Written and maintained by CASRAI Editorial Board
Published
On this page
The TSS test, also called the total suspended solids test, measures how much particulate material a water sample carries, in mg/L. In the gravimetric method Standard Methods 2540 D describes, a well-mixed sample is filtered through a weighed standard glass fiber filter, the filter and residue are dried at 103–105 °C to constant weight, and the gain in filter weight divided by the volume filtered is the result.
Short answer
TSS is a weight-based measurement, not an optical one. The lab weighs a clean, dried filter, passes a measured volume of sample through it, dries the filter again at 103–105 °C and weighs it a second time. The increase in mass is the suspended solids retained. This page follows the written procedure in Georgia EPD laboratory SOP 3-005 (Rev. 8) for Standard Methods 2540 D and cross-checks the regulatory status against 40 CFR 136.3, so the numbers below are the ones a regulated wastewater or surface-water laboratory writes into its own procedure.
What TSS is and what it is not
The SOP defines the measurement operationally: solids that stay on a standard glass fiber filter after a well-mixed sample is filtered, then dried to constant weight at 103–105 °C. Anything that passes the filter, including dissolved material, is not counted. The method applies to drinking, surface and saline waters and to domestic and industrial wastes, with a practical range of 1.0 to 20,000 mg/L.
EPA describes the particles behind the number as soil, clay, silt, organic matter, algae and microorganisms. They matter because they can adsorb nutrients, metals and organic contaminants and, in drinking water, can shield microorganisms from ultraviolet disinfection.
TSS versus turbidity
EPA’s turbidity factsheet states that turbidity and TSS are different ways to measure similar water quality characteristics. Turbidity is an optical reading of how much light the sample scatters; TSS is a mass concentration obtained by weighing. A sample can therefore give a high turbidity reading with a modest TSS, or the reverse, depending on particle size and composition, so the two are not interchangeable in a permit or a report. For the optical side, see the guides on the turbidity NTU unit and how a nephelometer reads it.
How the total suspended solids test works in practice
The gravimetric procedure has four parts that the SOP spells out: preparing the filter, choosing the sample volume, drying and weighing, and running quality control. The description below summarises what the SOP states; a laboratory should work from its own controlled copy of the method rather than from a web summary (see the standard operating procedure definition).
Filters, apparatus and sample handling
The SOP lists 47 mm glass fiber filters (Whatman 934-AH, Gelman A/E or equivalent, pre-washed and dried filters allowed), a drying oven held at 103–105 °C, an analytical balance reading to 0.1 mg, a desiccator with colour-indicating desiccant, a vacuum filtration manifold, Class A glassware and a magnetic stirrer. Samples are collected in a plastic half-gallon container, cooled on ice and stored at 0–6 °C without freezing. No chemical preservative is added and the holding time is 7 days. Weighing quality depends on the balance; see the guide on analytical balance calibration and weighing technique, and on desiccator care in laboratory desiccator types and maintenance.
Preparing the filter
Each filter is washed with at least 60 mL of reagent water, dried at 103–105 °C for at least 1 hour (overnight is preferable), cooled in the desiccator for at least 1 hour and weighed. The cycle repeats until the weight is constant, defined in the SOP as a change of less than 4% of the previous weighing or 0.5 mg, whichever is less. The SOP notes a laboratory study in which overnight drying was as effective as drying to constant weight for clean filters.
Choosing the sample volume
The analyst filters as much sample as possible so that the dried residue weighs between 2.5 and 200 mg. If the residue is below 2.5 mg or above 200 mg and less than 1000 mL was filtered, the SOP calls for corrective action and a J-flag on the result. If filtration takes more than 10 minutes, the analysis is restarted with a clean filter and a smaller volume.
Drying, weighing and the calculation
The sample filter is dried overnight at 103–105 °C, or on the same day by drying for at least 1 hour, cooling and re-drying until the constant-weight rule above is met. The result is:
TSS (mg/L) = (A − B) × 1000 / C, where A is the dried residue plus dish in mg, B is the dish alone in mg and C is the sample volume in mL.
