Digital Pipette: What It Is, Modes and Specifications

A digital pipette is either motor-driven (electronic) or a manual pipette with a digital readout. Compare modes, ranges and autoclave limits.

Written and maintained by CASRAI Editorial Board

Published

On this page

A digital pipette is a handheld liquid-handling device whose volume setting is shown on a digital display. The term covers two different instruments. In manufacturer usage, an electronic (digital) pipette is motor-driven: an electronically controlled piston aspirates and dispenses the liquid, and the user selects a mode rather than pressing a plunger through its full stroke. A manual pipette with a digital volume readout is still operated by thumb; only the volume window is numeric. A third, research-hardware meaning is covered below. This page separates them and compares them using the manufacturers’ own instructions for use.

What a digital pipette is

The Eppendorf Xplorer and Xplorer plus operating manual defines its pipettes as “motor-driven and electronically controlled piston-stroke pipettes for aspirating and dispensing liquids” that “work according to the air cushion principle” and need a suitable pipette tip fitted before use. Sartorius describes its Picus 2 electronic pipette in the same terms: motor-driven aspiration and dispensing, with a brake system that stops the piston at the correct position.

The second meaning is the manual pipette with a numeric readout. The Thermo Scientific Finnpipette Digital instructions for use call it “an autoclavable digital pipette” that operates on the air displacement principle with detachable, disposable tips, and state that “the adjusted delivery volume is displayed digitally on a readout window on the handle.” Nothing in that document describes a motor. Searchers and catalogues use “digital pipette” for both, so check whether a product page says motor-driven or electronic before assuming either.

Who uses it: laboratory staff trained in pipetting, in settings where the instructions for use apply. Both the Eppendorf and Thermo documents state that the pipettes are for laboratory use by trained personnel and are not intended for in vivo use.

How a digital pipette works and is built

Motor-driven (electronic) pipettes

In the Eppendorf Xplorer, a motor moves the piston and a central rocker control starts aspiration and dispensing. The housing colour varies with the volume range, and models cover 0.5 µL to 10 mL. The 10–1000 µL models have a spring-loaded tip cone, and a protection filter can be inserted in the 2.5, 5 and 10 mL models. Power comes from a lithium-polymer battery (750 mAh, 3.7 V) that the manual says needs about 3 hours to charge and supports about 1,400 dispensings in pipetting mode at medium speed with volumes up to 1000 µL on a single-channel pipette, measured continuously. The manual allows only Eppendorf batteries, warns against piercing or crushing them, and requires disposal according to legal requirements. Software updates are installed with the manufacturer’s update tool.

Sartorius lists, for the Picus 2, automated tip ejection, Bluetooth connection to a mobile app that adjusts settings and updates software, calibration reminders, password protection for saved programs and user-definable protocols.

Manual pipettes with a digital readout

The Finnpipette Digital IFU describes a mechanical air-displacement instrument made of “mechanically durable and chemically resistant materials” that allow repeated autoclaving of the complete pipette at 121 °C. Its tips (Finntips) are made of natural-colour polypropylene and are also autoclavable at 121 °C. Each model is colour-coded by volume range and listed against the tip types it takes.

For how tips and tip filters fit into this, see the guide to filtered pipette tips; for the basic instrument, see what a pipette is.

A third meaning: open-hardware digital pipettes for robots

The phrase also appears in robotics. A 2023 paper in Digital Discovery by Yoshikawa, Darvish, Vakili, Garg and Aspuru-Guzik, titled “Digital pipette: open hardware for liquid transfer in self-driving laboratories”, proposes an economical 3D-printed pipette intended to overcome the limitations of two-finger robot grippers. According to the paper’s abstract, it lets general-purpose robot arms achieve high-precision liquid transfer comparable to commercial devices. This is a research-hardware design for automated laboratories, not a handheld product.

