left to the user. For a full routine, use the pipette calibration guide; the points below are the repair-relevant ones from Eppendorf’s Userguide No. 25.
- Calibration is a determination of how much water a pipette delivers under specified conditions; adjustment is a permanent change to the instrument. Calibration does not require an intervention that permanently modifies the device.
- The gravimetric test needs a draft-free room, relative humidity above 50%, a constant temperature (± 0.5 °C) between 15 and 30 °C, and equilibration of the pipette and test liquid in the room for at least two hours. Degassed, distilled or ion-free water is the test liquid.
- For a variable-volume pipette, three volumes are tested: the nominal volume, 50% of it, and the lower limit of the useful volume or 10% of nominal. Generally ten measurements per test volume are made.
- Readjustment is done by the manufacturer’s instructions with the special tool provided and, because adjustment affects the entire measuring range, a pipette that needs it is a service-centre or trained-technician job.
A worked gravimetric example
Eppendorf’s Userguide No. 25 gives a straightforward example for a pipette with a nominal volume of 1,000 µl. Distilled water at 20 °C is aspirated five times and weighed on an analytical balance, giving 0.998 g, 0.999 g, 0.995 g, 0.996 g and 0.975 g. Because 1,000 µl of distilled water weighs 0.9983 g at 20 °C, each mass is converted with the formula volume = mass ÷ 0.9983 g × 1,000 µl, which gives 1,000, 1,001, 997, 998 and 977 µl. The mean of those five volumes is 994.6 µl, so the pipette delivered on average 5.4 µl (0.54%) less than its nominal volume in this example. Four readings sit within a few microlitres of each other and the fifth, at 977 µl, is an outlier that would stand out in any repeatability calculation.
That pattern is the reason calibration is separate from the leak test. The spread tells you about random error (imprecision) and the offset of the mean tells you about systematic error (inaccuracy), and the two point at different causes: scatter points to handling, cleanliness or a failing seal, while a steady offset points to the adjustment. The guide also notes that for small volumes below 50 µl a weighing vessel suited to the volume, and if possible an evaporation trap, should be used to limit evaporation loss during weighing. In the same guide, the Gaussian distribution of repeated measurements is what the terms precise but inaccurate, accurate but imprecise and precise and accurate describe.
Frequently asked questions
What are some common pipette handling errors?
Not wiping the tip on the vessel wall, not pre-wetting the tip, uneven rhythm, a tilted pipette, the wrong immersion depth and tips that do not seal. Eppendorf attributes up to 3.0% error to a tip not wiped on the vessel wall and up to 2% to a tip that is not pre-wetted. The table above gives the others.
How do I clean glass pipettes?
This page covers piston pipettes, whose cleaning limits are set by the manufacturer’s manual. Glass pipettes are washed as laboratory glassware; the equipment used for it is explained in what a pipette washer is.
Can you reuse a pipette?
A piston pipette is a durable instrument and is reused; its parts are maintained as described above. The tips are different: Eppendorf says to change the tip after each pipetting, because reusing tips causes sample-to-sample carryover. Sterile and single-use pipette items are covered in pipette sterile.
How to calibrate the pipette in the lab?
Calibration is gravimetric: dispense distilled water at stated test volumes under controlled temperature and humidity, weigh each dose on a suitable balance, convert mass to volume and compare the mean and spread with the error limits. Eppendorf says compliance with the ISO 8655 limits must be checked at least once a year.
Which pipette faults should go to a service centre?
Per the Gilson manual: a damaged or corroded piston, any change of the piston assembly, readjustment, and any P2 or P10 fault beyond the replaceable external parts. Contamination must be removed first and the pipette returned with the contamination documentation the service centre requires.
Sources
- Gilson: PIPETMAN Classic User Manual (2012), checked 2026-11-21
- Eppendorf: Userguide No. 24, Cleaning and Inspection of Pipettes, checked 2026-11-21
- Eppendorf: Userguide No. 25, Calibration and adjustment of dispensing systems in the laboratory, checked 2026-11-21
- Eppendorf: Userguide No. 20, Impact of pipetting techniques on precision and accuracy, checked 2026-11-21
Data: Pipette fault triage from the Gilson PIPETMAN Classic troubleshooting table
Each symptom listed in the Gilson PIPETMAN Classic user manual troubleshooting chapter, with the cause and corrective action the manual gives, and a column classifying whether the action is a user-level step or a return to the supplier. Compiled from the manual's symptom, cause and action table (read 2026-11-21); the who-performs column is our classification of the manual's stated action.
| Symptom | Cause | Action stated in the manual | Who performs it |
|---|---|---|---|
| Leaking sample | Damaged tip-holder | Replace the tip-holder | User |
| Leaking sample | Worn O-ring or seal | Replace both parts | User |
| Does not aspirate | Worn O-ring | Replace both O-ring and seal | User |
| Does not aspirate | Damaged tip-holder | Replace the tip-holder | User |
| Does not aspirate | Loose connecting nut | Tighten the nut | User |
| Does not aspirate | Damaged or corroded piston | Return the pipette to the supplier | Supplier or service centre |
| Does not aspirate | Improper repair or assembly | Follow the maintenance chapter to reassemble | User |
| Inaccurate volume | Improper repair or assembly | Follow the maintenance chapter | User |
| Inaccurate volume | Unscrewed tip-holder or loose connecting nut | Tighten | User |
| Not precise | Loose tip-holder or connecting nut | Tighten | User |
| Not precise | Incorrect operator technique | Operator training | User |
| Not precise | Damaged or corroded piston | Return the pipette to the supplier | Supplier or service centre |
| Not precise | Damaged tip-holder, worn O-ring or seal | Replace the part | User |
| Tips fall off or do not fit | Low quality tips | Use the manufacturer's own tips (PIPETMAN DIAMOND) | User |
| Tips fall off or do not fit | Damaged tip-holder, damaged or loose tip-ejector, missing or damaged ejector spacer, misaligned ejector lock | Replace or refit the part | User |
Sources
- Gilson: PIPETMAN Classic User Manual (2012) (accessed 2026-11-21)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.







