A digital pipette is a handheld liquid-handling device whose volume setting is shown on a digital display. The term covers two different instruments. In manufacturer usage, an electronic (digital) pipette is motor-driven: an electronically controlled piston aspirates and dispenses the liquid, and the user selects a mode rather than pressing a plunger through its full stroke. A manual pipette with a digital volume readout is still operated by thumb; only the volume window is numeric. A third, research-hardware meaning is covered below. This page separates them and compares them using the manufacturers’ own instructions for use.
What a digital pipette is
The Eppendorf Xplorer and Xplorer plus operating manual defines its pipettes as “motor-driven and electronically controlled piston-stroke pipettes for aspirating and dispensing liquids” that “work according to the air cushion principle” and need a suitable pipette tip fitted before use. Sartorius describes its Picus 2 electronic pipette in the same terms: motor-driven aspiration and dispensing, with a brake system that stops the piston at the correct position.
The second meaning is the manual pipette with a numeric readout. The Thermo Scientific Finnpipette Digital instructions for use call it “an autoclavable digital pipette” that operates on the air displacement principle with detachable, disposable tips, and state that “the adjusted delivery volume is displayed digitally on a readout window on the handle.” Nothing in that document describes a motor. Searchers and catalogues use “digital pipette” for both, so check whether a product page says motor-driven or electronic before assuming either.
Who uses it: laboratory staff trained in pipetting, in settings where the instructions for use apply. Both the Eppendorf and Thermo documents state that the pipettes are for laboratory use by trained personnel and are not intended for in vivo use.
How a digital pipette works and is built
Motor-driven (electronic) pipettes
In the Eppendorf Xplorer, a motor moves the piston and a central rocker control starts aspiration and dispensing. The housing colour varies with the volume range, and models cover 0.5 µL to 10 mL. The 10–1000 µL models have a spring-loaded tip cone, and a protection filter can be inserted in the 2.5, 5 and 10 mL models. Power comes from a lithium-polymer battery (750 mAh, 3.7 V) that the manual says needs about 3 hours to charge and supports about 1,400 dispensings in pipetting mode at medium speed with volumes up to 1000 µL on a single-channel pipette, measured continuously. The manual allows only Eppendorf batteries, warns against piercing or crushing them, and requires disposal according to legal requirements. Software updates are installed with the manufacturer’s update tool.
Sartorius lists, for the Picus 2, automated tip ejection, Bluetooth connection to a mobile app that adjusts settings and updates software, calibration reminders, password protection for saved programs and user-definable protocols.
Manual pipettes with a digital readout
The Finnpipette Digital IFU describes a mechanical air-displacement instrument made of “mechanically durable and chemically resistant materials” that allow repeated autoclaving of the complete pipette at 121 °C. Its tips (Finntips) are made of natural-colour polypropylene and are also autoclavable at 121 °C. Each model is colour-coded by volume range and listed against the tip types it takes.
For how tips and tip filters fit into this, see the guide to filtered pipette tips; for the basic instrument, see what a pipette is.
A third meaning: open-hardware digital pipettes for robots
The phrase also appears in robotics. A 2023 paper in Digital Discovery by Yoshikawa, Darvish, Vakili, Garg and Aspuru-Guzik, titled “Digital pipette: open hardware for liquid transfer in self-driving laboratories”, proposes an economical 3D-printed pipette intended to overcome the limitations of two-finger robot grippers. According to the paper’s abstract, it lets general-purpose robot arms achieve high-precision liquid transfer comparable to commercial devices. This is a research-hardware design for automated laboratories, not a handheld product.
Types and variants
The data table below the article compares three documented products by drive type, volume range, weight, autoclaving and operating limits. The variants that the sources themselves describe are these:
- Single-channel and multichannel. Eppendorf lists single-channel and eight-channel Xplorer models (the manual gives about 135 g for a 100 µL single-channel and about 213 g for a 100 µL eight-channel pipette). Thermo lists ten Finnpipette Digital single-channel models (0.2 µL to 10 mL) and seven multichannel models (0.5 µL to 300 µL, with 8, 12 or 16 channels).