As a purely arithmetic illustration (not data from any real sample): if the dried filter plus residue weighs 84.6 mg and the clean dried filter weighed 80.1 mg, then A − B = 4.5 mg; filtering 250 mL gives (4.5 × 1000) / 250 = 18 mg/L. The residue of 4.5 mg also sits inside the 2.5–200 mg window, so no flag is needed.
Quality control
The data table below lists the quality-control and specification values in the SOP side by side. In short, each batch carries a method blank of 1000 mL reagent water, a laboratory control sample and duplicate (LCS/LCSD) per batch of 20 made from Celite 545 stock at 100 mg/L, sample duplicates at 10% of samples, and a daily balance check with certified weights. General background on this practice is in the guide to internal quality control in the laboratory.
Specification checklist
The data table below compares the filter, drying, constant-weight, residue-window, holding-time and quality-control values from the SOP, so a lab can check a draft procedure against a published one line by line. It is an original compilation from the sources listed at the bottom of the page.
Regulatory status: which methods does 40 CFR 136.3 list?
For compliance reporting under the Clean Water Act, the approved test procedures are in 40 CFR 136.3, Table IB. The row for “Residue, non-filterable (TSS)” in mg/L lists the gravimetric technique with drying at 103–105 °C and post-washing of the residue, and names three approved references:
| Approved reference in Table IB | What the table says |
|---|---|
| Standard Methods 2540 D-2020 | Listed for non-filterable residue (TSS), gravimetric, 103–105 °C, post-washing of residue |
| ASTM D5907-18 (A) | Listed for the same parameter in the same row |
| USGS I-3765-85 | Listed, with a footnote pointing to USGS TWRI Book 5, Chapter A1 (1989) |
EPA method 160.2, a name that still turns up in older SOPs and vendor pages for TSS, does not appear in that row. A laboratory reporting under a permit should cite one of the three references the table lists, and confirm the edition its permit or accreditation body requires. NEMI also catalogues 2540 D as “Total Suspended Solids Dried at 103–105 °C” from Standard Methods Online (APHA, AWWA, WEF), noting it is approved by EPA for regulatory use, and refers readers to the Standard Methods site for the full method text. The wider rule set is covered in the guide to 40 CFR Part 136 water analysis methods.
Common misunderstandings and sources of error
- Thinking the filter and washing do not matter. The SOP says the filtration apparatus, filter material, pre-washing, post-washing and drying temperatures are specified because they affect results. Substituting a different filter or skipping the wash changes the number.
- Filtering too much sample. Residue above about 200 mg can form a crust that traps water, which biases the weight. This is why the SOP caps the residue window rather than just the volume.
- Ignoring slow filtration. If filtering takes a long time because the filter is clogging, the captured colloids can produce a high result, so the SOP restarts the analysis after 10 minutes.
- High dissolved solids. A sample rich in dissolved material needs a thorough wash of the filter so that dried dissolved salts are not weighed as suspended solids.
- Unrepresentative particles. The method allows large floating particles and agglomerates to be excluded when they are not representative of the sample.
- Treating TSS and turbidity as one measurement. See above: they are related, not the same.
Where it matters for laboratories
Wastewater, surface-water and drinking-water laboratories report TSS because suspended particles carry contaminants and affect treatment performance. For a laboratory writing or reviewing a procedure, the practical value of the SOP is its precision about acceptance limits: 90–110% recovery on the laboratory control sample, a relative percent difference of 15 between LCS and LCSD, and duplicate agreement within 5% of the average (equal to a 10% RPD) as Standard Methods 2540 D requires.
Frequently asked questions
What does a TSS test measure?
It measures the mass of solids retained by a standard glass fiber filter per litre of sample, after drying at 103–105 °C, reported in mg/L.