Types and variants

The data table below the article compares three documented products by drive type, volume range, weight, autoclaving and operating limits. The variants that the sources themselves describe are these:

  • Single-channel and multichannel. Eppendorf lists single-channel and eight-channel Xplorer models (the manual gives about 135 g for a 100 µL single-channel and about 213 g for a 100 µL eight-channel pipette). Thermo lists ten Finnpipette Digital single-channel models (0.2 µL to 10 mL) and seven multichannel models (0.5 µL to 300 µL, with 8, 12 or 16 channels).
  • Fixed-volume. The Xplorer has a Fix operating mode for fixed volumes.
  • Large-volume. Sartorius presents a single-channel Picus 2 for large-volume pipetting from 1.2 to 10 mL.

For the single- versus multichannel choice on manual instruments, see the guide to micropipette types, and for one manufacturer’s electronic model see Sartorius pipettes.

Operating modes of an electronic pipette

What most separates an electronic pipette from a manual one is that the pipetting sequence is selected, not performed by plunger travel. The Eppendorf Xplorer manual defines these settings on its mode dial:

Mode What the manual says it does
Pip (pipetting) Aspirates in one step and dispenses in one step.
Dis (dispensing) Dispenses equal partial volumes.
Ads (automatic dispensing) Dispenses equal partial volumes at fixed time intervals.
P/M (pipetting and mixing) Pipettes and then mixes.
Man (manual pipetting) Aspirates and dispenses while the rocker is held pressed.
Prg (program) Runs saved dispensing programs.
Fix (fixed volume) Pipettes a fixed volume.
Spc (special) Groups multiple aspiration, diluting, sequential dispensing, reverse pipetting (a larger volume taken up and dispensed with blow-out) and sequential pipetting.

The dial also has Edit and Opt positions for editing programs and changing options. Sartorius names diluting, titrating and serial dispensing among the modes of the Picus 2. The Eppendorf “reverse pipetting” setting is the electronic version of the technique described in the CASRAI guide to reverse pipetting.

Specifications that matter

The values below are those the instructions for use state. Compare them on the same basis: a battery life figure, for example, applies only to the test condition the manufacturer gives.

  • Volume range. Eppendorf Xplorer: 0.5 µL to 10 mL depending on model. Finnpipette Digital: 0.2 µL to 10 mL (single-channel) and 0.5 µL to 300 µL (multichannel). The Sartorius page gives no volume ranges for its electronic line other than 1.2–10 mL for the large-volume single-channel model.
  • Operating environment. Eppendorf: 5–40 °C and 10–95 % humidity, indoor use. Thermo: +4 °C to +40 °C.
  • Weight. Eppendorf: about 135 g (100 µL single-channel) and about 213 g (100 µL eight-channel). Sartorius gives about 100 g for the Picus 2.
  • Battery. Eppendorf Xplorer: Li-polymer, 750 mAh, 3.7 V, about 3 h charge, about 1,400 dispensings under the stated condition.
  • Performance testing. Eppendorf states that its test conditions follow DIN EN ISO 8655-6: 10 determinations per volume, water conforming to ISO 3696, 20–25 °C held to ±0.5 °C, standard pipetting mode at speed level 5, at 10 %, 50 % and 100 % of nominal volume, against ISO 8655 Part 2 or Part 5 limits. Thermo states that its pipettes are factory calibrated with distilled or deionized water using forward pipetting, that an analytical balance must be used for checks, and that ISO 8655 calibration limits are recommended. Its IFU ties balance readability to the test volume: 0.001 mg below 10 µL, 0.01 mg for 10–100 µL and 0.1 mg above 100 µL.

For calibration practice in general, see how and when to calibrate a pipette.

How to choose, using only what the instructions state

The manufacturers’ documents do not rank products, and this page does not either. They do give criteria that decide whether a given pipette is suitable:

  • Volume range against your working volumes. Each model is limited to its stated range; Thermo lists the tip types matched to each model, and Eppendorf says pipettes are used with matching tips.
  • Intended use. Both Eppendorf and Thermo exclude in vivo use and any liquid to be injected into the human body.
  • Sterilisation need. Thermo permits autoclaving the complete Finnpipette Digital at 121 °C; Eppendorf permits autoclaving only the lower part of the Xplorer (conditions below). This difference matters if the pipette must be steam-sterilised.
  • Ambient conditions. Operating temperature and humidity limits are stated for each product.
  • Software and programs. The Eppendorf and Sartorius electronic pipettes store programs and receive software updates, which the Finnpipette Digital instructions do not mention.