- Fixed-volume. The Xplorer has a Fix operating mode for fixed volumes.
- Large-volume. Sartorius presents a single-channel Picus 2 for large-volume pipetting from 1.2 to 10 mL.
For the single- versus multichannel choice on manual instruments, see the guide to micropipette types, and for one manufacturer’s electronic model see Sartorius pipettes.
Operating modes of an electronic pipette
What most separates an electronic pipette from a manual one is that the pipetting sequence is selected, not performed by plunger travel. The Eppendorf Xplorer manual defines these settings on its mode dial:
| Mode | What the manual says it does |
|---|---|
| Pip (pipetting) | Aspirates in one step and dispenses in one step. |
| Dis (dispensing) | Dispenses equal partial volumes. |
| Ads (automatic dispensing) | Dispenses equal partial volumes at fixed time intervals. |
| P/M (pipetting and mixing) | Pipettes and then mixes. |
| Man (manual pipetting) | Aspirates and dispenses while the rocker is held pressed. |
| Prg (program) | Runs saved dispensing programs. |
| Fix (fixed volume) | Pipettes a fixed volume. |
| Spc (special) | Groups multiple aspiration, diluting, sequential dispensing, reverse pipetting (a larger volume taken up and dispensed with blow-out) and sequential pipetting. |
The dial also has Edit and Opt positions for editing programs and changing options. Sartorius names diluting, titrating and serial dispensing among the modes of the Picus 2. The Eppendorf “reverse pipetting” setting is the electronic version of the technique described in the CASRAI guide to reverse pipetting.
Specifications that matter
The values below are those the instructions for use state. Compare them on the same basis: a battery life figure, for example, applies only to the test condition the manufacturer gives.
- Volume range. Eppendorf Xplorer: 0.5 µL to 10 mL depending on model. Finnpipette Digital: 0.2 µL to 10 mL (single-channel) and 0.5 µL to 300 µL (multichannel). The Sartorius page gives no volume ranges for its electronic line other than 1.2–10 mL for the large-volume single-channel model.
- Operating environment. Eppendorf: 5–40 °C and 10–95 % humidity, indoor use. Thermo: +4 °C to +40 °C.
- Weight. Eppendorf: about 135 g (100 µL single-channel) and about 213 g (100 µL eight-channel). Sartorius gives about 100 g for the Picus 2.
- Battery. Eppendorf Xplorer: Li-polymer, 750 mAh, 3.7 V, about 3 h charge, about 1,400 dispensings under the stated condition.
- Performance testing. Eppendorf states that its test conditions follow DIN EN ISO 8655-6: 10 determinations per volume, water conforming to ISO 3696, 20–25 °C held to ±0.5 °C, standard pipetting mode at speed level 5, at 10 %, 50 % and 100 % of nominal volume, against ISO 8655 Part 2 or Part 5 limits. Thermo states that its pipettes are factory calibrated with distilled or deionized water using forward pipetting, that an analytical balance must be used for checks, and that ISO 8655 calibration limits are recommended. Its IFU ties balance readability to the test volume: 0.001 mg below 10 µL, 0.01 mg for 10–100 µL and 0.1 mg above 100 µL.
For calibration practice in general, see how and when to calibrate a pipette.
How to choose, using only what the instructions state
The manufacturers’ documents do not rank products, and this page does not either. They do give criteria that decide whether a given pipette is suitable:
- Volume range against your working volumes. Each model is limited to its stated range; Thermo lists the tip types matched to each model, and Eppendorf says pipettes are used with matching tips.
- Intended use. Both Eppendorf and Thermo exclude in vivo use and any liquid to be injected into the human body.
- Sterilisation need. Thermo permits autoclaving the complete Finnpipette Digital at 121 °C; Eppendorf permits autoclaving only the lower part of the Xplorer (conditions below). This difference matters if the pipette must be steam-sterilised.