What temperature is used to dry the filter in the total suspended solids test?
103–105 °C, in both the Georgia EPD SOP for Standard Methods 2540 D and the Table IB entry in 40 CFR 136.3.
What is the practical range of the method?
The SOP states 1.0 to 20,000 mg/L, with the filtered residue targeted at 2.5 to 200 mg.
How long can a sample be held before TSS analysis?
The SOP gives a 7-day holding time with the sample stored at 0–6 °C, not frozen, and no chemical preservation.
Is EPA 160.2 still an approved TSS method?
It is not listed in the TSS row of 40 CFR 136.3 Table IB as read on 2026-11-15, which names Standard Methods 2540 D-2020, ASTM D5907-18 (A) and USGS I-3765-85.
Sources
- Georgia Environmental Protection Division, Laboratory SOP 3-005 Rev. 8, SM 2540 D Total Suspended Solids (effective 2021-06-10), https://epd.georgia.gov/document/document/sop-3-005-rev-8-sm2540d-tsspdf/download, read 2026-11-15.
- Electronic Code of Federal Regulations via Cornell LII, 40 CFR 136.3, Table IB, https://www.law.cornell.edu/cfr/text/40/136.3, read 2026-11-15.
- NEMI, Method Summary 2540 D, Total Suspended Solids Dried at 103–105 °C, https://www.nemi.gov/methods/method_summary/9819/, read 2026-11-15.
- US EPA, Parameter Factsheet: Turbidity (2021-07), https://www.epa.gov/system/files/documents/2021-07/parameter-factsheet_turbidity.pdf, read 2026-11-15.
Data: TSS gravimetric test (SM 2540 D): specification and QC values from a state laboratory SOP
Compiled from Georgia EPD laboratory SOP 3-005 Rev. 8 for Standard Methods 2540 D: filter, drying, constant-weight, residue window, holding-time and quality-control values for the total suspended solids test, cross-referenced with the 40 CFR 136.3 Table IB drying temperature.
| Parameter | Value in the SOP | Role in the test |
|---|---|---|
| Filter | 47 mm standard glass fiber filter (Whatman 934-AH, Gelman A/E or equivalent) | Retains the suspended solids |
| Drying temperature | 103-105 C (same in 40 CFR 136.3 Table IB) | Dries filter and residue |
| Filter pre-wash | At least 60 mL reagent water | Removes loose fibres before the tare weight |
| Filter drying time before weighing | At least 1 h (overnight preferable), then desiccator cooling at least 1 h | Brings filter to tare weight |
| Constant-weight criterion | Change below 4% of previous weight or 0.5 mg, whichever is less | Confirms drying is complete |
| Balance resolution | 0.1 mg, calibration verified each day of use with certified weights | Weighing |
| Dried residue window | 2.5 to 200 mg | Below 2.5 mg or above 200 mg with under 1000 mL filtered: corrective action and J-flag |
| Maximum filtration time | 10 minutes | Longer: restart with clean filter and smaller volume |
| Practical range | 1.0 to 20,000 mg/L | Applicable concentration range |
| Sample storage | 0-6 C, not frozen, no chemical preservative | Preservation |
| Holding time | 7 days | Time allowed to analysis |
| Method blank | 1000 mL reagent water per batch | Contamination check |
| Laboratory control sample | Celite 545 stock, 100 mg/L, 100 mL used, per batch of 20; limits 90-110% recovery, RPD 15 | Accuracy check |
| Sample duplicates | 10% of samples (minimum 1 for 1-10 samples, 2 for 11-20); agreement within 5% of average (10% RPD) | Precision check |
Sources
- Georgia EPD Laboratory SOP 3-005 Rev. 8, SM 2540 D Total Suspended Solids (accessed 2026-11-15)
- 40 CFR 136.3 Table IB (Cornell LII) (accessed 2026-11-15)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.