Safe use and handling basics

  • Use only trained personnel. Both manuals say so, and Thermo adds that the end user is responsible for validating the pipette and tip as a component of the analysing system.
  • When handling hazardous or potentially infectious liquids, follow the safety instructions, wear protective clothing, eye protection and gloves, and follow the applicable biosafety level and safety data sheet. The Eppendorf manual refers to the WHO Laboratory Biosafety Manual and warns against use in explosive atmospheres. Thermo adds not to eject a tip towards anybody.
  • Do not use the pipette for liquids to be injected into a human body, or for in vivo applications.
  • Eppendorf: autoclave only the lower part, at 121 °C for 20 minutes at 1 bar, never above 121 °C, and without extra disinfectants or hypochlorite in the autoclave. A protection filter is autoclaved only once. Clean the lower parts after contamination or exposure to aggressive chemicals; they are wear parts.
  • Eppendorf recommends regular checks by trained service personnel. Sartorius notes that basic maintenance and calibration are very similar to mechanical pipettes. Thermo recommends storing the Finnpipette Digital in an upright position when not in use.
  • Charge and dispose of batteries only as the manual instructs.

Common mistakes

  • Assuming a digital display means a motor. A numeric readout alone does not make a pipette electronic; the Finnpipette Digital has one and is operated manually.
  • Autoclaving the whole electronic pipette. The Eppendorf manual limits autoclaving to the lower part, so the rest of the pipette is not autoclaved.
  • Comparing a battery or accuracy figure without its test condition. The Eppendorf dispensing count and its ISO 8655 test conditions are each specific to the stated mode, speed, volume and temperature.
  • Changing technique without re-checking. Thermo states that factory calibration assumes distilled or deionized water and forward pipetting, and that other techniques may affect the results. See pipetting technique.
  • Using the wrong tip. Both manufacturers pair each volume range with specific tips.

Sources

Data: Digital pipette specifications from three manufacturer documents

Drive type, volume range, weight, autoclave limits and operating temperature for the Eppendorf Xplorer (motor-driven), Thermo Finnpipette Digital (manual with digital readout) and Sartorius Picus 2 (motor-driven), compiled from each manufacturer's instructions for use or product page.

Period covered: 2026.

Digital pipette specifications from three manufacturer documents
ProductDrive typeVolume rangeAutoclavingOperating temperature
Eppendorf Xplorer / Xplorer plusMotor-driven, electronically controlled0.5 µL to 10 mL by modelLower part only, 121 °C, 20 min, 1 bar5 to 40 °C
Thermo Scientific Finnpipette DigitalManual, digital volume readout0.2 µL to 10 mL single-channel; 0.5 to 300 µL multichannelComplete pipette, repeated, 121 °C+4 to +40 °C
Sartorius Picus 2Motor-drivenNot stated on the page (1.2 to 10 mL large-volume single-channel model)Not stated on the pageNot stated on the page

Sources

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.

Follow CASRAI

Research-administration guidance, standards updates and independent tool reviews.

Ask CASRAI · free to try

Ask about Digital Pipette: What It Is, Modes and Specifications

2 free questions a day, no account.

An AI assistant for research: funder rules, compliance, methods and publishing. It links the sources behind its answers and labels answers that come from a web search.

Answers draw on CASRAI's guides and dictionary plus the federal and funder documents we index: Federal Register, Grants.gov, Regulations.gov and UKRI.

Works on this site and inside Claude, Cursor and the AI tools you already use.

Everything CASRAI publishes — this page, the dictionary, the guides and the news — stays open to read, with no account and no card.

Referenced across the research world

University of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logoUniversity of Cambridge logoColumbia University logoCrossref logoUniversity of Edinburgh logoHarvard University logoUniversity of Oxford logoPrinceton University logoStanford School of Medicine logoUniversity College London logoORCID logo
  • University of Cambridge logo
  • Columbia University logo
  • Crossref logo
  • University of Edinburgh logo
  • Harvard University logo
  • University of Oxford logo
  • Princeton University logo
  • Stanford School of Medicine logo
  • University College London logo
  • ORCID logo

View CASRAI adoption →