- Ambient conditions. Operating temperature and humidity limits are stated for each product.
- Software and programs. The Eppendorf and Sartorius electronic pipettes store programs and receive software updates, which the Finnpipette Digital instructions do not mention.
Safe use and handling basics
- Use only trained personnel. Both manuals say so, and Thermo adds that the end user is responsible for validating the pipette and tip as a component of the analysing system.
- When handling hazardous or potentially infectious liquids, follow the safety instructions, wear protective clothing, eye protection and gloves, and follow the applicable biosafety level and safety data sheet. The Eppendorf manual refers to the WHO Laboratory Biosafety Manual and warns against use in explosive atmospheres. Thermo adds not to eject a tip towards anybody.
- Do not use the pipette for liquids to be injected into a human body, or for in vivo applications.
- Eppendorf: autoclave only the lower part, at 121 °C for 20 minutes at 1 bar, never above 121 °C, and without extra disinfectants or hypochlorite in the autoclave. A protection filter is autoclaved only once. Clean the lower parts after contamination or exposure to aggressive chemicals; they are wear parts.
- Eppendorf recommends regular checks by trained service personnel. Sartorius notes that basic maintenance and calibration are very similar to mechanical pipettes. Thermo recommends storing the Finnpipette Digital in an upright position when not in use.
- Charge and dispose of batteries only as the manual instructs.
Common mistakes
- Assuming a digital display means a motor. A numeric readout alone does not make a pipette electronic; the Finnpipette Digital has one and is operated manually.
- Autoclaving the whole electronic pipette. The Eppendorf manual limits autoclaving to the lower part, so the rest of the pipette is not autoclaved.
- Comparing a battery or accuracy figure without its test condition. The Eppendorf dispensing count and its ISO 8655 test conditions are each specific to the stated mode, speed, volume and temperature.
- Changing technique without re-checking. Thermo states that factory calibration assumes distilled or deionized water and forward pipetting, and that other techniques may affect the results. See pipetting technique.
- Using the wrong tip. Both manufacturers pair each volume range with specific tips.
Sources
- Eppendorf Xplorer / Xplorer plus operating manual (read 2026-11-18).
- Thermo Scientific Finnpipette Digital instructions for use (read 2026-11-18).
- Sartorius electronic pipettes product page (read 2026-11-18).
- Yoshikawa et al., Digital pipette: open hardware for liquid transfer in self-driving laboratories, Digital Discovery (2023), DOI 10.1039/D3DD00115F (record read 2026-11-18).
Data: Digital pipette specifications from three manufacturer documents
Drive type, volume range, weight, autoclave limits and operating temperature for the Eppendorf Xplorer (motor-driven), Thermo Finnpipette Digital (manual with digital readout) and Sartorius Picus 2 (motor-driven), compiled from each manufacturer's instructions for use or product page.
Period covered: 2026.
| Product | Drive type | Volume range | Autoclaving | Operating temperature |
|---|---|---|---|---|
| Eppendorf Xplorer / Xplorer plus | Motor-driven, electronically controlled | 0.5 µL to 10 mL by model | Lower part only, 121 °C, 20 min, 1 bar | 5 to 40 °C |
| Thermo Scientific Finnpipette Digital | Manual, digital volume readout | 0.2 µL to 10 mL single-channel; 0.5 to 300 µL multichannel | Complete pipette, repeated, 121 °C | +4 to +40 °C |
| Sartorius Picus 2 | Motor-driven | Not stated on the page (1.2 to 10 mL large-volume single-channel model) | Not stated on the page | Not stated on the page |
Sources
- Eppendorf Xplorer / Xplorer plus operating manual (accessed 2026-11-18)
- Thermo Scientific Finnpipette Digital instructions for use (accessed 2026-11-18)
- Sartorius electronic pipettes (accessed 2026-11-18)
Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.